Skip to content

ARROW IPN000261Recalls & reported problems

Teleflex Incorporated · Intra-Aortic Balloon: 9 Fr. 50 cc · Class 2

All models of ARROW (1494)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00801902010803), received between Aug 3, 2018 and Oct 14, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (1 report).

13 other recalls cover devices cleared under the same FDA submission (K010330); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Alarm System1
  • Device Dislodged or Dislocated1
  • Fluid/Blood Leak1
  • Image Display Error/Artifact1
  • Material Twisted/Bent1
  • Signal Artifact/Noise1
  • Unintended Movement1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 14, 2024Device Alarm System; Device Alarm System

    IT WAS REPORTED "POSSIBLE IAB RUPTURE". ADDITIONAL INFORMATION STATES "POSSIBLE HELIUM LOSS 3(2) ALARMS X 7 WITHIN A FEW MINUTES. NO BLOOD VISIBLE IN TUBING, CONNECTIONS TIGHT. ALARM PERSISTED AFTER REPOSITIONING THE PATIENT". THE IAB CATHETER WAS REMOVED AND NOT REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20440391 ↗
  • MalfunctionOct 27, 2021Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT WHEN THE STAFF ADVANCED THE SHEATH, IT BEGAN TO "FOLD UP ON ITSELF." THE STAFF REMOVED THAT SHEATH AND PLACED A 9 FR SHEATH. WHEN THEY TRIED TO ADVANCE THE BALLOON THROUGH THE 9 FR SHEATH, IT WOULD NOT ADVANCE DUE TO THE BALLOON "BUNCHING UP." AS A RESULT, A SECOND IAB WAS USED SUCCESSFULLY. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY, OR DEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12705815 ↗
  • MalfunctionOct 27, 2021Unraveled Material; Unraveled Material

    IT WAS REPORTED, THAT WHEN THE STAFF ADVANCED THE SHEATH, IT BEGAN TO "FOLD UP ON ITSELF". THE STAFF REMOVED THAT SHEATH. AND PLACED A 9 FR SHEATH. WHEN THEY TRIED TO ADVANCE THE BALLOON THROUGH THE 9 FR SHEATH, IT WOULD NOT ADVANCE, DUE TO THE BALLOON "BUNCHING UP". AS A RESULT, A SECOND IAB WAS USED SUCCESSFULLY. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY, OR DEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12705725 ↗
  • MalfunctionApr 30, 2021Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) WAS INSERTED INTO THE PATIENT IN THE CATH LAB FOR HEMODYNAMICALLY UNSTABLE STEMI, PREOP. OPEN HEART SURGERY (OHS) WAS PERFORMED ON THE PATIENT AND RETURNED TO OPERATING ROOM (OR). WHEN THE PATIENT WAS RETURNED TO THE UNIT AND THE IAB PLACEMENT WAS NOTED AS LOW ON CXR. THE DR. REPOSITIONED THE IAB AT BEDSIDE AND ALARMS OCCURRED AND BLOOD WAS NOTED. AS A RESULT, THE IAB WAS SUCCESSFULLY REMOVED AT BEDSIDE WITHOUT INCIDENT. THE PATIENT REMAINED STABLE POST REMOVAL, ANOTHER OTHER IAB WAS INSERTED, THERAPY WAS DISCONTINUED. THERE WAS NO REPORT OF PATIENT COMPLICATIONS, SERIOUS INJURY OR DEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11757305 ↗
  • MalfunctionFeb 4, 2020Signal Artifact/Noise; Image Display Error/Artifact; Signal Artifact/Noise; Image Display Error/Artifact

    IT WAS REPORTED THAT THE INTRA-AORTIC BALLOON (IAB) CATHETER HAD TOO MUCH WHIP WHICH CREATED TOO MUCH ARTIFACT. AS A RESULT, THE USER SWITCHED TO OUR 40CC CATHETER AND IT WORKED FINE. THERE WAS NO REPORT OF PATIENT INJURY OR CONSEQUENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9664871 ↗
  • MalfunctionAug 3, 2018Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement

    IT WAS REPORTED THE DOCTOR NOTED THAT THE INTRA-AORTIC BALLOON (IAB) WAS "CORK SCREWING ON ITSELF" AND APPEAR TO BE "SPIRALING OR COILING INSIDE THE CATHETER SHAFT". THE MOVEMENT CAUSE THE CATHETER PLACEMENT TO "MOVE BACKWARDS 4-5 INCHES". THE DOCTOR REPOSITIONED THE CATHETER TWO OR THREE TIMES. AS A RESULT, THE DOCTOR REMOVED AND REPLACED THE CATHETER WITH ANOTHER IAB. THERE WAS NO REPORT OF PATIENT COMPLICATION OR SERIOUS INJURY AND DEATH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7747044 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K010330. One submission often covers many models, so check the product and lot details in each record.

  • Z-1212-2009 · Feb 2, 2009 · Terminated
    8 Fr 30cc Ultra 8 IAB, IAB-05830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
  • Z-1220-2009 · Feb 2, 2009 · Terminated
    9 FR 50CC Rediguard IAB, IAB-R950-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the
  • Z-1211-2009 · Feb 2, 2009 · Terminated
    8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter
  • Z-1213-2009 · Feb 2, 2009 · Terminated
    8 Fr 40cc Ultra 8 IAB, IAB-05840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
  • Z-1222-2009 · Feb 2, 2009 · Terminated
    ARROW 50cc Drive Line Kits, REF IAK-02693, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation
  • Z-1217-2009 · Feb 2, 2009 · Terminated
    8 Fr 40cc FIBEROPTIX IAB, IAB-05840-LWS, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in
  • Z-1218-2009 · Feb 2, 2009 · Terminated
    7 FR 30CC Rediguard IAB, IAB-S730C, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a
  • Z-1215-2009 · Feb 2, 2009 · Terminated
    7.5 Fr 40cc UltraFlex IAB, IAB-06840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in t
  • Z-1210-2009 · Feb 2, 2009 · Terminated
    8 Fr 30cc Narrow Flex IAB Catheter Kit with Generica Adapters Product Number: IAB-04830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counte
  • Z-1221-2009 · Feb 2, 2009 · Terminated
    ARROW 40cc Drive Line Kits, REF IAK-02692, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation
  • Z-1214-2009 · Feb 2, 2009 · Terminated
    7.5 Fr 30cc UltraFlex IAB, IAB-06830-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in
  • Z-1216-2009 · Feb 2, 2009 · Terminated
    8 Fr 30cc FIBEROPTIX IAB, IAB-05830-LWS, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in
  • Z-1219-2009 · Feb 2, 2009 · Terminated
    8 FR 40CC Rediguard IAB, IAB-S840C, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter-pulsation therapy, whereby balloon inflation in the a

Other System, balloon, intra-aortic and control devices

All System, balloon, intra-aortic and control devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page