The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00801902099471), received between Mar 22, 2021 and Jan 15, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).
54 other recalls cover devices cleared under the same FDA submission (K071538); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Deformation Due to Compressive Stress2
- Physical Resistance/Sticking2
- Break1
- Difficult to Remove1
- Fluid/Blood Leak1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
- Needle Stick/Puncture1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2026Difficult to Remove
IT WAS REPORTED THAT "DURING THE PLACEMENT OF THE CVC, THE GUIDE ADVANCED WITHOUT DIFFICULTY, AND AFTER ADVANCING THE CATHETER, WHEN ATTEMPTING TO WITHDRAW THE GUIDE, IT WAS FOUND TO BE BLOCKED. THE CATHETER AND METALLIC GUIDE WERE THEN REMOVED." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24082382 ↗MalfunctionApr 15, 2025Break; Break
IT WAS REPORTED "WHEN THE CATHETER IS OPENED AND THE METAL GUIDE IS INSERTED, IT IS BROKEN." ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER STATES " THE PATIENT HAD TO BE PUNCTURED AGAIN. THE GUIDEWIRE WAS BROKEN WHEN IT WAS INSERTED AND ACCESS WAS NOT FEASIBLE. ONLY THE PHYSICIAN PERFORMING THE PROCEDURE CAN BE FITTED. THEREFORE, ANOTHER CATHETER WAS OPENED AND THE PATIENT WAS PUNCTURED AGAIN." THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "STABLE".
Patient problems: Needle Stick/Puncture
MAUDE report 21838448 ↗MalfunctionAug 9, 2023Unraveled Material; Deformation Due to Compressive Stress; Physical Resistance/Sticking; Unraveled Material; Deformation Due to Compressive Stress; Physical Resistance/Sticking
IT WAS REPORTED "THE GUIDE DID NOT FIT WITH THE CATHETER" DURING USE. A NEW PRODUCT WAS USED. NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17492325 ↗MalfunctionAug 9, 2023Deformation Due to Compressive Stress; Physical Resistance/Sticking; Deformation Due to Compressive Stress; Physical Resistance/Sticking
IT WAS REPORTED "THE GUIDE DID NOT FIT WITH THE CATHETER" DURING INSERTION. A NEW PRODUCT WAS USED.
Patient problems: Insufficient Information
MAUDE report 17492218 ↗MalfunctionMar 22, 2021Fluid/Blood Leak; Fluid/Blood Leak
THE COMPLAINT IS REPORTED AS: CENTRAL LINE WAS LEAKING AT THE EXTENSION LINE. NO PATIENT INJURY OR HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11542915 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071538. One submission often covers many models, so check the product and lot details in each record.
- Z-2376-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
- Z-1914-2026 · Mar 11, 2026 · Open, ClassifiedARROW, ARROWG+ARD AND ARROWG+ARD BLUE PLUS PRESSURE INJECTABLE CENTRAL VENOUS CATHETER REF ASK-42802-PLH UDI code: (01)10801902133110(17)261231(11)251211(10)33F25L0140 REF ASK-4
- Z-2123-2026 · Jan 7, 2026 · Open, ClassifiedMedline Convenience Kits: 1) KIT TRIPLE LUMEN INSERTION, Model Number: DYNDM1031A
- Z-0003-2024 · Aug 10, 2023 · Open, ClassifiedArrow Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
- Z-0002-2024 · Aug 10, 2023 · Open, ClassifiedArrow ErgoPack Complete, Pressure Injectable Arrowg+ard Blue Plus Four-Lumen CVC, REF CDC-42854- PCN1A; catheter, intravascular, therapeutic, long-term greater than 30 days
- Z-0352-2023 · Oct 27, 2022 · TerminatedPressure Injectable Arrowg+ard Blue Plus Three-Lumen CVC Kit, REF numbers: a) ASK-42703-NS, b) ASK-42703-PCCH2, c) ASK-42703-PUPM1, d) ASK-45703-NS, e) ASK-45703-PAU, f) ASK-45703-
- Z-0351-2023 · Oct 27, 2022 · TerminatedArrow MAC Two-Lumen Central Venous Access Kit fr use with 7 - 7.5 Fr Catheters, REF ASK-11242-UPM1; catheter introducer
- Z-1604-2022 · May 25, 2022 · Open, ClassifiedTELEFLEX LLC Arrow Central Venous Catheter Kit 16, 18 Gauge Multi Lumen Model Number:CDC-45703-XP1A (Model Numbers # IPN035619, IPN036437, IPN922677)
- Z-1980-2018 · Apr 11, 2018 · TerminatedPressure Injectable Multi-Lumen Central Venous Catheterization Kit; Product Code: ASK-15703-PHMC
- Z-1989-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PLUS Pressure Injectable Quad-Lumen CVC Kit; Product Codes: CDC-42854-P1A and CDC-45854-P1A
- Z-1997-2018 · Apr 11, 2018 · TerminatedPressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Codes: ASK-42703-PTJH
- Z-1979-2018 · Apr 11, 2018 · TerminatedPressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip(R) Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product Code: ASK-15703-LD
- Z-1990-2018 · Apr 11, 2018 · TerminatedPressure Injectable Two-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Code: ASK-45802-PHMW
- Z-1987-2018 · Apr 11, 2018 · TerminatedMulti-Lumen Central Venous Catheterization Kit with Blue FlexTip(R), ARROWg+ard Blue PLUS(R) Catheter and Sharps Safety Features; Product Codes: ASK-42703-MHW and ASK-42703-OLL1
- Z-2000-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue PLUS(R) Multi-Lumen Central Venous Catheterization Kit; Product Code: ASK-42703-SJH
…and 39 more.
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days660 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls527 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death608 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days582 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls374 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page