The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00810633021484), received between Feb 14, 2019 and Jun 21, 2022 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Break2
  • Device-Device Incompatibility2

Patient problems

  • No Code Available3
  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2022Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2022, HAMMER LOCK 2 IMPLANTS WERE BROKEN WITHOUT THE PACKAGE BEING OPENED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) HAERLOCK 2 IMPL KIT 12X5/0 DEG/SMALL THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14752865 ↗
  • InjuryAug 25, 2020Device-Device Incompatibility; Appropriate Device Problem Term/Code Not Available; Device-Device Incompatibility; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT ON (B)(6) 2020, DURING A HAMMERTOE CORRECTION ON DIGITS 2-5 THE INSERTER FOR THE HAMMERTOE CCI WOULD NOT RELEASE. THE FIRST INSERTER WAS TRIED AND WHEN IT DID NOT WORK, THE SURGEON DRILLED AGAIN TO MAKE SURE THAT IT WENT ALL THE WAY TO THE DRILL STOP. WHEN DRILLING DID NOT WORK, THE SURGEON LOOKED AT IT ON THE BACK TABLE AND EVENTUALLY MANAGED TO RELEASE THE IMPLANT FROM THE INSERTER. THE SURGEON THEN OPENED ANOTHER IMPLANT BUT THE NEW IMPLANT DID NOT RELEASE EITHER. WHEN THE HAMMERTOE CCI DID NOT WORK, A HAMMERLOCK 2 CCI WAS TRIED. THE SURGEON ATTEMPTED TO USE A MEDIUM HAMMERLOCK IMPLANT ON THE SECOND DIGIT AND SUCCESSFULLY INSERTED THE CCI IN THE PROXIMAL PHALANGE. HOWEVER, THE SURGEON COULD NOT GET THE INTERMEDIATE AND DISTAL PHALANGE UP AND ON THE DISTAL END OF THE CCI. A SMALL HAMMERLOCK IMPLANT WAS TRIED, AND STILL HAD THE SAME RESULTS. THE SURGEON THEN PUT A K-WIRE IN. A SMALL HAMMERLOCK 2 CCI WAS TRIED ON THE SECOND DIGIT BUT HAD THE SAME RESULTS. THE SURGEON THEN PUT K-WIRES IN THE REST OF THE DIGITS. THE SURGEON DECIDED TO USE K-WIRES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A THIRTY (30) MINUTE SURGICAL DELAY. PATIENT STATUS IS UNKNOWN CONCOMIT

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 10449390 ↗
  • InjuryAug 25, 2020Device-Device Incompatibility; Appropriate Device Problem Term/Code Not Available; Device-Device Incompatibility; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT ON (B)(6) 2020, DURING A HAMMERTOE CORRECTION ON DIGITS 2-5 THE INSERTER FOR THE HAMMERTOE CCI WOULD NOT RELEASE. THE FIRST INSERTER WAS TRIED AND WHEN IT DID NOT WORK, THE SURGEON DRILLED AGAIN TO MAKE SURE THAT IT WENT ALL THE WAY TO THE DRILL STOP. WHEN DRILLING DID NOT WORK, THE SURGEON LOOKED AT IT ON THE BACK TABLE AND EVENTUALLY MANAGED TO RELEASE THE IMPLANT FROM THE INSERTER. THE SURGEON THEN OPENED ANOTHER IMPLANT BUT THE NEW IMPLANT DID NOT RELEASE EITHER. WHEN THE HAMMERTOE CCI DID NOT WORK, A HAMMERLOCK 2 CCI WAS TRIED. THE SURGEON ATTEMPTED TO USE A MEDIUM HAMMERLOCK IMPLANT ON THE SECOND DIGIT AND SUCCESSFULLY INSERTED THE CCI IN THE PROXIMAL PHALANGE. HOWEVER, THE SURGEON COULD NOT GET THE INTERMEDIATE AND DISTAL PHALANGE UP AND ON THE DISTAL END OF THE CCI. A SMALL HAMMERLOCK IMPLANT WAS TRIED, AND STILL HAD THE SAME RESULTS. THE SURGEON THEN PUT A K-WIRE IN. A SMALL HAMMERLOCK 2 CCI WAS TRIED ON THE SECOND DIGIT BUT HAD THE SAME RESULTS. THE SURGEON THEN PUT K-WIRES IN THE REST OF THE DIGITS. THE SURGEON DECIDED TO USE K-WIRES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A THIRTY (30) MINUTE SURGICAL DELAY. PATIENT STATUS IS UNKNOWN CONCOMIT

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 10449244 ↗
  • InjuryFeb 14, 2019Break; Break

    THE PROCEDURE OCCURRED ON (B)(6) 2019. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Code Available

    MAUDE report 8339396 ↗

FDA clearance and classification

The FDA regulates product code HTY (pin, fixation, smooth) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the HammerLock2 HL2S

Has the HammerLock2 HL2S been recalled?

No FDA recall record in our data names this exact device identifier (DI 00810633021484), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the HammerLock2 HL2S?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Feb 14, 2019 and Jun 21, 2022. In those reports, the most common specific device problems are “Break” (2) and “Device-Device Incompatibility” (2); the most common patient problem is “Failure of Implant” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the HammerLock2 HL2S?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the HammerLock2 HL2S MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the HammerLock2 HL2S, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand HammerLock2, model HL2S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Pin, Fixation, Smooth devices

All Pin, Fixation, Smooth devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page