Skip to content

Vive MOB1106BLKRecalls & reported problems

Vive Health LLC · Core Rollator · Class 1

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810113577173), received between Aug 13, 2026 and Aug 13, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Collapse” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Collapse1

Patient problems

  • Concussion1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 13, 2026Collapse

    A customer reported an injury involving the use of a rollator walker. According to the initial report, the customer was being pushed by another individual while seated on the device in a parking lot area. When the walker encountered an uneven surface or bump transitioning from the parking lot to the sidewalk, the device reportedly collapsed, causing the customer to fall backward and strike the back of their head on the asphalt. The customer also reported additional minor injuries, including scrapes, and indicated that the assisting individual sustained a leg injury during the incident.The customer sought medical evaluation in the emergency room, where imaging (X-rays and CT scan) was performed to rule out serious complications.It is important to note that communication with the customer has ceased, and no further details or clarification have been obtained since the initial report.Additionally, the product's instruction manual clearly states: "Do not push a person seated on the rollator. The rollator is a walking aid only and not to be used as a transportation device." Based on the information provided, the device appears to have been used in a manner inconsistent with its intended

    Patient problems: Concussion

    MAUDE report 50234782 ↗

Other Walker, Mechanical devices

All Walker, Mechanical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.