ARROW IPN030162Recalls & reported problems
Teleflex Incorporated · Multi-Lumen Central Venous Catheterization Set with Blue FlexTip(R) ARROWg+ard Blue(R) Catheter · Class 2 · Implant
All models of ARROW (1488)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Teleflex Arrow central venous catheter set is a multi-lumen catheter used for short-term vein access. Of its 163 adverse event reports, 160 are malfunctions and 3 are injuries, with no deaths. Many describe guidewires that bent or would not pass, and some describe the catheter hub blocking or leaking. One describes redness, swelling and allergic reactions during placement. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 163
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 163 adverse event reports that list this exact device (DI 00801902102324), received between Jan 29, 2021 and Mar 30, 2026 — classified as 3 injuries, 160 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (58 reports).
12 other recalls cover devices cleared under the same FDA submission (K900263); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent58
- Physical Resistance/Sticking19
- Deformation Due to Compressive Stress16
- Fluid/Blood Leak13
- Gas/Air Leak9
- Material Separation9
- Break8
- Difficult to Advance8
Patient problems
- No Clinical Signs, Symptoms or Conditions153
- Needle Stick/Puncture5
- Hypersensitivity/Allergic reaction2
- Insufficient Information2
- Low Blood Pressure/ Hypotension2
- Rash2
- Anaphylactic Shock1
- Swelling/ Edema1
Recent adverse event reports
The 25 most recent of 163 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Obstruction of Flow
IT WAS REPORTED THAT "THE WHITE HUB EXHIBITED POOR FLOW IMMEDIATELY AFTER PLACEMENT AND BECAME COMPLETELY BLOCKED/OCCLUDED LATER THAT AFTERNOON". THE USER CHANGED TO A NEW SET TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24728641 ↗MalfunctionMar 9, 2026Failure to Advance
IT WAS REPORTED THAT "THE GUIDE WIRE COULD GET NOT THROUGH THE ARS." THE USER CHANGED TO A NEW SET TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24546171 ↗MalfunctionMar 5, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT "THE DEFECTIVE PRODUCT CAUSED REDNESS, SWELLING, AND ALLERGIC REACTIONS IN PATIENTS DURING PLACEMENT." THE SYMPTOMS WERE RESOLVED AFTER REMOVAL. THE USER CHANGED TO A NEW SET. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "SAFE".
Patient problems: Hypersensitivity/Allergic reaction; Rash; Swelling/ Edema
MAUDE report 24518818 ↗MalfunctionFeb 24, 2026Material Twisted/Bent
IT WAS REPORTED THAT " GUIDE WIRE BENT". THE USER CHANGED TO A NEW SET TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24428771 ↗MalfunctionFeb 16, 2026Material Twisted/Bent
IT WAS REPORTED THAT "THE GUIDE WIRE BENT." THE USER CHANGED TO A NEW SET TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24344166 ↗MalfunctionFeb 16, 2026Material Twisted/Bent
IT WAS REPORTED THAT "THE GUIDE WIRE BENT." THE USER CHANGED TO A NEW SET TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24344151 ↗MalfunctionJan 21, 2026Material Twisted/Bent
IT WAS REPORTED THAT "DURING THE PROCEDURE ACCORDING TO IFU, THE MD FOUND THE DILATOR FAILED TO PENETRATE THE SUBCUTANEOUS ISSUE. WHEN I CHECKED WHETHER OR NOT TO USE GUIDEWIRE, IT WAS SAID TO HAVE BEEN USED, AND WHEN I TOOK IT OUT, IT WAS SAID THAT THE GUIDEWIRE WAS BENT. SO, THE MD OPENED UP THE NEW KIT TO FINISH THE PROCEDURE." ADDITIONAL INFORMATION RECEIVED STATES " ACCORDING TO THE USER, THE GUIDEWIRE WAS USED DURING THE PROCEDURE, AND UPON REMOVAL FROM THE PATIENT'S BODY, IT WAS OBSERVED TO BE BENT. THE USER ALSO REPORTED THAT THE INITIAL ATTEMPT WAS UNSUCCESSFUL IN PENETRATING THE SUBCUTANEOUS LAYER, AFTER WHICH A NEW DEVICE WAS OPENED AND USED." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24131288 ↗MalfunctionJan 15, 2026Material Twisted/Bent; Physical Resistance/Sticking
