ARROW IPN000082Recalls & reported problems

Teleflex Incorporated · Berman Angiographic Balloon Catheter · Class 2

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00801902002266), received between Mar 7, 2019 and Mar 25, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

3 other recalls cover devices cleared under the same FDA submission (K892530); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation3
  • Material Rupture3
  • Inflation Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient3
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2025Material Deformation; Material Deformation

    IT WAS RPEORTED "BALLOON IS TILTED TO ONE SIDE AND UNABLE TO DETECT PRESSURE." THE DEVICE WAS REMOVED AND NOT REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". PLEASE SEE ASSOCAITED MDR#: 3010532612-2025-00234.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21687518 ↗
  • MalfunctionMar 25, 2025Material Deformation; Material Deformation

    IT WAS RPEORTED "BALLOON IS TILTED TO ONE SIDE AND UNABLE TO DETECT PRESSURE." THE DEVICE WAS REMOVED AND NOT REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21687476 ↗
  • MalfunctionOct 3, 2023Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT "THE BALLOON DID NOT INFLATE DURING USE ON A PATIENT. THEREFORE, THE CATHETER WAS REMOVED AND REPLACED WITH ANOTHER KIT, INSERTED AT THE SAME SITE AND COMPLETED THE PROCEDURE. NO HARM TO THE PATIENT WAS REPORTED". THE PATIENT STATUS IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17863552 ↗
  • MalfunctionOct 19, 2022Inflation Problem; Material Deformation; Inflation Problem; Material Deformation

    IT WAS REPORTED THAT THE "BALLOON INFLATED ASYMMETRICALLY". THEREFORE, ANOTHER KIT WAS USED INSTEAD. ADDITIONAL INFORMATION HAS BEEN REQUESTED TO CUSTOMER AS NO PATIENT DETAILS WAS PROVIDED AT TIME OF REPORT.

    Patient problems: Insufficient Information

    MAUDE report 15631121 ↗
  • MalfunctionMar 10, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT "THE BALLOON BURSTS DURING USAGE ON A PATIENT. THE 4-FRENCH CATHETER WAS INFLATED AND DEFLATED SEVERAL TIMES DURING THE APPLICATION (ANGIOGRAPHY); THE CUSTOMER DESCRIBED IT AS "PERMANENT USE"." IT WAS REPORTED THAT THERE WAS NO HARM OR INJURY TO THE PATIENT AND THE PATIENT CURRENT CONDITION IS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9813585 ↗
  • MalfunctionApr 15, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE EVENT OCCURRED PRIOR TO USE ON A PATIENT (PRE-TESTING). IT WAS REPORTED THAT THE BALLOON RUPTURED OR PERFORATED BEFORE TESTING. AS A RESULT, A NEW BERMAN ANGIOGRAPHIC CATHETER WAS USED. THERE WAS NO REPORT OF PATIENT INJURY OR CONSEQUENCE. ADDITIONAL INFORMATION. IT WAS REPORTED THAT THE RUPTURED OCCURRED PRIOR TO USE. THE SAMPLE WAS RECEIVED FOR INVESTIGATION. A PRELIMINARY INVESTIGATION WAS DONE, DRIED BLOOD WAS NOTED WITHIN THE INJECTION EXTENSION LINE OF THE RETURNED SAMPLE. THE COMPLAINT WAS REASSESSED AND DEEMED REPORTABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8514706 ↗
  • MalfunctionMar 7, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE CATHETER WAS INJECTED AT 600PSI AND THE BALLOON LUMEN SYRINGE BACKED UP AND THEY HEARD A POP (ASSUMED IT WAS BALLOON WHICH SHOULD NOT BE AFFECTED WHEN INJECTING DYE) SO THE DEVICE WAS IMMEDIATELY REMOVED. THERE WAS NO REPORT OF PATIENT COMPLICATION OR SERIOUS INJURY AND DEATH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8401241 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K892530. One submission often covers many models, so check the product and lot details in each record.

  • Z-0716-2020 · Nov 15, 2019 · Terminated
    Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J
  • Z-0715-2020 · Nov 15, 2019 · Terminated
    Arrow Berman Angiographic Catheter 50cc Item Number: AI-07130
  • Z-1057-2013 · Mar 4, 2013 · Terminated
    Arrow/Teleflex Medical Berman Angiographic Balloon Catheter and Reverse Berman Angiographic Balloon Catheter. The Arrow Berman(TM) Angiographic Catheter is indicated for use in ri

Other CATHETER, FLOW DIRECTED devices

All CATHETER, FLOW DIRECTED devices →