ARROW IPN000082Recalls & reported problems
Teleflex Incorporated · Berman Angiographic Balloon Catheter · Class 2
All models of ARROW (1494)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801902002266), received between Mar 7, 2019 and Mar 25, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K892530); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation3
- Material Rupture3
- Inflation Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 25, 2025Material Deformation; Material Deformation
IT WAS RPEORTED "BALLOON IS TILTED TO ONE SIDE AND UNABLE TO DETECT PRESSURE." THE DEVICE WAS REMOVED AND NOT REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". PLEASE SEE ASSOCAITED MDR#: 3010532612-2025-00234.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21687518 ↗MalfunctionMar 25, 2025Material Deformation; Material Deformation
IT WAS RPEORTED "BALLOON IS TILTED TO ONE SIDE AND UNABLE TO DETECT PRESSURE." THE DEVICE WAS REMOVED AND NOT REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21687476 ↗MalfunctionOct 3, 2023Inflation Problem; Inflation Problem
IT WAS REPORTED THAT "THE BALLOON DID NOT INFLATE DURING USE ON A PATIENT. THEREFORE, THE CATHETER WAS REMOVED AND REPLACED WITH ANOTHER KIT, INSERTED AT THE SAME SITE AND COMPLETED THE PROCEDURE. NO HARM TO THE PATIENT WAS REPORTED". THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17863552 ↗MalfunctionOct 19, 2022Inflation Problem; Material Deformation; Inflation Problem; Material Deformation
IT WAS REPORTED THAT THE "BALLOON INFLATED ASYMMETRICALLY". THEREFORE, ANOTHER KIT WAS USED INSTEAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED TO CUSTOMER AS NO PATIENT DETAILS WAS PROVIDED AT TIME OF REPORT.
Patient problems: Insufficient Information
MAUDE report 15631121 ↗MalfunctionMar 10, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT "THE BALLOON BURSTS DURING USAGE ON A PATIENT. THE 4-FRENCH CATHETER WAS INFLATED AND DEFLATED SEVERAL TIMES DURING THE APPLICATION (ANGIOGRAPHY); THE CUSTOMER DESCRIBED IT AS "PERMANENT USE"." IT WAS REPORTED THAT THERE WAS NO HARM OR INJURY TO THE PATIENT AND THE PATIENT CURRENT CONDITION IS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9813585 ↗MalfunctionApr 15, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE EVENT OCCURRED PRIOR TO USE ON A PATIENT (PRE-TESTING). IT WAS REPORTED THAT THE BALLOON RUPTURED OR PERFORATED BEFORE TESTING. AS A RESULT, A NEW BERMAN ANGIOGRAPHIC CATHETER WAS USED. THERE WAS NO REPORT OF PATIENT INJURY OR CONSEQUENCE. ADDITIONAL INFORMATION. IT WAS REPORTED THAT THE RUPTURED OCCURRED PRIOR TO USE. THE SAMPLE WAS RECEIVED FOR INVESTIGATION. A PRELIMINARY INVESTIGATION WAS DONE, DRIED BLOOD WAS NOTED WITHIN THE INJECTION EXTENSION LINE OF THE RETURNED SAMPLE. THE COMPLAINT WAS REASSESSED AND DEEMED REPORTABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8514706 ↗MalfunctionMar 7, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE CATHETER WAS INJECTED AT 600PSI AND THE BALLOON LUMEN SYRINGE BACKED UP AND THEY HEARD A POP (ASSUMED IT WAS BALLOON WHICH SHOULD NOT BE AFFECTED WHEN INJECTING DYE) SO THE DEVICE WAS IMMEDIATELY REMOVED. THERE WAS NO REPORT OF PATIENT COMPLICATION OR SERIOUS INJURY AND DEATH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8401241 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K892530. One submission often covers many models, so check the product and lot details in each record.
- Z-0716-2020 · Nov 15, 2019 · TerminatedArrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J
- Z-0715-2020 · Nov 15, 2019 · TerminatedArrow Berman Angiographic Catheter 50cc Item Number: AI-07130
- Z-1057-2013 · Mar 4, 2013 · TerminatedArrow/Teleflex Medical Berman Angiographic Balloon Catheter and Reverse Berman Angiographic Balloon Catheter. The Arrow Berman(TM) Angiographic Catheter is indicated for use in ri
Other CATHETER, FLOW DIRECTED devices
- Graftmaster 1012580-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED30 deaths225 reports
- SWAN-GANZ CCOMBO V 777F8Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls92 reports
- Graftmaster 1012581-16Abbott Vascular Inc. · CATHETER, FLOW DIRECTED21 deaths147 reports
- SWAN-GANZ CCOMBO V 774F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls2 deaths41 reports
- Graftmaster 1012581-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED12 deaths135 reports
- Graftmaster 1012582-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths77 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall23 reports
- Graftmaster 1012580-26Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths72 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page