IT WAS REPORTED THAT "DURING THE PROCEDURE ACCORDING TO IFU, THE MD DOESN'T GO INSIDE THE GUIDEWIRE, IT BOUNCES OUT. SO THE MD OPENED THE NEW KIT TO FINISH THE PROCEDURE". ADDITIONAL INFORMATION RECEIVED STATES "UPON WITHDRAWAL, THE GUIDEWIRE WAS FOUND TO BE BENT". THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24080767 ↗MalfunctionJan 15, 2026Material Twisted/Bent
IT WAS REPORTED THAT "DURING THE PROCEDURE ACCORDING TO IFU, THE MD FOUND THE DILATOR FAILED TO PENETRATE THE SUBCUTANEOUS ISSUE. WHEN I CHECKED WHETHER OR NOT TO USE GUIDEWIRE, IT WAS SAID TO HAVE BEEN USED, AND WHEN I TOOK IT OUT, IT WAS SAID THAT THE GUIDEWIRE WAS BENT. SO, THE MD OPENED UP THE NEW KIT TO FINISH THE PROCEDURE". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24080754 ↗MalfunctionJan 15, 2026Leak/Splash
IT WAS REPORTED "THE DEVICE LEAKED WHEN CVC WAS INSERTED INTO THE PATIENT". THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "UNKNOWN".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24080134 ↗MalfunctionJan 15, 2026Leak/Splash
IT WAS REPORTED THAT "THE DEVICE LEAKED WHEN CVP WAS INSERTED INTO THE PATIENT". THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24079709 ↗MalfunctionJan 13, 2026Material Twisted/Bent; Physical Resistance/Sticking
IT WAS REPORTED THAT " ON (B)(6) 2025, IN THE ANESTHESIOLOGY DEPARTMENT OF (B)(6) HOSPITAL, DURING THE INSERTION, THE DOCTOR FOUND THAT THE GUIDEWIRE COULD NOT PASS THE ARS SMOOTHLY, THE RESISTANCE WAS FOUND THEREFORE, THE CATHETER HAD TO BE REPLACED FOR REINSERTION." ADDITIONAL INFORMTION WAS RECIEVED THAT STATED "THE GUIDWIRE WAS KINKED". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24054284 ↗MalfunctionJan 13, 2026Material Twisted/Bent; Physical Resistance/Sticking
IT WAS REPORTED THAT "(B)(6) 2025, IN THE ANESTHESIOLOGY DEPARTMENT OF BEIJING HOSPITAL, DURING THE INSERTION, THE DOCTOR FOUND THAT THE GUIDEWIRE COULD NOT PASS THE ARS SMOOTHLY, THE RESISTANCE WAS FOUND THEREFORE, THE CATHETER HAD TO BE REPLACED FOR REINSERTION." ADDITIONAL INFORMTION WAS RECIEVED THAT STATED "THE GUIDWIRE WAS KINKED". THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24054148 ↗MalfunctionJan 13, 2026Fluid/Blood Leak
IT WAS REPORTED THAT "THE LUER-HUB (WHITE) LEAKED AFTER IT WAS INSERTED ON THE PATIENT ON THE 16TH DAY". THE DEVICE WAS REMOVED AND REPLACED. THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Insufficient Information
MAUDE report 24053968 ↗MalfunctionJan 7, 2026Positioning Problem
IT WAS REPORTED THAT "DURING THE PROCEDURE ACCORDING TO IFU, THE MD FOUND WIRE IS MISSING AND GETTING OUT OF JUGULAR WHILE REMOVING GUIDE WIRE (IT'S A NEW TYPE OF TRAY, AND 3 LOT OCCURRED, BUT ONLY 1 LOT WAS RECOVERED). SO, THE MD OPENED UP THE NEW KIT TO FINISH THE PROCEDURE." ADDITIONAL INFORMATION CLARIFIED THE GUIDE WIRE WAS NOT MISSING. HOWEVER, "THE OPERATOR FELT THAT THE GUIDE WIRE WAS SLIPPING OR FREE SPINNING." THERE WAS NO PATIENT HARM OR INJURY TO THE PATIENT. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 3006425876-2026-00014, AND 3006425876-2026-00007.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23989263 ↗MalfunctionJan 7, 2026Positioning Problem
IT WAS REPORTED THAT "DURING THE PROCEDURE ACCORDING TO IFU, THE MD FOUND WIRE IS MISSING AND GETTING OUT OF JUGULAR WHILE REMOVING GUIDE WIRE (IT'S A NEW TYPE OF TRAY, AND 3 LOT OCCURRED, BUT ONLY 1 LOT WAS RECOVERED). SO, THE MD OPENED UP THE NEW KIT TO FINISH THE PROCEDURE." ADDITIONAL INFORMATION CLARIFIED THE GUIDE WIRE WAS NOT MISSING. HOWEVER, "THE OPERATOR FELT THAT THE GUIDE WIRE WAS SLIPPING OR FREE SPINNING." THERE WAS NO PATIENT HARM OR INJURY TO THE PATIENT. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 3006425876-2026-00014, 3006425876-2026-00015, AND 3006425876-2026-00007.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23989252 ↗MalfunctionJan 7, 2026Positioning Problem
IT WAS REPORTED THAT "DURING THE PROCEDURE ACCORDING TO IFU, THE MD FOUND WIRE IS MISSING AND GETTING OUT OF JUGULAR WHILE REMOVING GUIDE WIRE (IT'S A NEW TYPE OF TRAY, AND 3 LOT OCCURRED, BUT ONLY 1 LOT WAS RECOVERED). SO, THE MD OPENED UP THE NEW KIT TO FINISH THE PROCEDURE." ADDITIONAL INFORMATION CLARIFIED THE GUIDE WIRE WAS NOT MISSING. HOWEVER, "THE OPERATOR FELT THAT THE GUIDE WIRE WAS SLIPPING OR FREE SPINNING." THERE WAS NO PATIENT HARM OR INJURY TO THE PATIENT. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23989097 ↗MalfunctionJan 7, 2026Material Twisted/Bent
