The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00810633021996), received between Oct 25, 2021 and Aug 15, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Device-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Bone Fracture(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 15, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2024, THE COMPRESSIVE FORCE OF THE STAPLE IMPLANT CAUSED AN IATROGENIC FRACTURE OF THE PATIENT'S BIG TOE. IT WAS QUESTIONED WHETHER POOR BONE QUALITY, DRILLING TECHNIQUE, OR THE IMPLANT WAS THE CAUSE OF THE IATROGENIC FRACTURE. THE SURGEON ELECTED TO REMOVE THE STAPLE AND USED A K-WIRE FOR FIXATION INSTEAD. THE PROCEDURE WAS COMPLETED WITH A FIFTEEN MINUTE DELAY. THIS REPORT IS FOR A SPEED 11X8X8MM IMPLANT. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: Bone Fracture(s)

    MAUDE report 19995838 ↗
  • MalfunctionOct 14, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THAT THE STAPLE ITSELF HAS BROKEN AWAY FROM THE HOLD IT SITS IN WITHIN THE HANDLE ¿ THE STAPLE SHOULD NOT MOVE FROM THE LOADED HANDLE UNTIL THE TRIANGULAR RELEASE BUTTON IS PRESSED WITH FORCE WHEN IN THE PATIENT. PATIENT IMPACT ¿ NONE AS WAS NOT USED IN SURGERY. THIS REPORT IS FOR ONE (1) SPEED COMPRESSION IMPLANT KIT 11X08X08MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15605653 ↗
  • MalfunctionOct 25, 2021Break; Device-Device Incompatibility; Break; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2021 DURING THE PROCEDURE, THE SPEED COMPRESSION IMPLANT AND UNKNOWN INSERTER WERE UNABLE TO BE RELEASED FROM ONE ANOTHER. WHILE TRYING TO GET THE INSTRUMENT AND IMPLANT TO DISASSEMBLE, THE SURGEON BROKE THE CORTEX OF THE FUSION SITE. THE IMPLANT WAS REMOVED FROM THE BONE WHILE ATTACHED TO THE INSERTER. DURING REMOVAL OF THE IMPLANT, THE INSERTER WAS BROKEN IN HALF. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT OUTCOME IS UNKNOWN. THIS REPORT IS FOR (1) SPEED COMPRESSION IMPLANT KIT 11X08X08MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12693570 ↗