ORIGEN™ EL ICD VR D050Recalls & reported problems
Boston Scientific Corporation · Implantable Cardioverter Defibrillator VR · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00802526535024), received between Apr 9, 2022 and Apr 9, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Premature Discharge of Battery” (1 report).
14 other recalls cover devices cleared under the same FDA submission (P960040); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Premature Discharge of Battery1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 9, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. DATA ANALYSIS SHOWED THE BATTERY APPEARED TO BE DEPLETING MORE QUICKLY THAN EXPECTED. DEVICE REPLACEMENT WAS RECOMMENDED. THIS DEVICE REMAINS IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14068905 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P960040. One submission often covers many models, so check the product and lot details in each record.
- Z-1670-2022 · Jul 7, 2022 · Open, ClassifiedThe PRM is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The software in use controls all
- Z-0077-2018 · Oct 4, 2017 · TerminatedRESONATE EL ICD VR, Model D432, Sterile.
- Z-0078-2018 · Oct 4, 2017 · TerminatedRESONATE EL ICD DR, Model D433, Sterile.
- Z-0079-2018 · Oct 4, 2017 · TerminatedVIGILANT X4 CRT-D, Model G247, Sterile. Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) are indicated for patients with heart failure who receive stable optimal pharma
- Z-1555-2010 · Mar 15, 2010 · TerminatedBoston Scientific CONFIENT, model E030, Implantable Cardioverter Defibrillator (ICD). The CONFIENT Implantable Cardioverter Defibrillator (ICDs) are designed to detect and termina
- Z-0610-2010 · Dec 1, 2009 · TerminatedBoston Scientific TELIGEN, Implantable Cardiac Defibrillator models E102, E110, the following models are not available in the US: models F102, F110, F111. Sterilized using ethyle
- Z-1621-2009 · Mar 4, 2009 · TerminatedGUIDANT VITALITY DS VR Model T135 VR; EL Model T127 DR; 2ICD Model T165 DR; 2ICD Model T175 VR; 2EL ICD Model T167 DR; 2 EL ICD Model T177 VR; DS Model T125 DR. Implantable Cardi
- Z-1513-2008 · Jan 7, 2008 · TerminatedVITALITY DR+ Implantable Cardioverter Defibrillator (ICD), model 1872
- Z-1514-2008 · Jan 7, 2008 · TerminatedVITALITY EL, models T127, T167, T177 Implantable Cardioverter Defibrillator (ICD), pulse-generator, dual Chamber
- Z-0050-2008 · Mar 16, 2007 · TerminatedGuidant VITALITY DS DR/VR (Models T125, T135), Guidant VITALITY EL (Model T127), Guidant VITALITY DR HE (Model T180), Implantable Cardioverter Defibrillators (ICD). Guidant Corpor
- Z-0051-2008 · Mar 16, 2007 · TerminatedGuidant VITALITY 2 DR/VR (Models T165, T175) and Guidant VITALITY 2 EL DR/VR (Models T167, T177). Implantable Cardioverter Defibrillator, ICD. Guidant Corporation, 4100 Hamline A
- Z-1297-06 · Jun 24, 2006 · TerminatedGuidant VENTAK PRIZM VR (model 1860) and VENTAK PRIZM DR (model 1861) Automatic Implantable Cardioverter Defibrillator (AICD), Dual Chamber, Sterilized with gaseous ethylene oxide.
- Z-1292-06 · May 15, 2006 · TerminatedGuidant VITALITY 2 (models T165, T175), VITALITY DS (model T125), VITALITY EL (model T135), Implantable Cardioverter Defibrillators (ICD), are designed to detect and terminate ven
- Z-0828-06 · Apr 7, 2006 · TerminatedGuidant VITALITY HE Implantable Cardioverter Defibrillator (model T180). Sterile EO. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 551
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page