INOmax 10095Recalls & reported problems
Mallinckrodt Manufacturing LLC · Evolve System Assembly US · Class 2
All models of INOmax (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810055280063), received between Mar 27, 2025 and Apr 21, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Low Readings” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Low Readings3
- Failure to Deliver1
- Intermittent Communication Failure1
Patient problems
- Low Blood Pressure/ Hypotension2
- Low Oxygen Saturation2
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 21, 2025Low Readings; Low Readings
THE CUSTOMER CONTACTED MALLINCKRODT TO REPORT THEY EXPERIENCED INJECTOR MODULE REVERSE FLOW - HIGH PRIORITY ALARM WITH THEIR INOMAX EVOLVE DS DEVICE WHILE A PATIENT WAS RECEIVING INHALED NITRIC OXIDE (NO) THERAPY. THE CUSTOMER REPORTED THE ADULT PATIENT WAS BEING TRANSITIONED FROM THE HAMILTON G5 VENTILATOR TO THE HAMILTON T1 VENTILATOR IN PREPARATION FOR INTRA-HOSPITAL TRANSPORT. THE CUSTOMER STATED INOMAX THERAPY WAS INITIATED ON (B)(6) 2025, IN CONJUNCTION WITH HAMILTON G5 VENTILATOR WITH A SET NO DOSE OF 20PPM. THE CUSTOMER STATED THAT THE EVOLVE DEVICE WAS FULLY OPERATIONAL WHEN USED IN CONJUNCTION WITH THE HAMILTON G5. HOWEVER, THE CUSTOMER REPORTED SOON AFTER THE PATIENT WAS PLACED ON THE HAMILTON T1 VENTILATOR WITH SETTINGS OF PC-AC, PRESSURE CONTROL OF 14, RESPIRATORY RATE OF 20, PEEP +12, INSPIRATORY TIME OF 0.8 SECONDS, FLOW TRIGGER OF 5 L/MIN, AND FIO2 OF 30%, THE EVOLVE DEVICE ALARMED FOR INJECTOR MODULE REVERSE FLOW. THE CUSTOMER REPORTED THE PATIENT'S OXYGEN SATURATION DECREASED FROM 95% TO 84% AND THE MAP (MEAN ARTERIAL PRESSURE) DECREASED FROM 63 MMHG TO 48 MMHG "IMMEDIATELY". THE CUSTOMER STATED AS SOON AS THE PATIENT DECOMPENSATED, THEY PLACED THE PATIENT BACK ON
Patient problems: Low Blood Pressure/ Hypotension; Low Oxygen Saturation
MAUDE report 21877708 ↗InjuryApr 21, 2025Low Readings; Low Readings
THE CUSTOMER CONTACTED MALLINCKRODT TO REPORT THEY EXPERIENCED AN INJECTOR MODULE REVERSE FLOW - HIGH PRIORITY ALARM WITH THEIR INOMAX EVOLVE DS DEVICE WHILE A PATIENT WAS RECEIVING INHALED NITRIC OXIDE (NO) THERAPY. THE PATIENT WAS INITIATED ON INOMAX ON (B)(6) 2025 WITH A SET DOSE OF 20PPM. THE PATIENT WAS VENTILATED USING A MAQUET SERVO-U VENTILATOR IN PC/SIMV MODE WITH A RR40, PC18, PS14, PEEP+7, FIO2 OF 50%. THE BASELINE O2 SATURATION WAS BETWEEN 95% AND 98% VIA PULSE OXIMETRY. THE RETURNED TIDAL VOLUME WAS CONSISTENTLY AT APPROXIMATELY 9ML. THE PATIENT WAS NOT BREATHING ABOVE THE VENTILATOR AT THE TIME. THE CUSTOMER REPORTED THAT WHEN THEY TRANSITIONED THE PATIENT FROM THE INOMAX DSIR PLUS DELIVERY SYSTEM TO THE INOMAX EVOLVE DS WHEN THE INJECTOR MODULE REVERSE FLOW - HIGH PRIORITY ALARM AND LOW NO - ALARM OCCURRED. DURING THIS TIME, THE PATIENT EXPERIENCED OXYGEN DESATURATION, AND THE PATIENT WAS PLACED BACK ONTO THE DSIR PLUS DEVICE THAT HAD REMAINED AT THE BEDSIDE. THE CUSTOMER REPORTED THE PATIENT'S OXYGEN SATURATION DECREASED FROM BASELINE TO A LOW OF 50% DURING THIS EPISODE. THE CUSTOMER REPORTED THE STAFF DID NOT BAG THE PATIENT; HOWEVER, INCREASED THE FIO2 ON THE VENT
