ARROW IPN033026Recalls & reported problems

Teleflex Incorporated · NextStep(R) Retograde Hemodialysis Catheter Long-Term Access. Catheter length, Cuff to Tip: 19 cm · Class 2 · Implant

All models of ARROW (1494)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00801902096289), received between Aug 29, 2023 and Oct 28, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K111117); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Connection Problem1
  • Incomplete or Inadequate Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT "AFTER COMPLETION OF PROCEDURE, ON FLUSHING AND ASPIRATION OF LUMENS, LEAKAGE EXPERIENCED AT HUB ASSEMBLY, DESPITE GREEN COMPRESSION SLEEVE IN PLACE. DR REMOVED GREEN COMPRESSION SLEEVE AND REPLACE WITH THE SPARE ONE IN PACK - STILL LEAKING ON FLUSHING AND ASPIRATION AT HUB ASSEMBLY. DR REQUESTED TO OPEN A CS-15232-X 23CM SET TO ACCESS THE HUB ASSEMBLY, AS (B)(6) HOSPITAL DID NOT STOCK (B)(6) HUB REPAIR KIT. AFTER USING NEW HUB ASSEMBLY FROM NEW PACK, THE CATHETER FUNCTIONED PERFECTLY WELL ON FLUSHING AND ASPIRATION OF CATHETER LUMENS, NO LEAKAGE." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: Insufficient Information

    MAUDE report 23401778 ↗
  • MalfunctionAug 21, 2024Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    IT WAS REPORTED "DOCTOR STATES THAT THE HUB IN THE SET WOULD NOT SCREW ONTO THE CATHETER ONCE TUNNELLING WAS COMPLETED AND THEY WERE ATTACHING THE HUBS. COLLAR WAS USED." THE DEVICE WAS NOT REMOVED OR REPLACED THE CATHETER WAS LEFT IN SITU. ANOTHER HUBSET WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO MEDICAL INTERVENTION NEEDED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20011874 ↗
  • MalfunctionFeb 9, 2024Connection Problem; Connection Problem

    IT WAS REPORTED "THE CATHETER WAS INSERTED AND WHEN DOCTOR TRIED TO CONNECT THE CATHETER USING THE SUPPLIED CONNECTOR , IT DID NOT CONNECT." ANOTHER CONNECTOR WAS USED FROM A NEW PACK AND THE SECOND ATTEMPT WAS SUCCESFUL. NO MEDICAL INTERVENTION. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18680016 ↗
  • MalfunctionAug 29, 2023Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THE EXTENSION LINE WAS FOUND LEAKING DURING USE ON THE PATIENT. THE PATIENT'S CONDITION IS REPORTED AS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17645364 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111117. One submission often covers many models, so check the product and lot details in each record.

  • Z-2159-2026 · Apr 10, 2026 · Open, Classified
    NextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep A
  • Z-2158-2026 · Apr 10, 2026 · Open, Classified
    Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter w
  • Z-2156-2026 · Apr 10, 2026 · Open, Classified
    Arrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I
  • Z-1007-2015 · Feb 4, 2013 · Terminated
    The Arrow Edge Hemodialysis Catheterization Set Long-Term Access and NextStep Antegrade Hemodialysis Catheterization Set, Long-Term Access, part numbers: AC-15192-SFX, AC-15232-SFX

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →