Scepter XC BC0411XC-BRecalls & reported problems
Microvention, Inc. · Balloon Catheter · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810170018565), received between Nov 10, 2016 and Apr 12, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Rupture1
- Material Separation1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 12, 2017Material Separation; Material Separation
IT WAS REPORTED THAT AFTER POSITIONING A SCEPTER XC BALLOON IN THE COMMON CAROTID ARTERY, THE BALLOON DID NOT INFLATE. THE BALLOON CATHETER WAS REMOVED, AND WAS OBSERVED TO BE BROKEN INTO TWO (2) SEPARATE PIECES. ALL PORTIONS OF THE BALLOON WERE REMOVED FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT WAS REPORTED TO BE STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6488871 ↗MalfunctionNov 10, 2016Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURED IN THE PATIENT. THE PHYSICIAN USED ANOTHER SCEPTER TO REPLACE THE BALLOON. THE PROCEDURE WAS SUCCESSFUL. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6092712 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page