ARROW IPN032917Recalls & reported problems
Teleflex Incorporated · Arrowg+ard Blue(R) Three-Lumen CVC · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801902007858), received between Aug 17, 2023 and Jan 26, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).
12 other recalls cover devices cleared under the same FDA submission (K900263); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent3
- Biocompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Adult Respiratory Distress Syndrome1
- Anaphylactic Shock1
- Cardiac Arrest1
- Erythema1
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "GUIDEWIRE KINKING WHILE TRYING TO THREAD, UNABLE TO PLACE CENTRAL LINE." GUIDEWIRE WAS REMOVED AND A NEW KIT WAS USED. NO HARM OR HEALTH CONSEQUENCE. PATIENT'S CURRENT CONDITION IS "FINE" AT THE TIME OF THIS REPORT. ASSOCIATED MDR NUMBERS 9680794-2024-00065 AND 9680794-2024-00066.
Patient problems: Insufficient Information
MAUDE report 18591831 ↗MalfunctionJan 26, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "GUIDEWIRE KINKING WHILE TRYING TO THREAD, UNABLE TO PLACE CENTRAL LINE." THE DEVICE WAS REMOVED AND A NEW KIT WAS USED. NO HARM OR HEALTH CONSEQUENCE. PATIENT'S CURRENT CONDITION IS "FINE" AT THE TIME OF THIS REPORT. ASSOCIATED MDR NUMBER 9680794-2024-00066 AND 9680794-2024-00067.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18591426 ↗MalfunctionJan 26, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "GUIDEWIRE KINKING WHILE TRYING TO THREAD, UNABLE TO PLACE CENTRAL LINE." GUIDEWIRE WAS REMOVED AND A NEW KIT WAS USED. NO PATIENT HARM OR INJURY. PATIENT'S CURRENT CONDITION IS "FINE" AT THE TIME OF THIS REPORT. ASSOCIATED MDR NUMBER 9680794-2024-00065 AND 9680794-2024-00067.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18591418 ↗InjuryAug 17, 2023Biocompatibility; Biocompatibility
IT WAS REPORTED THE PATIENT HAD AN ALLERGIC REACTION TO THE CATHETER. THE PATIENT HAD A KNOWN ALLERGY TO CHLORHEXIDINE WITH PREVIOUS ANAPHYLAXIS TO SKIN PREP FOR PERIPHERAL IV 4 MONTHS EARLIER. THE PATIENT EXPERIENCED ERYTHEMA AT THE INSERTION SITE, RESPIRATORY DISTRESS WITH AGONAL RESPIRATION, NO PULSE, AND PULSELESS ELECTRICAL ACTIVITY. THE PATIENT WAS TREATED WITH CPR FOR 45 MINUTES, EPINEPHRINE IV, INTUBATION, CARDIOVERSION, AND SOLUMEDROL. THE PATIENT WAS STABILIZED AND DISCHARGED HOME FROM THE ICU AFTER 5 DAYS. IT WAS REPORTED THE PATIENT EXPERIENCED SEVERE PSYCHOLOGICAL DISTRESS REQUIRING ON-GOING INTERVENTION.
Patient problems: Adult Respiratory Distress Syndrome; Anaphylactic Shock; Cardiac Arrest; Erythema
MAUDE report 17562993 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K900263. One submission often covers many models, so check the product and lot details in each record.
- Z-2376-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
- Z-0053-2020 · May 17, 2019 · TerminatedPediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, REF AK-25502 Product Usage; Provide short-term (< 30 days) central venous acces
- Z-2886-2018 · Jul 3, 2018 · TerminatedPediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous acc
- Z-1985-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue(R) Central Venous Catheterization Kit; Product Code: ASK-24306-WHC1 and ASK-24306-WHC2
- Z-1984-2018 · Apr 11, 2018 · TerminatedSingle-Lumen Central Venous Catheterization Kit with Blue FlexTip(R) ARROWg+ard Blue(R) Catheter and Sharps Safety Features; Product Code: ASK-24306-VC
- Z-1333-2018 · Dec 22, 2017 · TerminatedARROWg+ard Blue CVC Kit (1) Part Number (PN): CDC-24301-1A, (2) PN: CDC-24306-1A; (3) ARROWg+ard Blue Large-Bore Multi-Lumen CVC Kit, PN: CDC-22123-1A; (4) Central Venous Catheter
- Z-1001-2018 · Nov 27, 2017 · TerminatedARROWg+ard Blue Large-Bore Multi-Lumen CVC Kit Product Usage: The Arrow CVC is indicated to provide short-term (< 30 days) central venous access for treatment of diseases or condi
- Z-2337-2017 · May 9, 2017 · TerminatedARROWg+ard Blue CVC Kit The ARROWg+ard Blue CVCs are intended to permit venous access to the central circulation
- Z-1830-2012 · May 3, 2012 · TerminatedArrow Two-Lumen Central Venous Catheterization Set with Blue FlexTip(R) ARROWgard Blue(R) Catheter Product Usage: The multiple-lumen catheter permits venous access to central circ
- Z-2496-2010 · Jul 29, 2010 · TerminatedTwo-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter
- Z-1499-2009 · Feb 11, 2009 · TerminatedArrow ARROWg+ard CVCs (Central Venous Catheters)
- Z-0082-04 · Sep 16, 2003 · TerminatedArrow Howes multi-lumen central venous catheterization kit 7 Fr x 20 cm
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page