Phasix 1221015Recalls & reported problems
Davol Inc. · Phasix ST Mesh with Open Positioning System, 10 cm x 15 cm (4" x 6"), Oval · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00801741163203), received between Feb 25, 2025 and Feb 25, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 25, 2025Material Separation; Material Separation
AS REPORTED, THE PHASIX ST W/ OPS MESH WAS QUICKLY DIPPED IN SALINE PRIOR TO USE. REPORTEDLY, DURING THE LAPAROSCOPIC HERNIA REPAIR CONVERTED TO A PLANNED OPEN PROCEDURE ON (B)(6) 2025, THE SURGEON IMPLANTED THE PHASIX ST W/ OPS MESH AND FIXATED IT AROUND THE PERIMETER OF THE MESH. ONCE FIXATION WAS COMPLETE, HE LAPAROSCOPICALLY WENT TO OBSERVE THE MESH AND NOTICED SOME OF THE MESH WAS BOWING SO HE DECIDED TO ADD A FEW MORE FASTENERS. THE SURGEON FIXATED A FEW AND THEN THE PHYSICIAN ASSISTANT (PA) FIXATED A FEW BUT WHEN THE PA WAS ADDING THE EXTRA FIXATION, IT WAS NOTICED THAT THE SEPRA BARRIER WAS ¿DELAMINATED FROM THE PHASIX ST MESH, ABOUT 2 INCHES OF IT WAS HANGING FREELY.¿ THE SURGEON AND PA DECIDED TO USE A COUPLE MORE FASTENERS TO REATTACH THE SEPRA BARRIER TO THE MESH AND TO COVER THE AREAS THAT DID NOT HAVE THE SEPRA BARRIER WITH THE PATIENT¿S OMENTUM. THE SURGEON DECIDED TO LEAVE THE MESH IN PLACE AS IT WAS WELL FIXATED AND FELT IT WOULD HAVE BEEN MORE PROBLEMATIC IF HE HAD REMOVED IT AND REPLACED THE MESH. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21461207 ↗
Other Mesh, surgical, polymeric devices
- PROCEED PVPMJohnson & Johnson International Inc. · Mesh, surgical, polymeric1,233 reports
- PROCEED PCDN1Johnson & Johnson International Inc. · Mesh, surgical, polymeric744 reports
- PROCEED PCDG1Johnson & Johnson International Inc. · Mesh, surgical, polymeric1 recall1 death637 reports
- PROCEED PVPSJohnson & Johnson International Inc. · Mesh, surgical, polymeric622 reports
- ProGrip LPG1510Sofradim Production · Mesh, surgical, polymeric2 deaths616 reports
- VENTRALEX 0010302Davol Inc. · Mesh, surgical, polymeric4 deaths531 reports
- BARD 0112680Davol Inc. · Mesh, surgical, polymeric1 death494 reports
- ProGrip TEM1509GSofradim Production · Mesh, surgical, polymeric3 deaths471 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page