LUTONIX® Drug Coated Balloon LX1813053005FRecalls & reported problems

Lutonix, Inc. · Lutonix 018 5X300MMX130CM · Class 3

All models of LUTONIX Drug Coated Balloon (107)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00801741184710), received between Jan 28, 2021 and Mar 26, 2025 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent7
  • Inflation Problem6
  • Failure to Advance2
  • Material Rupture2
  • Deflation Problem1
  • Detachment of Device or Device Component1
  • Device-Device Incompatibility1
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2025Deflation Problem; Inflation Problem; Material Twisted/Bent; Deflation Problem; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG-COATED BALLOON WAS ALLEGEDLY TWISTED. IT WAS FURTHER REPORTED THAT THE DISTAL END OF THE BALLOON ALLEGEDLY WOULD NOT DEFLATE. REPORTEDLY, THE BALLOON WAS ALLEGEDLY INFLATED AND DEFLATED SEVERAL TIMES. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21695374 ↗
  • MalfunctionMar 12, 2025Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE DRUG COATED BALLOON ALLEGEDLY DEVELOPED A SMALL KINK IN MIDWAY DURING INFLATION. IT WAS FURTHER REPORTED THAT THE DRUG COATED BALLOON ALLEGEDLY TWISTED UPON INFLATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21581942 ↗
  • MalfunctionOct 1, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH, THE DRUG-COATED BALLOON ALLEGEDLY LEAKING FROM THE HUB. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20342499 ↗
  • MalfunctionJun 7, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG COATED BALLOON ALLEGEDLY RUPTURED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19482561 ↗
  • MalfunctionApr 26, 2023Failure to Advance

    DEVICE WOULDN'T ADVANCE THROUGH SHEATH FOR A SECOND PASS. NEW CATHETER OPENED AND WORKED. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16815218 ↗
  • MalfunctionDec 13, 2022Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG COATED BALLOON ALLEGEDLY FAILED TO INFLATE. IT WAS FURTHER REPORTED THAT THE BALLOON WAS ALLEGEDLY WRAPPED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15967292 ↗
  • MalfunctionNov 14, 2022Difficult to Remove; Failure to Advance; Detachment of Device or Device Component; Device-Device Incompatibility; Difficult to Remove; Failure to Advance; Detachment of Device or Device Component; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN SUPERFICIAL FEMORAL ARTERY, THE DRUG-COATED BALLOON ALLEGEDLY FAILED TO ADVANCE OVER THE WIRE. IT WAS FURTHER REPORTED THAT THE BALLOON WAS ALLEGEDLY DIFFICULT TO REMOVE. REPORTEDLY, THE CATHETER SHAFT GOT DETACHED AND THE WHOLE DEVICE WAS REMOVED AS ONE UNIT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15787054 ↗
  • MalfunctionSep 20, 2022Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE DISTAL SUPERFICIAL FEMORAL ARTERY VIA LEFT FEMORAL ARTERY, THE DRUG-COATED BALLOON ALLEGEDLY HAD INFLATION PROBLEM. IT WAS FURTHER REPORTED THAT THE BALLOON ALLEGEDLY GOT TWISTED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15454310 ↗
  • MalfunctionFeb 15, 2022Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN SUPERFICIAL FEMORAL ARTERY (SFA), THE DRUG COATED BALLOON ALLEGEDLY HAD DIFFICULTY TO CROSS THE LESION. IT WAS FURTHER REPORTED THAT UPON REACHING THE TARGET, THE DRUG COATED BALLOON ALLEGEDLY RUPTURED UPON INFLATION. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13530951 ↗
  • MalfunctionJan 4, 2022Difficult to Insert; Material Twisted/Bent; Difficult to Insert; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH THE GROIN FEMORAL TO TREAT THE SFA VIA A CONTRALATERAL APPROACH, THE BALLOON ALLEGEDLY TWISTED UPON INFLATION. IT WAS FURTHER REPORTED THAT BALLOON WAS REMOVED FORM PATIENT AND TRIED TO INSERT AGAIN, HOWEVER, THE BALLOON WAS ALLEGEDLY UNABLE TO INSERT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13151450 ↗
  • MalfunctionJun 30, 2021Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG COATED BALLOON ALLEGEDLY INFLATED WITH A NAPKIN RING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12092594 ↗
  • MalfunctionJan 28, 2021Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE ALLEGEDLY TWISTED AND FAILED TO INFLATE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11241937 ↗

Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices

All Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page