Patient problems: Low Oxygen Saturation
MAUDE report 21877485 ↗InjuryMar 27, 2025Failure to Deliver; Low Readings; Intermittent Communication Failure; Failure to Deliver; Low Readings; Intermittent Communication Failure
THE CUSTOMER CONTACTED MALLINCKRODT TO REPORT THEY EXPERIENCED A PRIMARY DELIVERY FAILURE ALARM WITH THEIR INOMAX EVOLVE DS DEVICE WHILE A PATIENT WAS RECEIVING INHALED NITRIC OXIDE (NO) THERAPY. THE PATIENT WAS AN ADULT BEING VENTILATED WITH A HAMILTON G5 VENTILATOR WITH THE FOLLOWING VENTILATOR PARAMETERS: VC-CMV, RR 12BPM, VT 220ML, FIO2 40% AND PEEP +5 CM H2O. THE PATIENT WAS RECEIVING INOMAX FOR APPROXIMATELY 12 HOURS WITH A SET DOSE OF 10PPM. THE RESPIRATORY THERAPIST WAS CALLED INTO THE PATIENT'S ROOM FOR A LOW NO ALARM AND UPON ARRIVAL A PRIMARY DELIVERY FAILURE ALARM WAS PRESENT. THE CUSTOMER REPORTED THE PATIENT WAS HYPOTENSIVE WITH BLOOD PRESSURE MONITORED BY ARTERIAL LINE OF 70/40 WITH A MEAN ARTERIAL PRESSURE OF 40MMHG. THE RESPIRATORY THERAPIST REPORTED THEY WERE "UNSURE IF THE EINOBLENDER WOULD STILL WORK" DUE TO THE PRIMARY DELIVERY FAILURE ALARM AND LEFT TO RETRIEVE A REPLACEMENT INOMAX EVOLVE DS. THE CUSTOMER REPORTED THAT ANOTHER HEALTHCARE PROVIDER ARRIVED SHORTLY AFTER AND NOTED THE ALARM, AND THEN IMMEDIATELY ACTIVATED THE EINOBLENDER AT 10PPM AND BEGAN TO VENTILATE THE PATIENT USING THE EINOBLENDER AND A MANUAL RESUSCITATOR BAG. THE PATIENT RESPONDED QUICKLY
Patient problems: Low Blood Pressure/ Hypotension
MAUDE report 21710364 ↗
FDA clearance and classification
Cleared through 510(k) K222930 on Dec 6, 2023 (decision: Substantially Equivalent). Submitted by Mallinckrodt Manufacturing, LLC as “Evolve Nitric Oxide Delivery System”. The FDA review took 436 days. FDA record ↗
The FDA regulates product code MRN (apparatus, nitric oxide delivery) under 21 CFR 868.5165 as a Class II device. 21 CFR 868.5165 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the INOmax 10095
Has the INOmax 10095 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00810055280063), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the INOmax 10095?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 27, 2025 and Apr 21, 2025. In those reports, the most common specific device problems are “Low Readings” (3), “Failure to Deliver” (1) and “Intermittent Communication Failure” (1); the most common patient problems are “Low Blood Pressure/ Hypotension” (2) and “Low Oxygen Saturation” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the INOmax 10095?
None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the INOmax 10095 FDA approved?
It was cleared through 510(k) K222930 on Dec 6, 2023. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the INOmax 10095, and is it still sold?
The FDA's device database lists Mallinckrodt Manufacturing LLC as the company responsible for it (brand INOmax, model 10095). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Apparatus, Nitric Oxide Delivery devices
- GENOSYL® 601588Vero Biotech Inc. · APPARATUS, NITRIC OXIDE DELIVERY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page