Skip to content

FlareHawk FHTAS100299SRecalls & reported problems

Integrity Implants Inc. · Shim, 8mm or 11mm x 29 mm, 0 degree, FH9 Sterile · Class 2 · Implant

All models of FlareHawk (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810045530499), received between Oct 25, 2023 and Oct 25, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 25, 2023Migration; Migration

    ON (B)(6) 2023 ACCELUS WAS INFORMED OF A REVISION SURGERY TO REMOVE AN UNLOCKED FLAREHAWK 9 SHIM FROM A TIHAWK 9 IMPLANTED CAGE. THE INITIAL SURGERY TOOK PLACE ON (B)(6) 2023. IT WAS LATER REPORTED THAT THE SHIM WAS UNLOCKED AND MIGRATING FROM THE SHELL. DURING THE REVISION PROCEDURE ON (B)(6) 2023, THE SURGEON MADE INCISION ON LEFT SIDE OF PATIENT AND DISSECTED DOWN TO THE L5-S1 POSTERIOR JOINT ON LEFT SIDE OF PATIENT. RETRACTION OF THE TISSUE WAS PERFORMED OVER LEFT SIDE L5-S1. THE SURGEON THEN REMOVED THE SHIM WITH THE USE OF PITUITARIES. THE SURGEON WAS ADVISED BY VINCENT FRAZIER (SALES REP) TO REMOVE THE SHELL PER THE SURGICAL TECHNIQUE MANUAL; HOWEVER, THEY CHOSE TO LEAVE THE SHELL IN PLACE AT L5-S1. IN-DEPTH ASSESSMENT OF EXITING NERVE ROOTS AND SPINAL WERE DONE, THE PROCEDURE WAS THEN COMPLETED WITHOUT ISSUE. REPORTEDLY, THE PATIENT WAS NOT EXPERIENCING ANY SYMPTOMS AS A RESULT OF THE MIGRATED SHIM. NO ADDITIONAL PATIENT HARM WAS REPORTED DURING THE REVISION SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18007440 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page