Skip to content

ENROUTE Transcarotid Stent SR-0730-CSRecalls & reported problems

Silk Road Medical, Inc. · 7 mm X 30 mm · Class 3 · Implant

All models of ENROUTE Transcarotid Stent (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00811311020430), received between Aug 7, 2023 and Dec 2, 2025 — classified as 7 injuries.

The most frequently reported specific device problem is “Defective Device” (2 reports).

1 other recall cover devices cleared under the same FDA submission (P140026); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Defective Device2
  • Positioning Problem1

Patient problems

  • Ischemia Stroke2
  • Stroke/CVA2
  • Thrombosis/Thrombus2
  • Low Blood Pressure/ Hypotension1
  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1
  • Vascular Dissection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A DISSECTION WAS IDENTIFIED RESULTING IN AN ADDITIONAL STENT PLACEMENT. AN ENROUTE TRANSCAROTID STENT WAS SELECTED FOR USE IN A LEFT SIDED TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE. DURING THE TCAR PROCEDURE, THE PHYSICIAN ADVANCED THE 0.014" GUIDEWIRE INTO THE INTERNAL CAROTID ARTERY (ICA), PRE-DILATED, PLACED AN ENROUTE STENT, AND POST-DILATED SUCCESSFULLY. IMAGING WAS THEN PERFORMED, AND IT REVEALED A DISSECTION JUST BEYOND THE DISTAL PORTION OF THE STENT IN THE ICA. THE PHYSICIAN ATTRIBUTED THE DISSECTION EVENT TO EITHER THE ENROUTE 0.014" GUIDEWIRE OR THE ENROUTE STENT AND ELECTED TO PLACE AN ADDITIONAL STENT TO COVER THE DISSECTION.

    Patient problems: Vascular Dissection

    MAUDE report 23695022 ↗
  • InjuryJul 2, 2025Positioning Problem; Positioning Problem

    IT WAS REPORTED THAT THE STENT WAS INADVERTENTLY DEPLOYED A COLLATERAL VESSEL THAT CAME OFF THE ICA, REQUIRING ADDITIONAL SURGICAL INTERVENTION. AN ENROUTE TRANSCAROTID STENT WAS SELECTED FOR USE IN A RIGHT SIDED TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE. THE DISTAL PORTION OF THE STENT WAS INADVERTENTLY DEPLOYED IN A COLLATERAL VESSEL THAT CAME OFF THE INTERNAL CAROTID ARTERY (ICA) AND THE PROXIMAL PORTION OF THE STENT WAS DEPLOYED IN THE ICA. THE PHYSICIAN ELECTED TO CONVERT THE PROCEDURE TO A CAROTID ENDARTERECTOMY (CEA), EXPLANT THE STENT, AND USE A POLYTETRAFLUOROETHYLENE (PTFE) GRAFT INSTEAD OF A CAROTID PATCH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22385659 ↗
  • InjuryJun 6, 2025Defective Device; Defective Device

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED STROKE-LIKE SYMPTOMS 1 WEEK POST T-CAR, AND THE STENT APPEARED NARROWED. TWO ENROUTE TRANSCAROTID STENTS WAS SELECTED FOR USE IN A RIGHT SIDED TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE. THERE WERE NO ISSUES WITH THE STENTS EXPANDING NOTED DURING THE PROCEDURE OR IN POST PROCEDURE IMAGING. ONE WEEK FOLLOWING THE TCAR PROCEDURE, THE PATIENT PRESENTED TO THE ER WITH STROKE-LIKE SYMPTOMS. THE PATIENT WAS TRANSFERRED TO A DIFFERENT FACILITY, AND THE PHYSICIAN HAD LIMITED ACCESS TO ADDITIONAL INFORMATION. THE PHYSICIAN WAS ABLE TO REPORT THAT ONE OF THE STENTS APPEARED NARROWED. NO FURTHER INFORMATION WAS AVAILABLE, DESPITE REQUEST BY BOSTON SCIENTIFIC.

