Elite EL-252013HRecalls & reported problems
Biomedical Enterprises, Inc. · ELITE 25x20x13mm, H-Shaped Bridge, 4 Legs · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810633021743), received between Sep 7, 2018 and Sep 25, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2018Break; Break
THIS REPORT IS FOR ONE (1) ELITE COMPRESSION IMPLANT 20X20X20X20X20 4 LEGS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7905193 ↗MalfunctionSep 7, 2018Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SURGEON USED ELITE H IMPLANTS FOR A MIDFOOT FUSION PROCEDURE. POSTOPERATIVE X-RAYS SHOWED TWO BROKEN STAPLES. ONE STAPLE HAD ONE LEG BROKEN AT TOP OF THE SECOND BARB IN THE TALONAVICULAR (TN) JOINT AND THE OTHER STAPLE HAD ONE BROKEN LEG AT THE SHOULDER IN THE CALCANEOCUBOID (CC) JOINT. PATIENT HAD NO COMPLAINTS OF PAIN OR ANY SYMPTOMS. THERE WAS NO REVISION PROCEDURE PERFORMED. THIS REPORT IS FOR AN ELITE 25 X 13 X 20 X 20 X 20 X 20 MM (4 LEGS) H SHAPED IMPLANT. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7856624 ↗
Other Staple, Fixation, Bone devices
- N/A StandardSynthes GmbH · Staple, Fixation, Bone2 recalls16 reports
- GII QUICKANCHOR ETHIBOND 212034Medos International Sàrl · STAPLE, FIXATION, BONE98 reports
- GII QUICKANCHOR ORTHOCORD 222983Medos International Sàrl · STAPLE, FIXATION, BONE94 reports
- Speed SE-0910Synthes GmbH · Staple, Fixation, Bone1 recall24 reports
- Speed Shift SE-2020-10Synthes GmbH · Staple, Fixation, Bone1 recall10 reports
- Lapiplasty SpeedPlate SK50Treace Medical Concepts, Inc. · Staple, Fixation, Bone59 reports
- Speed SE-1512Synthes GmbH · Staple, Fixation, Bone1 recall6 reports
- LUPINE ORTHOCORD 210709Depuy Mitek, LLC · STAPLE, FIXATION, BONE56 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page