Skip to content

Headway Advanced MC17215045XRecalls & reported problems

Microvention, Inc. · Microcatheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00810170015496), received between Dec 19, 2023 and Dec 19, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 19, 2023Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT RUPTURE OF THE PROXIMAL THIRD OF THE MICROCATHETER AFTER COMPLETING COILING AND PULL IT BACK OUTSIDE THE ANEURYSM SAC. ADDITIONAL INFORMATION RECEIVED INDICATED, THIS IS IN RELATION TO THE ABOVE REPORTED EVENT, I WOULD LIKE TO IDENTIFY THAT THE SENT MICROCATHETER HAS BEEN USED TO TREAT SUCCESSFULLY A RIGHT PARAOPTHALMIC SACCULAR ANEURYSM THROUGH A LONG 6F SHEATH AND A 5F SOPHIA INTERMEDIATE CATHETER AND AFTER FINISHING THE COILING PROCESS AND REMOVE THE SENT HEADWAY 17 MICROCATHETER OVER THE TRAXCESS 14 WIRE OUT OF THE ANEURYSM SAC INTO THE CAROTID SIPHON, WE HAVE LOST CONTROL OF THE MICROCATHETER AS WE PULLED OUT ITS PROXIMAL PART BUT NOT MOVING THE UPPER PART UP TO THE SIPHON. AS A RESULT, THE PROXIMAL PART OF THE CATHETER WHICH WE HAVE SENT EVENTUALLY HAD BEEN PULLED OUT BUT THE UPPER PART OF IT REMAINED INTO THE ARTERIAL NETWORK. HOPEFULLY, THE PATIENT WAS AND REMAINS VERY WELL AND THE CAROTID IS FULKLY PATENT UNDER DUAL ANTIPLATELET TREATMENT AND THE 3 MONTHS ANGIOGRAM IS FINE (SEE THE ATTACHED PICTURES - RT ICA COILED RT A2 ANEURYSM TREATED AS WELL DURING THIS PROCEDURE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18361487 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page