Skip to content

AccuShape® ZB-SPK40Recalls & reported problems

Medcad · PEEK Cranial Implant, Extra Large · Class 2 · Implant

All models of AccuShape (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810007630632), received between Aug 26, 2024 and Aug 11, 2026 — classified as 3 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Medical device site infection1
  • No Clinical Signs, Symptoms or Conditions1
  • Post Operative Wound Infection1
  • Unspecified Infection1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 11, 2026Adverse Event Without Identified Device or Use Problem

    Explant of cranial device.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50208228 ↗
  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CRANIAL DEVICE EXPLANT.

    Patient problems: Medical device site infection

    MAUDE report 22741739 ↗
  • InjuryAug 26, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IMPLANT WAS REMOVED AS A RESULT OF A PATIENT INFECTION.

    Patient problems: Post Operative Wound Infection; Unspecified Infection

    MAUDE report 20064839 ↗

FDA clearance and classification

Cleared through 510(k) K110684 on Jun 24, 2011 (decision: Substantially Equivalent). Submitted by Vanduzen Dba Medcad as “Medcad Accushape (Tm) Peek Patient Specific Cranial / Craniofacial Implant”. The FDA review took 105 days. FDA record ↗

The FDA regulates product code GXN (plate, cranioplasty, preformed, non-alterable) under 21 CFR 882.5330 as a Class II device. 21 CFR 882.5330 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AccuShape® ZB-SPK40

Has the AccuShape® ZB-SPK40 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00810007630632), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AccuShape® ZB-SPK40?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 26, 2024 and Aug 11, 2026. In those reports, the most common patient problems are “Medical device site infection” (1), “Post Operative Wound Infection” (1) and “Unspecified Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AccuShape® ZB-SPK40?

None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AccuShape® ZB-SPK40 FDA approved?

It was cleared through 510(k) K110684 on Jun 24, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the AccuShape® ZB-SPK40 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the AccuShape® ZB-SPK40, and is it still sold?

The FDA's device database lists Medcad as the company responsible for it (brand AccuShape®, model ZB-SPK40). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Plate, Cranioplasty, Preformed, Non-Alterable devices

All Plate, Cranioplasty, Preformed, Non-Alterable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page