Integrity Implants Inc. · Shim, 10mm or 12mm x 29 mm, 6 degree, FH9 Sterile · Class 2 · Implant
All models of FlareHawk (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810045530536), received between Jun 16, 2025 and Jun 16, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2025Malposition of Device; Malposition of Device
DURING A SURGICAL CASE, VINCENT FRAZIER CONTACTED ME (HEATHER CAIN) INTRAOPERATIVELY TO REVIEW AN IMAGE OF AN IMPLANTED DEVICE AND TO CONFIRM WHETHER THE IMPLANT WAS PROPERLY LOCKED. UPON REVIEWING THE IMAGE, I NOTED THAT THE IMPLANT WAS NOT LOCKED AND APPEARED TO BE ROTATED. VINCENT COMMUNICATED THAT THE GUIDE PIN HAD BROKEN OFF DURING THE EXPANSION PHASE. I ADVISED THE TEAM TO ATTEMPT LOCKING THE IMPLANT. THE SURGEON PROCEEDED BY PLACING THE LOCK GAUGE INTO THE POSTERIOR ASPECT OF THE IMPLANT AND MALLETING THE BACK OF THE SHIM TO SECURE IT WITHIN THE SHELL. DESPITE THESE EFFORTS, SUBSEQUENT IMAGES CONFIRMED THAT THE IMPLANT REMAINED UNLOCKED. GIVEN THE ROTATION AND FAILURE TO LOCK, I RECOMMENDED REMOVING THE IMPLANT. THE SURGICAL TEAM WAS ABLE TO SUCCESSFULLY EXTRACT THE SHIM USING THE SHIM REMOVAL TOOL. HOWEVER, THE SURGEON DETERMINED THAT IT WOULD BE SAFER TO LEAVE THE SHELL IN PLACE AND NOT ATTEMPT REMOVAL. IT WAS ALSO NOTED THAT THE SURGEON HAD SHAVED THE VERTEBRAL BODY DOWN TO 11 MM IN HEIGHT AND WAS ATTEMPTING TO PLACE A 10 MM IMPLANT. THERE WAS A 30 MINUTE DELAY. IT WAS REPORTED IN THE COMPLAINT AS YES THERE WAS IMPACT TO PATIENT BECAUSE OF THE 30 MINUTE DELAY BUT IN THE E
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22216481 ↗
Other Intervertebral Fusion Device With Bone Graft, Lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page