INOGEN™ EL ICD VR D140Recalls & reported problems

Boston Scientific Corporation · Implantable Cardioverter Defibrillator VR · Class 3 · Implant

All models of INOGEN EL ICD VR (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Boston Scientific Inogen defibrillator has 807 adverse event reports, 555 describing injuries and 4 describing deaths. Many describe batteries running down earlier than expected and shocks or pacing that were not appropriate, and infection and sepsis are also mentioned. No recalls name this device. Reports fell from 97 in the previous 12 months to 61 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
823
Deaths reported
4
Injuries reported
559
Recalls naming it
0

Summary

The FDA's MAUDE database contains 823 adverse event reports that list this exact device (DI 00802526534201), received between Apr 15, 2016 and Sep 29, 2026 — classified as 4 deaths, 559 injuries, 260 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (166 reports).

14 other recalls cover devices cleared under the same FDA submission (P960040); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 5 reportsJan 2024: 9 reportsFeb 2024: 7 reportsMar 2024: 12 reportsApr 2024: 9 reportsMay 2024: 5 reportsJun 2024: 8 reportsJul 2024: 6 reportsAug 2024: 7 reportsSep 2024: 8 reportsOct 2024: 13 reportsNov 2024: 13 reportsDec 2024: 8 reportsJan 2025: 11 reportsFeb 2025: 7 reportsMar 2025: 12 reportsApr 2025: 9 reportsMay 2025: 7 reportsJun 2025: 6 reportsJul 2025: 5 reportsAug 2025: 8 reportsSep 2025: 5 reportsOct 2025: 6 reportsNov 2025: 8 reportsDec 2025: 5 reportsJan 2026: 9 reportsFeb 2026: 8 reportsMar 2026: 4 reportsApr 2026: 5 reportsMay 2026: 2 reportsJun 2026: 5 reportsJul 2026: 7 reportsAug 2026: 7 reportsSep 2026: 17 reportsOct 2026: 0 reports20242025202617
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem240
  • Premature Discharge of Battery166
  • High impedance152
  • Inappropriate/Inadequate Shock/Stimulation124
  • Pacing Problem101
  • Over-Sensing60
  • Signal Artifact/Noise36
  • Incorrect, Inadequate or Imprecise Result or Readings30

Patient problems

  • No Clinical Signs, Symptoms or Conditions306
  • Unspecified Infection220
  • Sepsis90
  • No Known Impact Or Consequence To Patient74
  • Electric Shock63
  • Shock from Patient Lead(s)48
  • No Code Available43
  • No Consequences Or Impact To Patient26

Recent adverse event reports

The 25 most recent of 823 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938893 ↗
  • InjurySep 29, 2026Premature Discharge of Battery

    It was reported that this Implantable Cardioverter Defibrillator (ICD) recorded a Code 1003 indicative of battery voltage too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis low voltage fault and that this device power consumption recently began dropping in a gradual fashion. Subsequently, this device was explanted and replaced. A new device was implanted. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937004 ↗
  • MalfunctionSep 21, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848229 ↗
  • MalfunctionSep 20, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50834869 ↗
  • MalfunctionSep 20, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50834819 ↗
  • MalfunctionSep 18, 2026Premature Discharge of Battery

    It was reported that this implantable cardioverter defibrillator (ICD) recorded a Code 1003 indicative of battery voltage too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis showed the battery appeared to be depleting more quickly than expected. Device replacement was recommended. This device remains in service. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50814678 ↗
  • MalfunctionSep 17, 2026Interrogation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812693 ↗
  • MalfunctionSep 16, 2026Over-Sensing; Pacing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50796743 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable cardioverter defibrillator (ICD) was part of the system revision due to infection. Reportedly, the patient had a Coronary Artery Bypass Grafting (CABG) and the ejection fraction (EF) had improved. Subsequently, the ICD was removed, and the patient had a blood clot with sepsis anomaly. Furthermore, the patient recovered to the procedures but now had the atrial flutter (AF) with decreased EF. No adverse patient effects were reported. The ICD was explanted.

    Patient problems: Atrial Flutter; Sepsis; Thrombosis/Thrombus; Unspecified Infection

    MAUDE report 50784911 ↗
  • InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable cardioverter defibrillator (ICD) was part of a system revision due to infection. There were no additional adverse patient effects reported. The ICD was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50733327 ↗
  • MalfunctionSep 8, 2026Premature Discharge of Battery

    It was reported that this implantable cardioverter defibrillator (ICD) recorded a code 1003, which states that the voltage is too low for projected remaining capacity. A request was made to have data from this device analyzed. Data analysis confirmed that the code appears consistent with premature battery depletion due to hydrogen degradation of the bypass capacitors. Device replacement was recommended. This device remains in service at this time. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50659814 ↗
  • MalfunctionSep 7, 2026Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50650667 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable cardioverter defibrillator (ICD) was part of a system revision due to infection. There was no additional adverse patient effects reported. The ICD was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50612373 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient with implantable cardioverter defibrillator (ICD) felt as though the lead wire was pulling out and was experiencing sharp pain and dizziness. TS referred the patient to the physician for further evaluation. To date, this ICD remains in service. No adverse patient effects were reported.

