ARROW IPN000260Recalls & reported problems
Teleflex Incorporated · UltraFlex(TM) IAB · Class 2
Quick summary
This Teleflex Arrow UltraFlex intra-aortic balloon has 690 adverse event reports, with 4 deaths and 60 injuries. Common problems include the balloon failing to inflate, fluid or blood leaking, and material tearing or rupturing; some reports mention needle-stick injuries. No recalls name this device. Reports dropped sharply, with 32 in the last year compared with 236 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 698
- Deaths reported
- 4
- Injuries reported
- 60
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 698 adverse event reports that list this exact device (DI 00801902026804), received between Jul 19, 2018 and Sep 24, 2026 — classified as 4 deaths, 60 injuries, 634 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (138 reports).
27 other recalls cover devices cleared under the same FDA submission (K000729); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Inflation Problem138
- Fluid/Blood Leak114
- Leak/Splash69
- Patient Device Interaction Problem57
- Material Rupture40
- Material Integrity Problem39
- Material Twisted/Bent31
- Difficult to Insert27
Patient problems
- No Clinical Signs, Symptoms or Conditions489
- Insufficient Information125
- No Consequences Or Impact To Patient53
- Needle Stick/Puncture13
- Death9
- Calcium Deposits/Calcification3
- Hemorrhage/Blood Loss/Bleeding3
- Cardiac Arrest2
Recent adverse event reports
The 25 most recent of 698 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50889530 ↗MalfunctionSep 14, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50763512 ↗MalfunctionSep 10, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50719921 ↗MalfunctionSep 9, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50697402 ↗MalfunctionSep 2, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50565864 ↗MalfunctionSep 1, 2026Entrapment of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50548366 ↗InjuryAug 31, 2026Material Too Rigid or Stiff
Patient problems: Vascular Dissection
MAUDE report 50531314 ↗MalfunctionAug 31, 2026Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50531098 ↗MalfunctionAug 31, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50530974 ↗InjuryAug 31, 2026Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50528859 ↗MalfunctionAug 28, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50500911 ↗MalfunctionAug 28, 2026Material Twisted/Bent
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50499451 ↗MalfunctionAug 28, 2026Failure to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50498547 ↗MalfunctionAug 28, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50496367 ↗MalfunctionAug 26, 2026Failure to Advance
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50452949 ↗MalfunctionAug 25, 2026Difficult to Insert
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50437707 ↗MalfunctionAug 19, 2026Obstruction of Flow
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50327530 ↗MalfunctionNov 25, 2025Break; Break
IT WAS REPORTED "IABC FOUND TO BE RUPTURE AFTER INSERTION WITHIN 30 MIN". ADDITIONAL INFORMATION STATES THAT THE IAB CATHER WAS REMOVED AND REPALCED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". AT THE TIME OF THIS REPORT, THE CUSTOMER HAS NOT RETURNED OUR REQUESTS FOR ADDITIONAL INFORMATION. IF FURTHER INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE UPDATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23645403 ↗MalfunctionNov 25, 2025Device Alarm System; Device Alarm System
IT WAS REPORTED "AFTER THE CATHETER INTO THE PATIENT, THE PUMP ALARMED HELIUM LOSS ALARM. AFTER THE BALLOON WAS WITHDRAWN, IT WAS FOUND THAT THE FRONT END OF THE BALLOON WAS FOLDING". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REPLACED INTO THE SAME INSERTION SITE. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644873 ↗MalfunctionNov 25, 2025Increase in Pressure; Increase in Pressure
IT WAS REPORTED " THE CATHETER INSERTED INTO THE PATIENT SUCCESSFUL. AFTER ABOUT 3 MINUTES THE PUMP ALARMED THE BALLOON PRESSURE HIGH, AND THE COUNTERPULSATION WAS SUSPENDED. LATER, IT WAS DISCOVERED THAT THERE WAS BLOOD LEAKAGE, SO THE CATHETER WAS REMOVED. CHANGE THE NEW CATHETER". NO PATIENT'S INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644437 ↗MalfunctionNov 25, 2025Device Alarm System; Device Alarm System
IT WAS REPORTED "AFTER THE CATHETER INTO THE PATIENT, THE PUMP ALARMED HELIUM LOSS 2. AFTER THE CATHETER WAS REMOVED, IT WAS FOUND THAT THE FRONT END OF THE BALLOON WAS BENT". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REPLACED INTO THE SAME INSERTION SITE. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644330 ↗MalfunctionNov 25, 2025Entrapment of Device; Entrapment of Device
IT WAS REPORTED "THE PEEL-AWAY HEMOSTASIS DEVICE COULD NOT BE REMOVED DURING USE. A NEW CATHETER WAS REPLACED AND REINSERTED". ADDITIONAL INFORMATION STATES THAT THE SECOND IAB CATHETER WAS INSERTED INTO THE SAME INSERTION SITE. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644232 ↗MalfunctionNov 24, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED "DURING THE PROCESS OF IMPLANTING THE CATHETER INTO THE PATIENT, THE FRONT END OF THE BALLOON WAS BENT AND COULD NOT BE IMPLANTED INSIDE THE PATIENT'S BODY. AFTER REMOVAL, IT WAS FOUND THAT THE FRONT END OF THE BALLOON WAS BENT". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23633488 ↗MalfunctionNov 24, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED "AFTER THE CATHETER INSERTED INTO THE PATIENT, THE BALLOON HAD NOT INFLATE COMPLETELY. THE BALLOON WAS FOUND RUPTURED AFTER REMOVE THE CATHETER. CHANGE THE NEW CATHETER". NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENTS CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23633171 ↗MalfunctionNov 24, 2025Inflation Problem; Inflation Problem
IT WAS REPORTED "BALLOON NOT INFLATING IN BODY". ADDITIONAL INFORMATION STATES THAT THE IAB CATHETER WAS REMOVED AND REPLACED. NO PATIENT INJURY OR CONSEQUENCE REPORTED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".
