ARROW IPN033034Recalls & reported problems
Teleflex Incorporated · Cannon(R) II Plus Long-Term Access. Catheter length, Cuff to Tip: 19 cm · Class 2 · Implant
Quick summary
This Arrow Cannon II Plus catheter is a long-term implanted catheter used for hemodialysis. The 41 reports mostly describe cracking at the luer hub or connector with blood leaking or oozing, with no deaths and no injuries mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 00801902096326), received between May 13, 2024 and Jun 23, 2026 — classified as 41 malfunctions.
The most frequently reported specific device problem is “Crack” (36 reports).
4 other recalls cover devices cleared under the same FDA submission (K111117); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Crack36
- Fluid/Blood Leak9
- Material Deformation6
- Material Separation2
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions38
- Hemorrhage/Blood Loss/Bleeding2
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2026, THAT AN INDWELLING CANNON II PLUS CHRONIC HEMODIALYSIS CATHETER (CHDC), ORIGINALLY INSERTED ON (B)(6) 2026, EXHIBITED CRACKING AT THE LUER HUB/FITTING ACCOMPANIED BY BLOOD OOZING FROM THE AFFECTED AREA. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS "FINE." THE AFFECTED CATHETER WAS REMOVED AND REPLACED WITH ANOTHER DEVICE. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25593864 ↗MalfunctionJun 23, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2026, THAT AN INDWELLING CANNON II PLUS CHRONIC HEMODIALYSIS CATHETER (CHDC), ORIGINALLY INSERTED ON (B)(6) 2026, EXHIBITED CRACKING AT THE LUER HUB/FITTING ACCOMPANIED BY BLOOD OOZING FROM THE AFFECTED AREA. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS "FINE." THE AFFECTED CATHETER WAS REMOVED AND REPLACED WITH ANOTHER DEVICE. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25593791 ↗MalfunctionJun 23, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2026, THAT AN INDWELLING CANNON II PLUS CHRONIC HEMODIALYSIS CATHETER (CHDC), ORIGINALLY INSERTED ON (B)(6) 2026, EXHIBITED CRACKING AT THE LUER HUB/FITTING ACCOMPANIED BY BLOOD OOZING FROM THE AFFECTED AREA. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED AS "FINE." THE AFFECTED CATHETER WAS REMOVED AND REPLACED WITH ANOTHER DEVICE. NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25593705 ↗MalfunctionMay 11, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT ON (B)(6) 2026, DURING USE ON A PATIENT, THE LUER HUB/FITTING OF A CHRONIC HEMODIALYSIS CATHETER (CHDC) WAS FOUND TO BE CRACKED AND LEAKING. THE CATHETER HAD BEEN INSERTED ON (B)(6) 2026. THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THE EVENT. THE PATIENT'S CONDITION WAS REPORTED AS FINE. NO MEDICAL INTERVENTION WAS NECESSARY, AND THE PROCEDURE WAS COMPLETED AFTER CHANGING TO ANOTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25137503 ↗MalfunctionMay 4, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT "ON 09APR2026 DURING PATIENT USE, THE LUER HUB/FITTING WAS FOUND CRACKED AND LEAKING." THERE WERE NO REPORTS OF PATIENT INJURY, HARM, OR ADVERSE HEALTH CONSEQUENCES ASSOCIATED WITH THE EVENT. THE PATIENT'S CONDITION WAS REPORTED AS FINE. NO MEDICAL INTERVENTION WAS NECESSARY, AND THE PROCEDURE WAS COMPLETED AFTER SWITCHING TO ANOTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25071741 ↗MalfunctionApr 20, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT: "ON (B)(6) 2026, LEUR HUB WAS FOUND TO HAVE A CRACK AND BLEEDING DURING PRE MACHINE TESTING OF THE CATHETER. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24925101 ↗MalfunctionApr 17, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT: "ON (B)(6) 2026, LEUR HUB WAS FOUND TO HAVE A CRACK AND BLEEDING DURING PRE-MACHINE TESTING OF THE CATHETER. THERE WAS NO REPORTED PATIENT HARM."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24908201 ↗MalfunctionApr 10, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT: "(B)(6) 2026, BEFORE THE PATIENT UNDERWENT DIALYSIS ON THE MACHINE, IT WAS FOUND THAT THERE WERE CRACKS AND BLEEDING ON THE LUER HUB. NO HARM FOR THE PATIENT. THE PATIENT CONDITION WAS REPORTED AS FINE. NO MEDICAL INTERVENTION WAS REQUIRED. THEY REPLACED THE DEVICE TO RESOLVE THE ISSUE."
