Skip to content

LUTONIX® Drug Coated Balloon LX3513071505FRecalls & reported problems

Lutonix, Inc. · Lutonix 035 DCB, 7mm x 150mm x 130cm, 5F · Class 3

All models of LUTONIX Drug Coated Balloon (107)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00801741163463), received between May 26, 2022 and Jun 11, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture2
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1
  • Failure to Advance1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG COATED BALLOON WAS ALLEGEDLY RUPTURED WHEN IT REACHED NOMINAL PRESSURE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19508490 ↗
  • MalfunctionFeb 1, 2023Material Rupture; Failure to Advance; Device-Device Incompatibility; Material Deformation; Material Rupture; Failure to Advance; Device-Device Incompatibility; Material Deformation

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG COATED BALLOON WAS ALLEGEDLY UNABLE TO BE ADVANCED TO THE TARGET LESION. IT WAS FURTHER REPORTED THAT THE DRUG COATED BALLOON ALLEGEDLY STUCK ON THE WIRE. REPORTEDLY, UPON REMOVAL OF THE DEVICE, IT WAS FOUND THAT THE BALLOON ALLEGEDLY SNAGGED ON THE CATHETER AND LOOKED LIKE IT WAS STARTING TO ELONGATE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16277554 ↗
  • MalfunctionMay 26, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE CENTRAL PART OF THE DRUG-COATED BALLOON ALLEGEDLY BROKEN INSIDE THE PATIENT VESSEL. IT WAS FURTHER REPORTED THAT FRAGMENT OF THE BALLOON REMAINS INSIDE THE VESSEL. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14510995 ↗

Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices

All Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page