IT WAS REPORTED THAT " THE INSERTION NEEDLES BENT (USING FOUR DIFFERENT SETS FROM THE SAME BATCH) DURING PUNCTURE OF THE RIGHT JUGULAR VEIN. THE FOURTH NEEDLE EVEN BROKE. THERE WERE NO SERIOUS COMPLICATIONS FOR THE PATIENT." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 3006425876-2026-00056, 3006425876-2026-00057, 3006425876-2026-00058, AND 3006425876-2026-00059.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23988706 ↗MalfunctionJan 7, 2026Material Twisted/Bent
IT WAS REPORTED THAT " THE INSERTION NEEDLES BENT (USING FOUR DIFFERENT SETS FROM THE SAME BATCH) DURING PUNCTURE OF THE RIGHT JUGULAR VEIN. THE FOURTH NEEDLE EVEN BROKE. THERE WERE NO SERIOUS COMPLICATIONS FOR THE PATIENT." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 3006425876-2026-00056, 3006425876-2026-00057, 3006425876-2026-00058, AND 3006425876-2026-00059.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23988699 ↗MalfunctionJan 7, 2026Material Twisted/Bent
IT WAS REPORTED THAT "THE INSERTION NEEDLES BENT (USING FOUR DIFFERENT SETS FROM THE SAME BATCH) DURING PUNCTURE OF THE RIGHT JUGULAR VEIN. THE FOURTH NEEDLE EVEN BROKE. THERE WERE NO SERIOUS COMPLICATIONS FOR THE PATIENT." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 3006425876-2026-00056, 3006425876-2026-00057, 3006425876-2026-00058, AND 3006425876-2026-00059.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23988684 ↗MalfunctionJan 7, 2026Break
IT WAS REPORTED THAT " THE INSERTION NEEDLES BENT (USING FOUR DIFFERENT SETS FROM THE SAME BATCH) DURING PUNCTURE OF THE RIGHT JUGULAR VEIN. THE FOURTH NEEDLE EVEN BROKE. THERE WERE NO SERIOUS COMPLICATIONS FOR THE PATIENT." THERE WAS NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE". ASSOCIATED MDR NUMBERS INCLUDE: 3006425876-2026-00056, 3006425876-2026-00057, 3006425876-2026-00058, AND 3006425876-2026-00059.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23988678 ↗MalfunctionJan 6, 2026Crack
IT WAS REPORTED THAT "CRACKS WERE FOUND IN THE ARS, CAUSING BLOOD TO SEEP DURING THE INJECTION OF SALINE AND THE WITHDRAWAL OF BLOOD". THE USER CHANGED TO A NEW SET TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23973374 ↗MalfunctionDec 30, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT "DURING ASPIRATION, GAS WAS CONSISTENTLY RETRIEVED, RAISING SUSPICION OF A LEAK IN THE CATHETER BODY. FOR THE SAKE OF THE PATIENT'S SAFETY, A NEW CATHETER WAS REPLACED". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23926176 ↗MalfunctionNov 24, 2025Difficult to Remove; Unraveled Material; Difficult to Remove; Unraveled Material
IT WAS REPORTED THAT "THE GUIDE WIRE IS LOOSE AND STUCK INSIDE THE CATHETER, MAKING IT IMPOSSIBLE TO REMOVE NORMALLY. IT CAN ONLY BE PULLED OUT TOGETHER WITH THE LUMEN THAT HAS BEEN INSERTED, REQUIRING A RE-INSERTION." ADDITIONAL INFORMATION RECEIVED STATES "SWG IS UNRAVELED BUT DID NOT SEPARATE. UNRAVELED SWG AND CATHETER WERE REMOVED TOGETHER." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: Needle Stick/Puncture
MAUDE report 23631617 ↗MalfunctionNov 20, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT "THE NEEDLE CANNOT THROUGH THE ARS." THERE WAS NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23606219 ↗
FDA clearance and classification
Cleared through 510(k) K900263 on Jul 24, 1990. Submitted by Arrow Intl., Inc. as “Antimicrobial Multi-Lumen Central Venous Catheter”. The FDA review took 187 days. FDA record ↗
The FDA regulates product code FOZ (catheter,intravascular,therapeutic,short-term less than 30 days) under 21 CFR 880.5200 as a Class II device. 21 CFR 880.5200 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K900263. One submission often covers many models, so check the product and lot details in each record.
