Skip to content

Cardioplegia Pack (UDI 00803622150548)Recalls & reported problems

Livanova USA, Inc.

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00803622150548), received between Sep 13, 2023 and Sep 13, 2023 — classified as 1 malfunction.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2023

    UPON PRIMING THE PUMP AND PERFORMING THE PRE-BYPASS CHECKLIST, CARDIOPLEGIA CIRCUIT WAS PRESSURE CHECKED PER PROTOCOL. UPON BUILDING PRESSURE, THE BOTTOM OF THE HEATER/ COOLER EXCHANGER WAS LEAKING. COMPONENT WAS CHANGED OUT, NEW CIRCUIT WAS PRIMED AND CHECKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17735369 ↗