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24842951 ↗MalfunctionApr 10, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT: "16 MAR 2026, BEFORE THE PATIENT UNDERWENT DIALYSIS ON THE MACHINE, IT WAS FOUND THAT THERE WERE CRACKS AND BLEEDING ON THE LUER HUB. NO HARM FOR THE PATIENT. THE PATIENT CONDITION WAS REPORTED AS FINE. NO MEDICAL INTERVENTION WAS REQUIRED. THEY REPLACED THE DEVICE TO RESOLVE THE ISSUE."
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 24838382 ↗MalfunctionDec 1, 2025Crack; Crack
IT WAS REPORTED THAT"ON (B)(6) 2025, ACCORDING TO THE CUSTOMER'S FEEDBACK, WHEN THE SYRINGE WAS ROUTINELY USED TO TEST THE TIGHTNESS OF THE DIALYSIS TUBE BEFORE THE PATIENT WAS PUT ON THE MACHINE FOR DIALYSIS, THE LEAK WAS FOUND. AFTER CAREFUL INSPECTION OF THE JOINT, IT WAS FOUND THAT THE LUER HUB/FITTING HAS CRACK. THIS WAS USED ABOUT 5 MONTHS. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO MEDICAL DEVICE INTERVENTION NEEDED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23678728 ↗MalfunctionDec 1, 2025Crack; Crack
IT WAS REPORTED THAT "ON (B)(6) 2025, ACCORDING TO THE CUSTOMER'S FEEDBACK, WHEN THE SYRINGE WAS ROUTINELY USED TO TEST THE TIGHTNESS OF THE DIALYSIS TUBE BEFORE THE PATIENT WAS PUT ON THE MACHINE FOR DIALYSIS, THE LEAK WAS FOUND. AFTER CAREFUL INSPECTION OF THE JOINT, IT WAS FOUND THAT THE LUER HUB/FITTING HAS CRACK. THIS WAS USED ABOUT 5 MONTHS. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO MEDICAL DEVICE INTERVENTION NEEDED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23678707 ↗MalfunctionDec 1, 2025Crack; Crack
IT WAS REPORTED THAT: "ON (B)(6) 2025, ACCORDING TO THE CUSTOMER'S FEEDBACK, WHEN THE SYRINGE WAS ROUTINELY USED TO TEST THE TIGHTNESS OF THE DIALYSIS TUBE BEFORE THE PATIENT WAS PUT ON THE MACHINE FOR DIALYSIS, THE LEAK WAS FOUND. AFTER CAREFUL INSPECTION OF THE JOINT, IT WAS FOUND THAT THE LUER HUB/FITTING HAS CRACK. INVOLVED 2 PC." THIS WAS USED ABOUT 5.5 MONTHS. "
Patient problems: Insufficient Information
MAUDE report 23678641 ↗MalfunctionDec 1, 2025Crack; Crack
IT WAS REPORTED THAT: "ON (B)(6) 2025, ACCORDING TO THE CUSTOMER'S FEEDBACK, WHEN THE SYRINGE WAS ROUTINELY USED TO TEST THE TIGHTNESS OF THE DIALYSIS TUBE BEFORE THE PATIENT WAS PUT ON THE MACHINE FOR DIALYSIS, THE LEAK WAS FOUND. AFTER CAREFUL INSPECTION OF THE JOINT, IT WAS FOUND THAT THE LUER HUB/FITTING HAS CRACK. INVOLVED 2 PC." THIS WAS USED ABOUT 5.5 MONTHS. "
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23678616 ↗MalfunctionMay 8, 2025Crack; Crack
IT WAS REPORTED "ON (B)(6) 2025, IN THE COURSE OF DIALYSING A PATIENT, THE DOCTOR FOUND LUER HUB/FITTING HAS CRACK." THE USER CHANGED TO A NEW SET TO RESOLVE THE ISSUE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2025-00440, 9680794-2025-00441, AND 9680794-2025-00442.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21985410 ↗MalfunctionMay 8, 2025Crack; Crack
IT WAS REPORTED "(B)(6) 2025, IN THE COURSE OF DIALYSING A PATIENT, THE DOCTOR FOUND LUER HUB/FITTING HAS CRACK." THE USER CHANGED TO A NEW SET TO RESOLVE THE ISSUE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2025-00441, AND 9680794-2025-00442.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21985367 ↗MalfunctionMay 8, 2025Crack; Crack
IT WAS REPORTED "ON (B)(6) 2025, IN THE COURSE OF DIALYSING A PATIENT, THE DOCTOR FOUND LUER HUB/FITTING HAS CRACK." THE USER CHANGED TO A NEW SET TO RESOLVE THE ISSUE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 9680794-2025-00440 AND 9680794-2025-00442.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21985349 ↗MalfunctionFeb 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT "DURING OPERATION, WHEN THE DOCTOR OPENED THE PACKAGE, FOUND THE PEEL-AWAY SHEATH SPLIT". THE DEVICE WAS NOT USED ON THE PATIENT. THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21356098 ↗MalfunctionJan 29, 2025Crack; Crack
