Triever20 21-101 (UDI 00810123710065)Recalls & reported problems
Inari Medical, Inc. · Triever20 · Class 2
All models of Triever20 (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810123710065), received between Mar 4, 2025 and Jul 1, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation2
- Difficult to Remove1
- Entrapment of Device1
- Retraction Problem1
Patient problems
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 1, 2025Retraction Problem; Material Separation; Retraction Problem; Material Separation
ON (B)(6) 2025, A PATIENT UNDERWENT DEEP VEIN THROMBOSIS (DVT) THROMBECTOMY USING INARI DEVICES. THE PATIENT'S CLOT AGE WAS ESTIMATED AT 14 DAYS. WHEN ATTEMPTING TO REMOVE THE TRIEVER 20 AFTER THE FINAL ASPIRATION, THE CATHETER BECAME STUCK IN THE PATIENT'S VESSEL. FOLLOWING MULTIPLE ATTEMPTS TO REMOVE THE DEVICE USING BALLOONING AND MEDICATION, THE DEVICE SEPARATED AND APPROXIMATELY 10-15CM OF THE CATHETER REMAINED IN THE PATIENT. THE MEDICAL TEAM WAS UNSUCCESSFUL IN REMOVING THE REMAINING PART OF THE CATHETER DURING THE PROCEDURE, HOWEVER; THE MEDICAL TEAM WAS ABLE TO REMOVE THE PIECE OF THE DEVICE THE FOLLOWING DAY.
Patient problems: Foreign Body In Patient
MAUDE report 22375105 ↗MalfunctionMar 4, 2025Entrapment of Device; Difficult to Remove; Material Separation; Entrapment of Device; Difficult to Remove; Material Separation
ON (B)(6) 2025, A 49-YEAR-OLD MALE UNDERWENT DEEP VEIN THROMBOSIS (DVT) THROMBECTOMY USING INARI DEVICES. THE PATIENT'S CLOT AGE WAS ESTIMATED AT 90 DAYS. DURING THE THROMBECTOMY, THE TREIVER20 CATHETER (T20) WAS ADVANCED WITHOUT AN INTRODUCER SHEATH INTO THE LEFT POPLITEAL VEIN. AN ASPIRATION WAS PERFORMED, AND THE CATHETER WAS RETRACTED. THE TIP OF THE CATHETER BECAME STUCK WHILE RETRACTING THE CATHETER THROUGH THE COMMON ILIAC VEIN. THE CATHETER WAS RETRACTED A FEW MORE INCHES AND THEN FRACTURED. HALF OF THE CATHETER AND THE DILATOR WERE REMOVED AND APPROXIMATELY 20 CENTIMETERS OF THE T20 DELIVERY CATHETER REMAINED IN THE PATIENT'S VESSEL. SCISSORS WERE USED TO CUT THE WIRE BRAIDING OF THE T20. AN INTRI24 INTRODUCER SHEATH (INTRI24) WAS THEN ADVANCED TO WHERE THE TIP WAS JUST OVER THE PROXIMAL (BROKEN) END OF THE T20 AND A 10X40 BALLOON WAS PLACED OVER THE WIRE AT THE DISTAL END OF THE REMAINING PORTION OF THE T20 CATHETER INTO THE INTRI24. THE INTRI24, T20 CATHETER, AND BALLOON WERE THEN REMOVED TOGETHER AND THE INTRI24 WAS FLUSHED AND REPLACED INTO THE LEFT POPLITEAL VEIN. ALL MATERIAL WAS CONFIRMED TO BE REMOVED FROM THE PATIENT AND THE PATIENT SUSTAINED NO INJURIES. THE PATI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21520940 ↗
FDA clearance and classification
Cleared through Special 510(k) K213402 on May 23, 2022 (decision: Substantially Equivalent). Submitted by Inari Medical as “Triever24, Triever20”. The FDA review took 217 days. FDA record ↗
The FDA regulates product code QEW (peripheral mechanical thrombectomy with aspiration) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Triever20 21-101
Has the Triever20 21-101 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00810123710065), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Triever20 21-101?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Mar 4, 2025 and Jul 1, 2025. In those reports, the most common specific device problems are “Material Separation” (2), “Difficult to Remove” (1) and “Entrapment of Device” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Triever20 21-101?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Triever20 21-101 FDA approved?
It was cleared through Special 510(k) K213402 on May 23, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Triever20 21-101, and is it still sold?
The FDA's device database lists Inari Medical, Inc. as the company responsible for it (brand Triever20, model 21-101). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Peripheral Mechanical Thrombectomy With Aspiration devices
- Indigo System LITNG12HTORQ115-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration4 deaths105 reports
- Indigo System LITNG7XTORQ130-APenumbra, Inc. · Peripheral mechanical thrombectomy with aspiration100 reports
- Indigo System LITNG12XTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death80 reports
- ClotTriever XL Catheter AS-0235Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 recall9 deaths21 reports
- Indigo System LITNGBT7TQ130-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration46 reports
- Indigo System LITNG12HTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death46 reports
- Triever 24 22-101Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration20 deaths34 reports
- ClotTriever Catheter, 16 mm 40-102Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration11 deaths27 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.