    Patient problems: Dizziness; Pain

    MAUDE report 50610690 ↗
  • InjurySep 4, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation

    It was reported that this implantable cardioverter defibrillator (ICD) delivered inappropriate shock and anti-tachycardia pacing (ATP) therapy for supraventricular tachycardia (SVT). Technical Services (TS) provided reprograming options. This device remains in service. No adverse patient effects were reported.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50602100 ↗
  • MalfunctionSep 4, 2026High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50602071 ↗
  • MalfunctionSep 4, 2026Signal Artifact/Noise; Over-Sensing

    It was reported that during a review of the recorded episodes of this implantable cardioverter defibrillator (ICD), an isolated noise oversensing episode was observed. Technical Services (TS) noted that the noise was likely from electromagnetic interference during a procedure, since therapy was programmed off for the episode. No adverse patient effects were reported. This device remains in service.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50601131 ↗
  • InjuryAug 18, 2026Premature Discharge of Battery

    It was reported that this implantable cardioverter defibrillator (ICD) recorded a Code 1003 indicative of battery voltage too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis showed the battery appeared to be depleting more quickly than expected due to an internal battery short. This ICD was explanted, replaced with a different model and returned for analysis. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50304645 ↗
  • InjuryAug 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that this implantable cardioverter defibrillator (ICD) was part of the system revision due to infection. No adverse patient effects were reported. The ICD was explanted.

    Patient problems: Unspecified Infection

    MAUDE report 50261993 ↗
  • MalfunctionAug 14, 2026High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50248293 ↗
  • MalfunctionAug 12, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50223130 ↗
  • MalfunctionAug 11, 2026Premature Discharge of Battery; Noise, Audible

    It was reported that this implantable cardioverter defibrillator (ICD) emitted beeping tones and recorded a Code 1003, which states that the voltage is too low for projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified that the power consumption is increasing in a gradual fashion. Device replacement was recommended. This device remains in service at this time. No adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50209672 ↗
  • InjuryAug 5, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation

    It was reported that this implantable cardioverter defibrillator (ICD) delivered inappropriate anti-tachycardia pacing (ATP) and six shocks due to supraventricular tachycardia (SVT). Reprograming of the device was performed. This device remains in service. No further adverse patient effects were reported.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50144377 ↗
  • InjuryAug 3, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation

    It was reported that this implantable cardioverter defibrillator (ICD) delivered rounds of inappropriate anti-tachycardia pacing (ATP) due to an episode of sinus tachycardia. The device remains in service. No adverse patient effects were reported.Additional information obtained of six ATP sequences and six shocks for a rhythm with gradual onset and ventricular rates below the programmed therapy threshold. This device remains in service. No additional adverse patient effects were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50122481 ↗
  • MalfunctionJul 27, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50003503 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P960040. One submission often covers many models, so check the product and lot details in each record.

  • Z-1670-2022 · Jul 7, 2022 · Open, Classified
    The PRM is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The software in use controls all
  • Z-0078-2018 · Oct 4, 2017 · Terminated
    RESONATE EL ICD DR, Model D433, Sterile.
  • Z-0079-2018 · Oct 4, 2017 · Terminated
    VIGILANT X4 CRT-D, Model G247, Sterile. Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) are indicated for patients with heart failure who receive stable optimal pharma
  • Z-0077-2018 · Oct 4, 2017 · Terminated
    RESONATE EL ICD VR, Model D432, Sterile.
  • Z-1555-2010 · Mar 15, 2010 · Terminated
    Boston Scientific CONFIENT, model E030, Implantable Cardioverter Defibrillator (ICD). The CONFIENT Implantable Cardioverter Defibrillator (ICDs) are designed to detect and termina
  • Z-0610-2010 · Dec 1, 2009 · Terminated
    Boston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models F102, F110, F111. Sterilized using ethyle
  • Z-1621-2009 · Mar 4, 2009 · Terminated
    GUIDANT VITALITY DS VR Model T135 VR; EL Model T127 DR; 2ICD Model T165 DR; 2ICD Model T175 VR; 2EL ICD Model T167 DR; 2 EL ICD Model T177 VR; DS Model T125 DR. Implantable Cardi
  • Z-1513-2008 · Jan 7, 2008 · Terminated
    VITALITY DR+ Implantable Cardioverter Defibrillator (ICD), model 1872
  • Z-1514-2008 · Jan 7, 2008 · Terminated
    VITALITY EL, models T127, T167, T177 Implantable Cardioverter Defibrillator (ICD), pulse-generator, dual Chamber
  • Z-0050-2008 · Mar 16, 2007 · Terminated
    Guidant VITALITY DS DR/VR (Models T125, T135), Guidant VITALITY EL (Model T127), Guidant VITALITY DR HE (Model T180), Implantable Cardioverter Defibrillators (ICD). Guidant Corpor
  • Z-0051-2008 · Mar 16, 2007 · Terminated
    Guidant VITALITY 2 DR/VR (Models T165, T175) and Guidant VITALITY 2 EL DR/VR (Models T167, T177). Implantable Cardioverter Defibrillator, ICD. Guidant Corporation, 4100 Hamline A
  • Z-1297-06 · Jun 24, 2006 · Terminated
    Guidant VENTAK PRIZM VR (model 1860) and VENTAK PRIZM DR (model 1861) Automatic Implantable Cardioverter Defibrillator (AICD), Dual Chamber, Sterilized with gaseous ethylene oxide.
  • Z-1292-06 · May 15, 2006 · Terminated
    Guidant VITALITY 2 (models T165, T175), VITALITY DS (model T125), VITALITY EL (model T135), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ven
  • Z-0828-06 · Apr 7, 2006 · Terminated
    Guidant VITALITY HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 551

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →