LUTONIX® Drug Coated Balloon LX1813063005FRecalls & reported problems

Lutonix, Inc. · Lutonix 018 6X300MMX130CM · Class 3

All models of LUTONIX Drug Coated Balloon (107)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Lutonix drug-coated balloon catheter is used during angioplasty procedures. Reports mostly describe the balloon twisting or folding during inflation so it does not inflate fully, with trouble deflating and occasional rupture. There are 23 reports that name its identifier, with no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
22
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 00801741184727), received between Oct 1, 2020 and Apr 8, 2026 — classified as 22 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (19 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inflation Problem19
  • Material Twisted/Bent19
  • Deflation Problem3
  • Difficult to Remove2
  • Failure to Advance2
  • Material Rupture2
  • Device-Device Incompatibility1
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 8, 2026Inflation Problem; Material Twisted/Bent

    ON (B)(6), 2026, A PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE LUTONIX 018 DRUG COATED BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER WAS ALLEGEDLY TWISTED DURING INFLATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24814083 ↗
  • MalfunctionMar 16, 2026Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2026, A 60-YEAR-OLD MALE PATIENT UNDERWENT A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE USING THE LUTONIX 018 DRUG COATED BALLOON CATHETER. DURING THE PROCEDURE, THE BALLOON CATHETER ALLEGEDLY TWISTED DURING INFLATION. THE PROCEDURE WAS COMPLETED USING ANOTHER BALLOON. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603600 ↗
  • MalfunctionFeb 9, 2026Deflation Problem; Inflation Problem; Difficult to Remove; Material Twisted/Bent

    ON (B)(6) 2026, AN 80 YEAR OLD FEMALE PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE BY USING THE LUTONIX 018 DRUG COATED BALLOON. IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON ALLEGEDLY TWISTED AND WAS TROUBLE DEFLATING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24289671 ↗
  • MalfunctionFeb 5, 2026Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE BY USING THE LUTONIX 018 DRUG COATED BALLOON. IT WAS REPORTED THAT DURING THE PROCEDURE, THE BALLOON ALLEGEDLY TWISTED DURING INFLATION. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24265380 ↗
  • MalfunctionFeb 5, 2026Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2026, A PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE USING THE LUTONIX 018 DRUG COATED BALLOON. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY WRAPPED IN DEPLOYMENT AND NOT INFLATE FULLY. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24265302 ↗
  • MalfunctionDec 12, 2025Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS DRUG-COATED BALLOON ANGIOPLASTY (DCB) PROCEDURE USING THE LUTONIX 018 DRUG-COATED DILATATION CATHETER. DURING THE PROCEDURE, THE BALLOON ALLEGEDLY FAILED INFLATE, LEAVING A SMALL PINCH IN THE BALLOON. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23788174 ↗
  • MalfunctionNov 19, 2025Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2025, A PATIENT UNDERWENT A PERCUTANEOUS DRUG COATED BALLOON ANGIOPLASTY PROCEDURE USING THE LUTONIX BALLOON CATHETER. DURING THE PROCEDURE THERE WAS TIGHT PINCH MIDWAY IN THE BALLOON. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE, AS IT DID NOT POSE ANY DANGER TO THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23592095 ↗
  • MalfunctionAug 25, 2025Deflation Problem; Inflation Problem; Material Twisted/Bent; Deflation Problem; Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2025, 86-YEAR-OLD MALE PATIENT UNDERWENT AN ANGIOPLASTY PROCEDURE IN SUPERFICIAL FEMORAL ARTERY/POPLITEAL ARTERY VIA CONTRALATERAL APPROACH USING THE LUTONIX 018 DRUG COATED DILATATION CATHETER IN WHICH THE BALLOON ALLEGEDLY GOT TWISTED IN 2 AREAS AND TROUBLE DEFLATING. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22881927 ↗
  • MalfunctionAug 18, 2025Inflation Problem; Difficult to Remove; Material Twisted/Bent; Inflation Problem; Difficult to Remove; Material Twisted/Bent

    ON (B)(6) 2025, A PATIENT UNDERWENT A ANGIOPLASTY PROCEDURE USING THE LUTONIX 018 DRUG COATED BALLOON (DCB) PTA CATHETER. DURING THE PROCEDURE, IT WAS REPORTED THAT THE BALLOON ALLEGEDLY FAILED TO OPEN FULLY DURING INFLATION, LEAVING A SMALL PINCH IN THE BALLOON. IT WAS FURTHER REPORTED THAT BALLOON WAS DEFLATED AND STRUGGLED TO BE REMOVED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22824865 ↗
  • MalfunctionJun 10, 2025Deflation Problem; Inflation Problem; Material Twisted/Bent; Deflation Problem; Inflation Problem; Material Twisted/Bent

    ON (B)(6) 2025, A 69-YEAR-OLD FEMALE PATIENT UNDERWENT A PROCEDURE USING THE LUTONIX 018 DRUG COATED BALLOON (DCB) PTA CATHETER. DURING THE PROCEDURE, THE BALLOON WAS ALLEGEDLY TWISTED DURING INFLATION AND HAD TROUBLE IN DEFLATING. IT WAS FURTHER REPORTED THAT TRIED RE-INFLATING IT TWICE AND POPPED IT WITH ANOTHER WIRE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22175673 ↗
  • MalfunctionApr 22, 2025Material Rupture; Failure to Advance; Material Rupture; Failure to Advance

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA RIGHT FEMORAL ACCESS, THE DRUG-COATED BALLOON ALLEGEDLY DID NOT CROSS THE LESION TO THE INTENDED TREATMENT SITE IN AN OPEN VESSEL. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21879030 ↗
  • MalfunctionDec 24, 2024Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY VIA CONTRALATERAL APPROACH, THE DRUG-COATED BALLOON WAS ALLEGEDLY TWISTED IN TWO SPOTS. IT WAS FURTHER REPORTED THAT THE BALLOON WAS DEFLATED AND REINFLATED AND ITS STILL TWISTED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21009700 ↗
  • MalfunctionOct 14, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG-COATED BALLOON ALLEGEDLY NEVER HELD PRESSURE ON INFLATION AND GOT BLOOD RETURN FROM THE BALLOON. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20436944 ↗
  • MalfunctionFeb 29, 2024Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE CALCIFIED SUPERFICIAL FEMORAL ARTERY VIA THE RIGHT GROIN, THE DRUG-COATED BALLOON ALLEGEDLY TWISTED UPON INFLATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18808824 ↗
  • MalfunctionAug 6, 2023Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE, THE DRUG-COATED BALLOON ALLEGEDLY TWISTED UPON INFLATION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17467301 ↗
  • MalfunctionJun 25, 2023Difficult to Insert; Material Puncture/Hole; Failure to Advance; Device-Device Incompatibility; Difficult to Insert; Material Puncture/Hole; Failure to Advance; Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE GUIDEWIRE WAS ALLEGEDLY FAILED TO GO THROUGH THE WIRE LUMEN. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE ALLEGEDLY PROTRUDED THROUGH THE EITHER SIDE OF THE NYLON MATERIAL ON THE BALLOON OR THE PEEL AWAY PROTECTOR. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17196165 ↗
  • MalfunctionNov 30, 2022Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH COMMON FEMORAL ARTERY CONTRALATERAL APPROACH, THE DEVICE ALLEGEDLY TWISTED. IT WAS FURTHER REPORTED THAT THE DEVICE ALLEGEDLY FAILED TO INFLATE. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15884090 ↗
  • MalfunctionDec 30, 2021Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE VIA LEFT SUPERFICIAL FEMORAL ARTERY THROUGH CONTRALATERAL APPROACH, THE DRUG COATED BALLOON ALLEGEDLY HAD INFLATION ISSUE AND WRAPPED UPON ITSELF. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13117366 ↗
  • MalfunctionJun 10, 2021Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE IN THE RIGHT SFA VIA ANTIGRADE APPROACH, THE DEVICE ALLEGEDLY TWISTED AND HAD AN INFLATION ISSUE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11972520 ↗
  • MalfunctionDec 21, 2020Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A DRUG COATED BALLOON ANGIOPLASTY PROCEDURE THROUGH LEFT SUPERFICIAL FEMORAL ARTERY, THE DEVICE ALLEGEDLY HAD AN INFLATION ISSUE AND MATERIAL TWISTED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11044516 ↗
  • MalfunctionDec 3, 2020Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED DURING AN ANGIOPLASTY PROCEDURE, THE DEVICE BALLOON ALLEGEDLY TWISTED AND FAILED TO INFLATE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10939659 ↗
  • MalfunctionOct 1, 2020Inflation Problem; Material Twisted/Bent; Inflation Problem; Material Twisted/Bent

    IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE THROUGH SUPERFICIAL FEMORAL ARTERY, THE DRUG-COATED BALLOON WAS ALLEGEDLY FAILED TO INFLATE AND TWISTED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10612705 ↗