    Patient problems: Stroke/CVA

    MAUDE report 22159367 ↗
  • InjuryMay 30, 2025Defective Device; Defective Device

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED STROKE-LIKE SYMPTOMS 1 WEEK POST TCAR, AND THE STENT APPEARED NARROWED. TWO ENROUTE TRANSCAROTID STENTS WERE SELECTED FOR USE IN A RIGHT SIDED TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE. THERE WERE NO ISSUES WITH THE STENTS EXPANDING NOTED DURING THE PROCEDURE OR IN POST PROCEDURE IMAGING. ONE WEEK FOLLOWING THE TCAR PROCEDURE, THE PATIENT PRESENTED TO THE ER WITH STROKE-LIKE SYMPTOMS. THE PATIENT WAS TRANSFERRED TO A DIFFERENT FACILITY, AND THE PHYSICIAN HAD LIMITED ACCESS TO ADDITIONAL INFORMATION. THE PHYSICIAN WAS ABLE TO REPORT THAT ONE OF THE STENTS APPEARED NARROWED. NO FURTHER INFORMATION WAS AVAILABLE, DESPITE REQUEST BY BOSTON SCIENTIFIC.

    Patient problems: Stroke/CVA

    MAUDE report 22114138 ↗
  • InjuryApr 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER A LEFT TRANSCAPHOID ARTERY REVASCULARIZATION (TCAR) PROCEDURE, THE PATIENT EXPERIENCED BARORECEPTOR DYSFUNCTION. THE PATIENT REMAINED AN INPATIENT FOR THREE DAYS POST PROCEDURE AND WAS GIVEN NOREPINEPHRINE UNTIL THE BARORECEPTORS NORMALIZED. IT IS UNKNOWN IF THE ENROUTE TRANSCAPHOID STENT SYSTEM (TSS) CONTRIBUTED TO BARORECEPTOR DYSFUNCTION AND THEREFORE, THIS REPORT WILL BE REPORTED OUT OF AN ABUNDANCE OF CAUTION.

    Patient problems: Low Blood Pressure/ Hypotension

    MAUDE report 19076246 ↗
  • InjuryOct 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT 9 DAYS AFTER A TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE, THE PATIENT EXPERIENCED ACUTE STROKE WITH AN OCCLUDED STENT. THROMBECTOMY WAS PERFORMED TO ADDRESS THE ISSUE. ADDITIONAL INFORMATION INDICATED THAT THE PATIENT'S CONDITION HAD NOT YET RESOLVED. AT THIS TIME, IT IS UNKNOWN IF THE REPORTED EVENT IS RELATED TO PROCEDURAL ISSUES, NON-COMPLIANCE OR RESISTANCE TO MEDICATION, OR A SILK ROAD MEDICAL DEVICE FAILURE, HENCE, THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION.

    Patient problems: Ischemia Stroke; Obstruction/Occlusion; Thrombosis/Thrombus

    MAUDE report 17873337 ↗
  • InjuryAug 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER THE COMPLETION OF A LEFT TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE, THE PATIENT HAD A STROKE WITH A LEFT-SIDED INFARCT. THE PHYSICIAN PERFORMED A THROMBECTOMY TO ADDRESS THE ISSUE. AT THIS TIME, IT IS UNKNOWN IF THE REPORTED EVENT IS RELATED TO PROCEDURAL ISSUES, PATIENT RESISTANCE OR NON-COMPLIANCE TO MEDICATION, OR A SILK ROAD MEDICAL DEVICE FAILURE, HENCE, THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION.

    Patient problems: Ischemia Stroke; Thrombosis/Thrombus

    MAUDE report 17476382 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140026. One submission often covers many models, so check the product and lot details in each record.

  • Z-1146-2021 · Jan 13, 2021 · Terminated
    ENROUTE Transcarotid Stent System REF SR-XXYY-CS

Other Stent, Carotid devices

All Stent, Carotid devices →