Skip to content

PowerGlide F218101TRecalls & reported problems

Bard Access Systems, Inc. · PowerGlide Pro with Reinforced Tip, 18G 10cm, Full Kit wo GIVa · Class 2

All models of PowerGlide (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801741142758), received between Jul 23, 2024 and Aug 5, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Activation Problem1
  • Material Frayed1

Patient problems

  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 5, 2026Break

    It was reported by the customer that Powerglide shearing off the tip yesterday and getting stuck in a patient.Additional information received:It was retained in the tissue per CT scan. It was not removed. They monitored patient for multiple days and patient had no symptoms and the vascular surgery provider recommended no follow up surgery. During insertion, guidewire mechanism was engaged with no resistance but attempt to move midline mechanism forward and halfway in met resistance. Attempted to retract and was met again with resistance. Midline was removed entirely with the retracting resistance. They had done a CT scan that had shown it in patient arm, venogram was done to confirm it was not in the vein. No additional intervention aside from the diagnostic testing post procedure for catheter location. No further details provided.

    Patient problems: Foreign Body In Patient

    MAUDE report 50144100 ↗
  • MalfunctionSep 23, 2025Material Frayed; Material Frayed

    IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE WAS COILED. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23128912 ↗
  • MalfunctionAug 18, 2025Activation Problem; Activation Problem

    IT WAS REPORTED THAT POWERGLIDE NEEDLE FAILED TO SAFETY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22834185 ↗
  • InjuryJul 23, 2024Break; Break

    IT WAS REPORTED, CATHETER REPORTED TO BE IN AT 4CM, LEFT BASILIC VEIN WITH A BLOOD RETURN. GUIDEWIRE ADVANCED WITHOUT PROBLEMS, ATTEMPTED TO ADVANCE CATHETER, MET RESISTANCE, PULLED THE CATHETER BACK FLUSH. ATTEMPTED TO RETRACT THE GUIDEWIRE AND IT FAILED. PULLED DEVICE BUT ONLY CATHETER AND NEEDLE CAME OUT BUT NO GUIDEWIRE. GUIDEWIRE SHEARING ALONG WITH 3CM OF CATHETER TIP. UPON INSPECTION NOTED ONLY 7CM OF CATHETER REMOVED. NO COLLATERAL BRANCHING OF VESSEL NOTED. GUIDEWIRE CHECKED PRIOR TO PLACEMENT ATTEMPT, NOTHING NOTED TO BE UNUSUAL. X-RAY CONFIRMATION OF GUIDEWIRE IN MID UPPER ARM. REQUIRED SURGICAL INTERVENTION TO REMOVE GUIDEWIRE AND CATHETER TIP FROM BASILIC VEIN.

    Patient problems: Foreign Body In Patient

    MAUDE report 19807148 ↗

Other Catheter,intravascular,therapeutic,short-term less than 30 days devices