Bard Access Systems, Inc. · PowerGlide Pro with Reinforced Tip, 18G 8cm, Basic Kit · Class 2

All models of PowerGlide (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00801741142642), received between Sep 29, 2022 and Oct 15, 2025 — classified as 2 injuries, 8 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation3
  • Material Frayed3
  • Detachment of Device or Device Component2
  • Break1
  • Fluid/Blood Leak1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient2
  • Hematoma1
  • Infiltration into Tissue1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Material Frayed; Material Frayed

    ADDITIONAL INFORMATION: WE HAD TO REDO THE CATHETER INSERTION WITH A DIFFERENT DEVICE. THERE WERE NO COMPLICATIONS FOR THE PATIENT, OTHER THAN A SECOND PUNCTURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23292307 ↗
  • InjuryMar 11, 2025Break; Break

    IT WAS REPORTED VIA MEDWATCH "NURSE AT PATIENT BEDSIDE TO REMOVE MIDLINE IV DUE TO LEAKING. WHEN THE NURSE ATTEMPTED TO REMOVE THE MIDLINE, THE CATHETER BROKE AT THE HUB LEAVING PART OF THE CATHETER IN THE PATIENT'S ARM. INTERVENTION REQUIRED TO REMOVE REMAINING CATHETER."

    Patient problems: Foreign Body In Patient

    MAUDE report 21574471 ↗
  • MalfunctionDec 13, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED A MIDLINE BROKE AND INFILTRATED IN THE PATIENT ARM. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Infiltration into Tissue

    MAUDE report 20924903 ↗
  • MalfunctionDec 11, 2024Material Frayed; Material Frayed

    IT WAS REPORTED THAT A MIDLINE JUST INFILTRATED IN THE PATIENT ARM AND THE GUIDEWIRE DISINTEGRATED IN THE PATIENT ARM WHEN WE WERE ADVANCING IT AND IT CAUSED A BIG HEMATOMA ON OUR PATIENT. NO OTHER INFORMATION WAS PROVIDED.

    Patient problems: Hematoma

    MAUDE report 20907085 ↗
  • MalfunctionOct 19, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE CUSTOMER "UPON PULLING BACK ON DEVICE HOUSING TO SAFETY WITH RIGHT HAND WHILE HOLDING WINGS STEADY WITH LEFT HAND THE NEEDLE SEPARATED FROM THE HOUSING." THERE WAS NO IMPACT TO THE PATIENT. NO OTHER INFORMATION WAS PROVIDED. (B)(6) 2023 - SIXTEEN SAMPLES WERE RETURNED FOR EVALUATION. THE NEEDLE DETACHED EASILY FROM THE HOUSING ON TWO OF THE RETURNED SAMPLES. THIS REPORT ADDRESSES THE SECOND SAMPLE WITH AN EASILY DETACHED NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17968422 ↗
  • MalfunctionMar 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE CUSTOMER "UPON PULLING BACK ON DEVICE HOUSING TO SAFETY WITH RIGHT HAND WHILE HOLDING WINGS STEADY WITH LEFT HAND THE NEEDLE SEPARATED FROM THE HOUSING." THERE WAS NO IMPACT TO THE PATIENT. NO OTHER INFORMATION WAS PROVIDED. ON 10/13/2023, SIXTEEN SAMPLES WERE RETURNED FOR EVALUATION. THE NEEDLE DETACHED EASILY FROM THE HOUSING ON TWO OF THE RETURNED SAMPLES. THIS REPORT ADDRESSES THE FIRST SAMPLE WITH AN EASILY DETACHED NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16613131 ↗
  • MalfunctionFeb 8, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER ISSUES WITH 2 MIDLINE CATHETERS KINKING IN THE VESSEL UPON INSERTION. SONOSITE GUIDANCE WAS USED AND THE SURGEON WAS PRESENT. IT WAS STATED THERE WAS NO PATIENT OR USER IMPACT. THIS REPORT ADDRESSES THE SECOND DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 16318668 ↗
  • MalfunctionFeb 8, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER ISSUES WITH 2 MIDLINE CATHETERS KINKING IN THE VESSEL UPON INSERTION. SONOSITE GUIDANCE WAS USED AND THE SURGEON WAS PRESENT. IT WAS STATED THERE WAS NO PATIENT OR USER IMPACT. THIS REPORT ADDRESSES THE FIRST DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 16318627 ↗
  • MalfunctionFeb 3, 2023Material Deformation; Material Deformation

    IT WAS REPORTED BY THE CUSTOMER, HAVING ISSUES WITH MIDLINE CATHETERS KINKING IN THE VESSEL UPON INSERTION. SONOSITE GUIDANCE WAS USED, AND THE SURGEON WAS PRESENT.

    Patient problems: Insufficient Information

    MAUDE report 16296339 ↗
  • InjurySep 29, 2022Material Frayed; Improper or Incorrect Procedure or Method; Material Frayed; Improper or Incorrect Procedure or Method

    IT WAS REPORTED VIA MEDWATCH "DURING A MIDLINE CATHETER PLACEMENT, THE VEIN WAS ACCESSED, HOWEVER, THE CATHETER WAS UNABLE TO BE SUCCESSFULLY THREADED OVER THE DEVICE'S GUIDE WIRE. THE PROCEDURE WAS STOPPED AND THE DEVICE WAS WITHDRAWN. UPON REMOVAL AND INSPECTION OF THE MIDLINE DEVICE, THE GUIDEWIRE WAS NOT PRESENT OR ATTACHED TO THE DEVICE. X-RAY OBTAINED AND REVEALED THE PRESENCE OF FOREIGN BODY TO THE LEFT LOWER CALF. PATIENT WAS TAKEN TO THE OPERATING ROOM FOR FOREIGN BODY REMOVAL." ADDITIONAL INFORMATION RECEIVED 09/27/2022: WERE ALL THE MISSING PIECES RETRIEVED AND ACCOUNTED FOR? YES.

    Patient problems: Foreign Body In Patient

    MAUDE report 15505333 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices

All Catheter,intravascular,therapeutic,short-term less than 30 days devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page