Skip to content

LATITUDE® Communicator 6476Recalls & reported problems

Boston Scientific Corporation · LATITUDE ® Communicator · Class 3

All models of LATITUDE Communicator (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00802526516115), received between Jul 4, 2018 and Jan 26, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Power Problem” (3 reports).

10 other recalls cover devices cleared under the same FDA submission (P910077); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Power Problem3
  • Energy Output Problem2
  • Defective Device1
  • Device Operates Differently Than Expected1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT THE COMMUNICATOR HAD BLOWN UP. THE PATIENT TOOK THE COMMUNICATOR TO THE DOCTOR'S OFFICE, AND THE COMPANY REPRESENTATIVE OPTED TO REPLACE THE COMMUNICATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18586219 ↗
  • MalfunctionSep 15, 2023Energy Output Problem; Energy Output Problem

    IT WAS REPORTED THAT A LIGHTNING STORM HAS IMPACTED THE COMMUNICATOR ABOUT A WEEK AGO AND HAD A BURNING SMELL. A BOSTON SCIENTIFIC COMPANY REPRESENTATIVE HAS OPTED TO SEND FOR THE COMMUNICATOR AND CELLULAR ADAPTER. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE COMMUNICATOR IS NO LONGER IN SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17755820 ↗
  • MalfunctionJan 16, 2023Power Problem; Power Problem

    IT WAS REPORTED THAT THE COMMUNICATOR POWER CORD GOT SO HOT. THE COMMUNICATOR BURNED UP AND STARTED TO SMOKE. THE PATIENT BROUGHT THE COMMUNICATOR TO CLINIC AND THE PATIENT WAS GIVEN A NEW ONE. THE COMMUNICATOR WAS DEACTIVATED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16173319 ↗
  • MalfunctionFeb 6, 2022Power Problem; Power Problem

    IT WAS REPORTED THAT THE COMMUNICATOR POWER CORD WAS HOT, EMITTING SMOKE AND WAS SPARKING. THE COMMUNICATOR WAS FOR REPLACEMENT AND WILL BE RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13467430 ↗
  • MalfunctionJul 16, 2020Power Problem; Power Problem

    IT WAS REPORTED THAT THE COMMUNICATOR POWER CORD SHORTED OUT AND NO LONGER HAD LIGHTS ON. THE PATIENT MENTIONED THAT THE POWER CORD SPARKED WHEN TRIED TO PLUG IN. DUE TO THE POTENTIAL DAMAGE, THE COMPANY REPRESENTATIVE ADVISED ABOUT THE PRODUCT RETURN AND REPLACEMENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10287057 ↗
  • MalfunctionJun 18, 2019Energy Output Problem; Energy Output Problem

    BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT REPORTED THAT THE COMMUNICATOR SMELLED LIKE BURNT WIRES AFTER THE LIGHTNING STRIKE NEAR HIS HOME. THE COMPANY REPRESENTATIVE ADVISED THE PATIENT THAT A NEW COMMUNICATOR AND A CELLULAR ADAPTER WILL BE REPLACED. NO ADVERSE PATIENTS EFFECT REPORTED. THE COMMUNICATOR WAS DEACTIVATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8707212 ↗
  • MalfunctionJul 4, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected

    BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE COMMUNICATOR WAS SMOKING. COMMUNICATOR WAS DEACTIVATED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7662649 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P910077. One submission often covers many models, so check the product and lot details in each record.

  • Z-0673-2024 · Nov 30, 2023 · Open, Classified
    Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: AL
  • Z-1216-2023 · Feb 2, 2023 · Open, Classified
    LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
  • Z-1669-2022 · Jul 7, 2022 · Open, Classified
    The Programmer/Recorder/Monitor (PRM) is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The
  • Z-3260-2018 · Sep 10, 2018 · Terminated
    Model 3300 LATITUDE(TM) Programming System Product Usage - The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communic
  • Z-1445-2012 · Feb 24, 2012 · Terminated
    Boston Scientific LATITUDE Patient Management System Communicator, Model 6476, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, Made in Malaysi
  • Z-1096-2012 · Dec 19, 2011 · Terminated
    LATITUDE Patient management System, Model 6488, software version 7.2. It is indicated for use to remotely communicate with a compatible Boston Scientific pulse generator and transf
  • Z-0333-2012 · Oct 12, 2011 · Terminated
    LATITUDE Patient Management System, Models 6441*, 6442*, 6465*, 6488 (USA). There are two software applications: the programmer software and the LATITUDE remote monitoring system
  • Z-0037-2012 · Aug 12, 2011 · Terminated
    LATITUDE Paceart Integration Software, version 1.00 Model 6472. LATITUDE Paceart Integration is software and therefore does not have any serial numbers. However, the software ver
  • Z-1631-2010 · Sep 10, 2009 · Terminated
    Boston Scientific, LATITUDE Communicator, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112. Models 6476 (USA) and 6468 (OUS)
  • Z-0002-2007 · May 19, 2006 · Terminated
    Guidant Zoom Latitude Programming System, Model 3120, Programmer/Recorder/Monitor (PRM). A portable cardiac rhythm management system designed to be used with Guidant implantable p

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page