- Z-2376-2026 · Apr 21, 2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-
- Z-0053-2020 · May 17, 2019 · TerminatedPediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, REF AK-25502 Product Usage; Provide short-term (< 30 days) central venous acces
- Z-2886-2018 · Jul 3, 2018 · TerminatedPediatric Two-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter, Cat. No. AK-22502 Product Usage: The multiple-lumen catheter permits venous acc
- Z-1985-2018 · Apr 11, 2018 · TerminatedARROWg+ard Blue(R) Central Venous Catheterization Kit; Product Code: ASK-24306-WHC1 and ASK-24306-WHC2
- Z-1984-2018 · Apr 11, 2018 · TerminatedSingle-Lumen Central Venous Catheterization Kit with Blue FlexTip(R) ARROWg+ard Blue(R) Catheter and Sharps Safety Features; Product Code: ASK-24306-VC
- Z-1333-2018 · Dec 22, 2017 · TerminatedARROWg+ard Blue CVC Kit (1) Part Number (PN): CDC-24301-1A, (2) PN: CDC-24306-1A; (3) ARROWg+ard Blue Large-Bore Multi-Lumen CVC Kit, PN: CDC-22123-1A; (4) Central Venous Catheter
- Z-1001-2018 · Nov 27, 2017 · TerminatedARROWg+ard Blue Large-Bore Multi-Lumen CVC Kit Product Usage: The Arrow CVC is indicated to provide short-term (< 30 days) central venous access for treatment of diseases or condi
- Z-2337-2017 · May 9, 2017 · TerminatedARROWg+ard Blue CVC Kit The ARROWg+ard Blue CVCs are intended to permit venous access to the central circulation
- Z-1830-2012 · May 3, 2012 · TerminatedArrow Two-Lumen Central Venous Catheterization Set with Blue FlexTip(R) ARROWgard Blue(R) Catheter Product Usage: The multiple-lumen catheter permits venous access to central c
- Z-2496-2010 · Jul 29, 2010 · TerminatedTwo-Lumen Central Venous Catheterization Kit with Blue FlexTip ARROWg+ard Blue Catheter
- Z-1499-2009 · Feb 11, 2009 · TerminatedArrow ARROWg+ard CVCs (Central Venous Catheters)
- Z-0082-04 · Sep 16, 2003 · TerminatedArrow Howes multi-lumen central venous catheterization kit 7 Fr x 20 cm
Common questions about the ARROW IPN030162
Has the ARROW IPN030162 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00801902102324), based on the FDA recall file dated Nov 27, 2026. However, 12 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ARROW IPN030162?
The FDA's MAUDE database lists 163 adverse event reports for this device, received between Jan 29, 2021 and Mar 30, 2026. In those reports, the most common specific device problems are “Material Twisted/Bent” (58), “Physical Resistance/Sticking” (19) and “Deformation Due to Compressive Stress” (16); the most common patient problems are “Needle Stick/Puncture” (5), “Hypersensitivity/Allergic reaction” (2) and “Low Blood Pressure/ Hypotension” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ARROW IPN030162?
None of the reports are classified as deaths, 3 as injuries and 160 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ARROW IPN030162 FDA approved?
It was cleared through 510(k) K900263 on Jul 24, 1990. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ARROW IPN030162 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the ARROW IPN030162, and is it still sold?
The FDA's device database lists Teleflex Incorporated as the company responsible for it (brand ARROW, model IPN030162). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter,intravascular,therapeutic,short-term less than 30 days devices
- BD Insyte Autoguard 382523Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days659 reports
- BD Insyte Autoguard 382533Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls528 reports
- BD PosiFlush 306546Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 death607 reports
- BD Insyte Autoguard Shielded IV Catheter 381823Becton, Dickinson And Company · Catheter, intravascular, therapeutic, short-term less than 30 days579 reports
- BD Insyte Autoguard 381423Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall528 reports
- BD Insyte Autoguard 382544Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days1 recall437 reports
- BD Nexiva, BD Q-Syte 383536Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls372 reports
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death330 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page