Skip to content

LATITUDE® Communicator 6280 (UDI 00802526544118)Recalls & reported problems

Boston Scientific Corporation · LATITUDE ® Communicator · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00802526544118), received between May 6, 2016 and Dec 27, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Power Problem” (2 reports).

10 other recalls cover devices cleared under the same FDA submission (P910077); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Power Problem2
  • Connection Problem1
  • Defective Device1
  • Device Operates Differently Than Expected1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 27, 2024Defective Device; Power Problem; Defective Device; Power Problem

    IT WAS REPORTED THAT THE COMMUNICATOR WAS FLASHING YELLOW CALL DOCTOR ICON (YCD. THE COMMUNICATOR POWER CORD WAS NOT POWER ON AND INSIDE OF THE COMMUNICATOR EXHIBITED HOT. A BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) ASSISTED THE PATIENT IN TROUBLESHOOTING. THERE WERE NO STORMS OR OUTAGES OBSERVED. THE COMMUNICATOR ISSUE PERSISTED. THE COMPANY REPRESENTATIVE ADVISED TO REPLACE THE COMMUNICATOR. THE COMMUNICATOR WAS NO LONGER IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21031031 ↗
  • MalfunctionJul 12, 2023Connection Problem; Power Problem; Connection Problem; Power Problem

    IT WAS REPORTED THAT THE COMMUNICATOR POWER CORD HAD A SHORT AND THE COMMUNICATOR WAS NOT CONNECTING. A BOSTON SCIENTIFIC COMPANY REPRESENTATIVE OPTED TO REPLACE THE COMMUNICATOR POWER CORD AND THE FOURTH-GENERATION COMMUNICATOR CELLULAR ADAPTER. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE COMMUNICATOR REMAINS IN SERVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17305179 ↗
  • MalfunctionMay 6, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected

    BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE COMMUNICATOR POWER CORD WAS MELTED. A BOSTON SCIENTIFIC COMPANY REPRESENTATIVE VERIFIED NO ONE WAS INJURED. THE COMPANY REPRESENTATIVE OFFERED TO REPLACE THE POWER CORD BUT THE PATIENT DOES NOT WANT TO USE THE COMMUNICATOR ANYMORE. THE COMPANY REPRESENTATIVE ADVISED THE PATIENT TO CALL THE PHYSICIAN IN REGARDS OF THE COMMUNICATOR USAGE. A RETURN LABEL WILL BE SEND WITH THE POWER CORD. THE COMMUNICATOR WAS OUT OF SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5636545 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P910077. One submission often covers many models, so check the product and lot details in each record.

  • Z-0673-2024 · Nov 30, 2023 · Open, Classified
    Boston Scientific Model 3120 ZOOM LATITUDE Programmer/Recorder/Monitor (PRM) installed with Model 2892 Programmer Application Software used for the following pacemaker products: AL
  • Z-1216-2023 · Feb 2, 2023 · Open, Classified
    LATITUDE NXT Remote Patient Management System, LATITUDE NXT System Server Software Model 6460.
  • Z-1669-2022 · Jul 7, 2022 · Open, Classified
    The Programmer/Recorder/Monitor (PRM) is intended to be used as part of the ZOOM LATITUDE Programming System to communicate with Boston Scientific implantable pulse generators. The
  • Z-3260-2018 · Sep 10, 2018 · Terminated
    Model 3300 LATITUDE(TM) Programming System Product Usage - The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communic
  • Z-1445-2012 · Feb 24, 2012 · Terminated
    Boston Scientific LATITUDE Patient Management System Communicator, Model 6476, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, Made in Malaysi
  • Z-1096-2012 · Dec 19, 2011 · Terminated
    LATITUDE Patient management System, Model 6488, software version 7.2. It is indicated for use to remotely communicate with a compatible Boston Scientific pulse generator and transf
  • Z-0333-2012 · Oct 12, 2011 · Terminated
    LATITUDE Patient Management System, Models 6441*, 6442*, 6465*, 6488 (USA). There are two software applications: the programmer software and the LATITUDE remote monitoring system
  • Z-0037-2012 · Aug 12, 2011 · Terminated
    LATITUDE Paceart Integration Software, version 1.00 Model 6472. LATITUDE Paceart Integration is software and therefore does not have any serial numbers. However, the software ver
  • Z-1631-2010 · Sep 10, 2009 · Terminated
    Boston Scientific, LATITUDE Communicator, Distributed by Boston Scientific, 4100 Hamline Avenue North, St. Paul, MN 55112. Models 6476 (USA) and 6468 (OUS)
  • Z-0002-2007 · May 19, 2006 · Terminated
    Guidant Zoom Latitude Programming System, Model 3120, Programmer/Recorder/Monitor (PRM). A portable cardiac rhythm management system designed to be used with Guidant implantable p

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →

Sources and updates