ARROW IPN000261Recalls & reported problems
Teleflex Incorporated · Intra-Aortic Balloon: 9 Fr. 50 cc · Class 2
All models of ARROW (1494)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801902010803), received between Aug 3, 2018 and Oct 14, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (1 report).
13 other recalls cover devices cleared under the same FDA submission (K010330); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Alarm System1
- Device Dislodged or Dislocated1
- Fluid/Blood Leak1
- Image Display Error/Artifact1
- Material Twisted/Bent1
- Signal Artifact/Noise1
- Unintended Movement1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 14, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED "POSSIBLE IAB RUPTURE". ADDITIONAL INFORMATION STATES "POSSIBLE HELIUM LOSS 3(2) ALARMS X 7 WITHIN A FEW MINUTES. NO BLOOD VISIBLE IN TUBING, CONNECTIONS TIGHT. ALARM PERSISTED AFTER REPOSITIONING THE PATIENT". THE IAB CATHETER WAS REMOVED AND NOT REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20440391 ↗MalfunctionOct 27, 2021Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT WHEN THE STAFF ADVANCED THE SHEATH, IT BEGAN TO "FOLD UP ON ITSELF." THE STAFF REMOVED THAT SHEATH AND PLACED A 9 FR SHEATH. WHEN THEY TRIED TO ADVANCE THE BALLOON THROUGH THE 9 FR SHEATH, IT WOULD NOT ADVANCE DUE TO THE BALLOON "BUNCHING UP." AS A RESULT, A SECOND IAB WAS USED SUCCESSFULLY. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY, OR DEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12705815 ↗MalfunctionOct 27, 2021Unraveled Material; Unraveled Material
IT WAS REPORTED, THAT WHEN THE STAFF ADVANCED THE SHEATH, IT BEGAN TO "FOLD UP ON ITSELF". THE STAFF REMOVED THAT SHEATH. AND PLACED A 9 FR SHEATH. WHEN THEY TRIED TO ADVANCE THE BALLOON THROUGH THE 9 FR SHEATH, IT WOULD NOT ADVANCE, DUE TO THE BALLOON "BUNCHING UP". AS A RESULT, A SECOND IAB WAS USED SUCCESSFULLY. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY, OR DEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12705725 ↗MalfunctionApr 30, 2021Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) WAS INSERTED INTO THE PATIENT IN THE CATH LAB FOR HEMODYNAMICALLY UNSTABLE STEMI, PREOP. OPEN HEART SURGERY (OHS) WAS PERFORMED ON THE PATIENT AND RETURNED TO OPERATING ROOM (OR). WHEN THE PATIENT WAS RETURNED TO THE UNIT AND THE IAB PLACEMENT WAS NOTED AS LOW ON CXR. THE DR. REPOSITIONED THE IAB AT BEDSIDE AND ALARMS OCCURRED AND BLOOD WAS NOTED. AS A RESULT, THE IAB WAS SUCCESSFULLY REMOVED AT BEDSIDE WITHOUT INCIDENT. THE PATIENT REMAINED STABLE POST REMOVAL, ANOTHER OTHER IAB WAS INSERTED, THERAPY WAS DISCONTINUED. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY OR DEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11757305 ↗MalfunctionFeb 4, 2020Signal Artifact/Noise; Image Display Error/Artifact; Signal Artifact/Noise; Image Display Error/Artifact
IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) CATHETER HAD TOO MUCH WHIP WHICH CREATED TOO MUCH ARTIFACT. AS A RESULT, THE USER SWITCHED TO OUR 40CC CATHETER AND IT WORKED FINE. THERE WAS NO REPORT OF PATIENT INJURY OR CONSEQUENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9664871 ↗MalfunctionAug 3, 2018Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement
IT WAS REPORTED THE DOCTOR NOTED THAT THE INTRA-AORTIC BALLOON (IAB) WAS "CORK SCREWING ON ITSELF" AND APPEAR TO BE "SPIRALING OR COILING INSIDE THE CATHETER SHAFT". THE MOVEMENT CAUSE THE CATHETER PLACEMENT TO "MOVE BACKWARDS 4-5 INCHES". THE DOCTOR REPOSITIONED THE CATHETER TWO OR THREE TIMES. AS A RESULT, THE DOCTOR REMOVED AND REPLACED THE CATHETER WITH ANOTHER IAB. THERE WAS NO REPORT OF PATIENT COMPLICATION OR SERIOUS INJURY AND DEATH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7747044 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K010330. One submission often covers many models, so check the product and lot details in each record.
- Z-1212-2009 · Feb 2, 2009 · Terminated8 Fr 30cc Ultra 8 IAB, IAB-05830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
- Z-1220-2009 · Feb 2, 2009 · Terminated9 FR 50CC Rediguard IAB, IAB-R950-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the
- Z-1211-2009 · Feb 2, 2009 · Terminated8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter
- Z-1213-2009 · Feb 2, 2009 · Terminated8 Fr 40cc Ultra 8 IAB, IAB-05840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
- Z-1222-2009 · Feb 2, 2009 · TerminatedARROW 50cc Drive Line Kits, REF IAK-02693, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation
- Z-1217-2009 · Feb 2, 2009 · Terminated8 Fr 40cc FIBEROPTIX IAB, IAB-05840-LWS, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in
- Z-1218-2009 · Feb 2, 2009 · Terminated7 FR 30CC Rediguard IAB, IAB-S730C, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
- Z-1215-2009 · Feb 2, 2009 · Terminated7.5 Fr 40cc UltraFlex IAB, IAB-06840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in t
- Z-1210-2009 · Feb 2, 2009 · Terminated8 Fr 30cc Narrow Flex IAB Catheter Kit with Generica Adapters Product Number: IAB-04830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counte
- Z-1221-2009 · Feb 2, 2009 · TerminatedARROW 40cc Drive Line Kits, REF IAK-02692, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation
- Z-1214-2009 · Feb 2, 2009 · Terminated7.5 Fr 30cc UltraFlex IAB, IAB-06830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in
- Z-1216-2009 · Feb 2, 2009 · Terminated8 Fr 30cc FIBEROPTIX IAB, IAB-05830-LWS, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in
- Z-1219-2009 · Feb 2, 2009 · Terminated8 FR 40CC Rediguard IAB, IAB-S840C, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
Other System, balloon, intra-aortic and control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page