Skip to content

FlareHawk CMP-00897Recalls & reported problems

Integrity Implants Inc. · Shim, 10mm or 13mm x 30mm, 0 Degree, FlareHawk 7 · Class 2 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810004721616), received between Apr 7, 2025 and Apr 7, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Activation Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2025Activation Failure; Activation Failure

    ON (B)(6) 2025, IT WAS REPORTED THAT A DOCTOR HAD INITIALLY DETERMINED THAT A FLAREHAWK SHIM AND SHELL WERE PROPERLY LOCKED. THE DOCTOR INDICATED HE HEARD A LOUD POP WHEN DEPLOYING THE SHIM INTO THE SHELL. HE USED THE LOCK GAUGE, AND THE GUIDE PROTRUDED OUT THE BACK OF THE LOCK GAUGE. HE BELIEVED THE IMPLANT WAS LOCKED BASED UPON THE READING. WHEN HE REMOVED THE GUIDE PIN THE THREADS OF THE CORE (PART OF THE SHELL) WERE INSIDE THE GUIDE PIN. UPON SUBSEQUENT REVIEW OF THE PATIENT X-RAYS, IT IS BELIEVED THAT THE SHIM AND SHELL ARE NOT LOCKED AS INTENDED. THERE WAS A DELAY OF 10 MINUTES. THE PATIENT IS DOING WELL AND WILL CONTINUE TO BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21785650 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.