IT WAS REPORTED THAT "THE DOCTOR FOUND THE LUER HUB/FITTING HAS CRACK DURING USED ON THE PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21264258 ↗MalfunctionJan 15, 2025Crack; Crack
IT WAS REPORTED THAT "THE DOCTOR FOUND THE LUER HUB/FITTING HAS CRACK DURING USED ON THE DIFFERENT PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED. THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2025-00088.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21159595 ↗MalfunctionJan 15, 2025Crack; Crack
IT WAS REPORTED THAT "THE DOCTOR FOUND THE LUER HUB/FITTING HAS CRACK DURING USED ON THE DIFFERENT PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED. THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2025-00087.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21159585 ↗MalfunctionJan 10, 2025Crack; Crack
IT WAS REPORTED THAT "THE DOCTOR FOUND LUER HUB/FITTING HAS CRACK DURING USED ON THE PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED. THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2025-00045 AND 9680794-2025-00046.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21124401 ↗MalfunctionJan 10, 2025Crack; Crack
IT WAS REPORTED THAT "THE DOCTOR FOUND LUER HUB/FITTING HAS CRACK DURING USED ON THE PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED. THE ASSOCIATED EMDR REPORT NUMBERS ARE 9680794-2025-00046.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21124118 ↗MalfunctionJan 8, 2025Crack; Crack
IT WAS REPORTED THAT "THE DOCTOR FOUND THE LUER HUB/FITTING HAS CRACK DURING USED ON THE PATIENT". THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21101316 ↗MalfunctionDec 6, 2024Crack; Crack
IT WAS REPORTED "ON (B)(6) 2024, THE DOCTOR FOUND THE LUER HUB/FITTING HAS A CRACK DURING USE ON THE PATIENT. THEY CHANGED A NEW ONE TO RESOLVE THE ISSUE. NO HARM FOR THE PATIENT. THE PATIENT'S CONDITION IS FINE. NO MEDICAL INTERVENTION WAS REQUIRED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20866123 ↗MalfunctionDec 4, 2024Crack; Crack
IT WAS REPORTED " ON (B)(6) 2024, THE DOCTOR FOUND THE LUER HUB/FITTING HAS A CRACK DURING USE ON THE PATIENT." THE OPERATOR CHANGED TO A NEW SET. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20843283 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111117. One submission often covers many models, so check the product and lot details in each record.
- Z-2158-2026 · Apr 10, 2026 · Open, ClassifiedCannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter w
- Z-2156-2026 · Apr 10, 2026 · Open, ClassifiedArrow Edge Hemodialysis Catheterization Product, REF: CS-15242-I, CS-15282-I, CS-15322-I
- Z-2159-2026 · Apr 10, 2026 · Open, ClassifiedNextStep Antegrade Chronic Hemodialysis Catheter, REF: AC-15192-SFX, AC-15232-SFX, CS-15192-SFX, CS-15232-SFX, CS-15272-SFX, CS-15312-SFX, CS-15422-SFX, CS-15502-SFX; NextStep A
- Z-1007-2015 · Feb 4, 2013 · TerminatedThe Arrow Edge Hemodialysis Catheterization Set Long-Term Access and NextStep Antegrade Hemodialysis Catheterization Set, Long-Term Access, part numbers: AC-15192-SFX, AC-15232-SFX
Other Catheter, hemodialysis, implanted devices
- Palindrome 8888145014Mozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall312 reports
- Palindrome 8888145015Mozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death219 reports
- HemoStar Long-Term Hemodialysis Catheter 5833690Bard Access Systems, Inc. · Catheter, hemodialysis, implanted2 deaths290 reports
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- HemoStar Long-Term Hemodialysis Catheter 5833730Bard Access Systems, Inc. · Catheter, hemodialysis, implanted1 death141 reports
- GlidePath Long-Term Hemodialysis Catheter with Preloaded Stylet 5393230Bard Access Systems, Inc. · Catheter, hemodialysis, implanted5 deaths138 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.