EMBLEM™ S-ICD Programmer 3200Recalls & reported problems
Boston Scientific Corporation · EMBLEM S-ICD Programmer · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00802526586507), received between May 5, 2020 and Jun 7, 2022 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (2 reports).
11 other recalls cover devices cleared under the same FDA submission (P110042); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Defibrillation/Stimulation Problem2
- Application Program Problem1
- Communication or Transmission Problem1
- Data Problem1
- Device Displays Incorrect Message1
- Failure to Power Up1
- Interrogation Problem1
- Noise, Audible1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Code Available2
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 7, 2022Use of Device Problem; Application Program Problem; Interrogation Problem; Use of Device Problem; Application Program Problem; Interrogation Problem
IT WAS REPORTED THAT THIS PROGRAMMER COULD NOT INTERROGATE THE PULSE GENERATOR. THE PROGRAMMER HAD AN ERROR MESSAGE. TECHNICAL SERVICES SUSPECTS THIS IS DUE TO THE PROGRAMMER HAVING OLDER SOFTWARE THAN THE DEVICE. A LOCAL REPRESENTATIVE WILL BE CONTACTED TO GET THE LATEST SOFTWARE. THERE WERE NO ADVERSE PATIENT EFFECTS. THE DEVICE REMAINS IMPLANTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14628505 ↗InjuryDec 19, 2020Defibrillation/Stimulation Problem; Device Displays Incorrect Message; Communication or Transmission Problem; Defibrillation/Stimulation Problem; Device Displays Incorrect Message; Communication or Transmission Problem
IT WAS REPORTED THAT THIS PATIENT IMPLANTED WITH A SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (S-ICD) UNDERWENT A VENTRICULAR TACHYCARDIA (VT) ABLATION PROCEDURE. DURING THE PROCEDURE, THE PATIENT WENT INTO VENTRICULAR FIBRILLATION (VF). A HEALTH CARE PROFESSIONAL (HCP) ATTEMPTED TO DELIVER A SHOCK THROUGH THE IMPLANTED DEVICE BY PRESSING THE SHOCK BUTTON ON THIS PROGRAMMER. HOWEVER, THE PROGRAMMER DISPLAYED A MESSAGE THAT THERE WERE NO STORED S-ECGS FOR THE PATIENT AND SHOCK THERAPY WAS NOT DELIVERED. EXTERNAL RESCUE WAS DELIVERED TO TERMINATE THE RHYTHM. FURTHER REVIEW WAS PERFORMED BY A LOCAL FIELD REPRESENTATIVE AND DETERMINED THAT THE PROGRAMMER LOST TELEMETRY WITH THE S-ICD AND RESULTED IN THE INABILITY TO DELIVER THE SHOCK COMMAND TO THE S-ICD. THIS PRODUCT REMAINS IN SERVICE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. IT WAS REPORTED THAT THE PROGRAMMER WAS LATER RETURNED WITH NO ADDITIONAL INFORMATION RELATED TO THE PREVIOUS REPORTED CLINICAL OBSERVATIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Code Available
MAUDE report 11042762 ↗InjuryMay 5, 2020Failure to Power Up; Defibrillation/Stimulation Problem; Data Problem; Noise, Audible; Failure to Power Up; Defibrillation/Stimulation Problem; Data Problem; Noise, Audible
THIS SUPPLEMENTAL REPORT IS BEING FILLED TO INCLUDE DEVICE ANALYSIS INFORMATION.
Patient problems: No Code Available
MAUDE report 10027211 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P110042. One submission often covers many models, so check the product and lot details in each record.
- Z-2441-2023 · Jul 11, 2023 · Open, ClassifiedEMBLEM S-ICD Pulse Generator Model A209; Subcutaneous Implantable Cardioverter Defibrillator
- Z-2442-2023 · Jul 11, 2023 · Open, ClassifiedEMBLEM MRI S-ICD Pulse Generator Model A219; Subcutaneous Implantable Cardioverter Defibrillator
- Z-0151-2023 · Sep 30, 2022 · Open, ClassifiedEMBLEM S-ICD, Subcutaneous Implantable Defibrillators, Model A209, sterile.
- Z-0152-2023 · Sep 30, 2022 · Open, ClassifiedEMBLEM MRI S-ICD, Subcutaneous Implantable Defibrillator, Model A219, sterile.
- Z-0474-2019 · Nov 1, 2018 · TerminatedThe SQ-RX Pulse Generator a component of the Subcutaneous Implantable Cardioverter Defibrillator (S-ICD) System The S-ICD System is intended to provide defibrillation therapy for
- Z-3038-2017 · Jun 29, 2017 · TerminatedEMBLEM MRI S-ICD Model 219, Subcutaneous-Implantable Cardioverter Defibrillator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the trea
- Z-3039-2017 · Jun 29, 2017 · TerminatedSQ-RX 1010 Pulse Generator, Rx. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias i
- Z-3037-2017 · Jun 29, 2017 · TerminatedEMBLEM S-ICD, Model A209, Subcutaneous Implantable Cardioverter Defibrillator, RX. Product Usage: The S-ICD system is intended to provide defibrillation therapy for the treatmen
- Z-1178-2017 · Jan 12, 2017 · TerminatedBoston Scientific, EMBLEM S-ICD Programmer Model 3200 Programmer.
- Z-0232-2016 · Sep 17, 2015 · TerminatedBoston Scientific SQ-RX Model 1010 subcutaneous pulse generator, (S-ICD). Intended to provide defibrillation therapy.
- Z-0965-2013 · Mar 1, 2013 · TerminatedBoston Scientific/Cameron Health SQ-RX Model 1010 subcutaneous pulse generators. The S-ICD System is intended to provide therapy for the treatment of life-threatening ventricular t
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page