Skip to content

FlareHawk FHPACT0925TTSRecalls & reported problems

Integrity Implants Inc. · Shell, Lordotic 25 mm, TiH9 Sterile · Class 2 · Implant

All models of FlareHawk (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810045530918), received between Sep 11, 2024 and Sep 11, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2024Migration; Migration

    ON 08/23/2024, AFTER REACHING OUT TO A REP TO CLARIFY WHY A SHIM AND SHELL WERE LISTED AS WASTED ON THE CHARGE SHEET, THE REP HAD REPORTED THAT THE SURGEON TOOK OUT THE SHIM AND LEFT THE SHELL IN THE PATIENT. THREE FOLLOW UPS HAVE BEEN MADE WITH THE REP TO GET THE COMPLAINT FORM POPULATED. NO COMPLAINT FORM WAS COMPLETED, BUT A PHONE CALL WITH THE REP PROVIDED THE FOLLOWING INFORMATION. ON A FOLLOW UP CALL WITH THE REP, EVAN GORGES, IT WAS REPORTED THAT THE SURGEON DIDN'T CHECK A/P BEFORE INSERTING THE SHIM. WHEN THEY EXPANDED THE IMPLANT, AFTER THE SHIM HAD LOCKED INTO THE SHELL THE SURGEON DETERMINED THE CONSTRUCT WENT TOO FAR FORWARD AND THE SURGEON DIDN'T LIKE PLACEMENT. DURING THE REMOVAL, THE SHIM WAS REMOVED AND WHILE TRYING TO USE THE SHELL REMOVAL TOOL, THE SURGEON PULLED THE SHELL BACK TO CENTER AND NOTED THE SHELL PROVIDED DISC HEIGHT AS IT WAS REPORTED THE PATIENT HAD A COLLAPSED DISC SPACE AND A LORDOTIC SHELL WAS USED. THE SURGEON THEN OPTED TO LEAVE THE SHELL IN THE PATIENT AND USED THE SHELL AS A STATIC AFTER THE SURGEON GOT THE SHELL AT THE MIDLINE. THE SURGEON HAD NO CONCERNS WITH PATIENT AND HE BELIEVES THE SHELL IS ACTING AS INTENDED AND LEAVING IT IN WAS THE BE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20194407 ↗

Other Intervertebral Fusion Device With Bone Graft, Lumbar devices

All Intervertebral Fusion Device With Bone Graft, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page