SSPC3 Delivery Catheter 9183Recalls & reported problems
Centerpoint Systems LLC · Catheter and Dilator · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810024671830), received between Jul 5, 2024 and Sep 30, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
- Perforation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this accessory guidewire exhibited perforation during procedure. Which possibly caused by the guiding being directed toward the hinge side. This accessory was out of service. No additional adverse patient effects were reported.
Patient problems: Perforation
MAUDE report 50945912 ↗MalfunctionFeb 22, 2026Material Integrity Problem
IT WAS REPORTED THAT THIS ACCESSORY GUIDING CATHETER EXHIBITED A CURVE FORMATION FAILURE WHICH MIGHT BE DUE TO HEAT RELATED ANOMALY OR REPEATED REPOSITIONING WHICH THE LEAD COULD NO LONGER BE DIRECTED TANGENT TOWARD THE SEPTUM AND APPEARED UNABLE TO REACH THE SITE WHERE PLACEMENT WAS INITIALLY ATTEMPTED. THIS ACCESSORY WAS OUT OF SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24410961 ↗InjuryDec 19, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE CHECK VALVE COMPONENT OF THIS ACCESSORY GUIDE CATHETER DETACHED AND REMAINED WHEN SECURING THE RIGHT VENTRICULAR LEAD TO THE SUCTION SLEEVE. THE ATTENDING PHYSICIAN VERIFIED THE DETACHED COMPONENT TO BE SIMILAR WITH THE TRANSLUCENT RING SHAPED PART FROM A NEW GUIDE CATHETER. THE INVOLVED RV LEAD DID NOT EXHIBIT LEAD DISLODGEMENT WITH NORMAL LEAD DATA MEASUREMENTS AS REPEATEDLY CHECKED WITH NO SIGNIFICANT CHANGES. THIS ACCESSORY WAS OUT OF SERVICE AND WILL BE RETURNED FOR ANALYSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 23855602 ↗MalfunctionJul 5, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THIS CATHETER WAS DISCONTINUED TO USE DUE TO HEMOSTASIS VALVE BROKE AND THE BACK FLOW OF BLOOD COULD NOT BE STOPPED. THE LEAD FELL OFF WHILE SLITTING THE CATHETER. PHYSICIAN STATED THAT THE CATHETER'S HEMOSTASIS VALVE WAS STIFFER THAN USUAL SO CUTTING. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19684505 ↗
Other Catheter, Percutaneous devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page