Patient problems: Insufficient Information
MAUDE report 23632890 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K000729. One submission often covers many models, so check the product and lot details in each record.
- Z-1879-2024 · Apr 29, 2024 · Open, ClassifiedArrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-U
- Z-1880-2024 · Apr 29, 2024 · Open, ClassifiedArrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U
- Z-1059-2016 · Feb 10, 2016 · TerminatedUltraFlex IAB: 7.5Fr 30cc; Product Code: IAB-06830-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during di
- Z-1060-2016 · Feb 10, 2016 · TerminatedUltraFlex IAB: 7.5Fr 40cc; Product Code: IAB-06840-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during di
- Z-1056-2016 · Feb 10, 2016 · TerminatedUltra 8 IAB: 8Fr 30cc; Product Code: IAB-05830-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diasto
- Z-1058-2016 · Feb 10, 2016 · TerminatedUltra 8 IAB: 8Fr 40cc; Product Code: IAB-05840-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diasto
- Z-1062-2016 · Feb 10, 2016 · TerminatedPercutaneous Insertion Tray; Product Code: IAK-06845 The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during di
- Z-1065-2013 · Nov 27, 2012 · TerminatedArrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 40 cc; Product Number: IAB-05840-LWS. The IAB is utilized for intra-aortic balloon counter-pulsation th
- Z-1064-2013 · Nov 27, 2012 · TerminatedArrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 30 cc; Product Number: IAB-05830-LWS. The IAB is utilized for intra-aortic balloon counter-pulsation t
- Z-0793-2011 · Oct 11, 2010 · TerminatedFiberOptix IAB Ultra 8 Fr, 40cc Model: IAB-05840-LWS
- Z-0969-2011 · Oct 11, 2010 · TerminatedPercutaneous Insertion Tray For Use with: Arrow Ultra 8 IAB (Intra Aortic Balloon) Catheters Model: IAK-05845
- Z-0790-2011 · Oct 11, 2010 · TerminatedArrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr, 30cc. Model: IAB-05830-U
- Z-0791-2011 · Oct 11, 2010 · TerminatedArrow Ultra 8 IAB (Intra-Aortic Balloon) Catheter, 8 Fr, 40cc. Model: IAB-05840-U
- Z-0792-2011 · Oct 11, 2010 · TerminatedFiberOptix IAB Sheath, Ultra 8 Fr, 30cc. Model: IAB-05830-LWS
- Z-1211-2009 · Feb 2, 2009 · Terminated8 Fr 40cc Narrow Flex IAB Catheter Kit with Generic Adapters Product Number: IAB-04840-U, Intra-aortic balloon catheter. The Intra-aortic balloon (IAB) is utilized for IAB counter
…and 12 more.
Other System, balloon, intra-aortic and control devices
- CARDIOSAVE HYBRID, TYPE B PLUG 0998-00-0800-53Datascope Corp. · System, Balloon, Intra-Aortic And Control21 recalls39 deaths10,316 reports
- CS300 Intra-Aortic Balloon Pump, English, 110V 0998-00-3023-53Datascope Corp. · System, Balloon, Intra-Aortic And Control12 deaths3,164 reports
- CARDIOSAVE HYBRID W/ E/F PLUG 0998-00-0800-55Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls8 deaths1,686 reports
- CARDIOSAVE HYBRID, TYPE I PLUG 0998-00-0800-45Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls5 deaths1,181 reports
- CARDIOSAVE HYBRID, TYPE G PLUG 0998-00-0800-52Datascope Corp. · System, Balloon, Intra-Aortic And Control19 recalls2 deaths1,157 reports
- SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet) 0684-00-0576-01Datascope Corp. · System, Balloon, Intra-Aortic And Control2 recalls33 deaths1,541 reports
- CARDIOSAVE RESCUE 0998-00-0800-83Datascope Corp. · System, Balloon, Intra-Aortic And Control26 recalls167 reports
- CARDIOSAVE HYBRID, 3.1 EDITION 0998-00-0800-65Datascope Corp. · System, Balloon, Intra-Aortic And Control20 recalls3 deaths456 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.