EPOC BGEM CREA TEST CARD CT-1006-00-00Recalls & reported problems
Epocal Inc · EPOC BGEM CREA TEST CARD · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00809708072254), received between Dec 1, 2017 and Dec 1, 2017 — classified as 1 death.
The most frequently reported specific device problem is “Device Operational Issue” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 19, 2025Class IIOpen, Classified
The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
Z-1685-2025 · Root cause: Process control · Recall details · FDA record ↗
- Jun 1, 2023Class IIOpen, Classified
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Z-2477-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Operational Issue1
- Use of Device Problem1
Patient problems
- Death1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
DeathDec 1, 2017Use of Device Problem; Device Operational Issue; Use of Device Problem; Device Operational Issue
IT WAS REPORTED TO EPOCAL ON (B)(6) 2017 (CASE NUMBER 00973317) THAT TWO "IQC SAMPLE DELIVERY" AND ONE "IQC INSUFFICIENT SAMPLE DETECTED" ERROR MESSAGES WERE DISPLAYED DURING TESTING ON A PATIENT EXPERIENCING A CODE MEDICAL EMERGENCY. THE PATIENT PASSED AWAY TWO DAYS LATER DUE TO THE CARDIORESPIRATORY EVENT THAT OCCURRED ON (B)(6) 2017. THE INITIAL REPORTER STATED: "POINT-OF-CARE COORDINATOR CALLED TO ASK THE CAUSES OF TWO TYPES OF IQC FAILURES: SAMPLE DELIVERY AND INSUFFICIENT SAMPLE DETECTED. THESE FAILURES BEGAN (B)(6) 2017 AND CUSTOMER WENT TO THE ICU TO RUN SOME SAMPLES AS PRELIMINARY TROUBLESHOOTING. CUSTOMER ALSO RAN QC WHICH RESULTED FINE. DURING THE PHONE CALL, CUSTOMER MENTIONED EPOC SAMPLE DELIVERY IQC FAILURES OCCURRED DURING PATIENT CODE ON (B)(6) 2017. CALLER WAS NOT THE TEST OPERATOR. TWO OPERATORS RAN THE EPOC SYSTEM. -PATIENT ID:(B)(6). -PATIENTS' AGE AND MEDICAL HISTORY: DATE OF BIRTH=(B)(6), MALE, (B)(6), DOWN'S SYNDROME, HEART FAILURES AND DEFECTS. -PATIENTS' MEDICAL SYMPTOMS? NOT BREATHING (TIME NOTED WAS NOT PROVIDED). CHRONOLOGY: (B)(6) 2017. A CODE WAS CALLED. THE CUSTOMER ESTIMATED THE TIME TO BE 13:55, BUT THEY COULDN'T CONFIRM: 1:55PM: THE NC WAS REMOVED.
Patient problems: Death
MAUDE report 7076683 ↗
Other Electrode, Ion Specific, Potassium devices
- FlexLab FLXInpeco SA · Electrode, Ion Specific, Potassium8 recalls17 reports
- Aptio Automation AP2Inpeco SA · Electrode, Ion Specific, Potassium5 recalls6 deaths16 reports
- Accelerator a3600 ACPInpeco SA · Electrode, Ion Specific, Potassium5 recalls13 reports
- FlexLab X FXXInpeco SA · Electrode, Ion Specific, Potassium2 recalls2 reports
- cobas b 221 <6> system 3337154001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM1 recall43 reports
- EPOC READER HR-1002-00-00Epocal Inc · Electrode, Ion Specific, Potassium11 deaths88 reports
- cobas b 221 <4> system 3337138001Roche Diagnostics GmbH · ELECTRODE, ION SPECIFIC, POTASSIUM1 recall7 reports
- HOST 2, EPOC HH-1009-00-00Epocal Inc · Electrode, Ion Specific, Potassium1 recall1 death7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page