Skip to content

NovoHip Total Hip System 1345-2803Recalls & reported problems

Novosource, LLC · Biolox Ball Head, 28-12/14L · Class 2 · Implant

All models of NovoHip Total Hip System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00811382037818), received between Oct 9, 2024 and Oct 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Wrong Label” (1 report).

1 other recall cover devices cleared under the same FDA submission (K132158); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Wrong Label1

Patient problems

  • Joint Dislocation1
  • Joint Laxity1
  • Metal Related Pathology1
  • Osteolysis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2024Wrong Label; Wrong Label

    AN INITIAL COMPLAINT WAS RECEIVED ON SEPTEMBER 12, 2024, INDICATING THAT IT WAS DISCOVERED INTRAOPERATIVELY THAT A BOX LABELED AS CONTAINING A SIZE 28 L CERAMIC HEAD ACTUALLY CONTAINED A SIZE 32 L CERAMIC HEAD (CORRECT SIZE DIRECTLY MARKED ON THE DEVICE). NO PATIENT INJURY OR DEATH WAS REPORTED. NOVOSOURCE BECAME AWARE THAT BIOLOX CERAMIC HEADS, A COMPONENT IN THE NOVOSOURCE TOTAL HIP SYSTEM HAD BEEN MISLABELED. SIZES 28 L AND 32 L HAD SEEMINGLY BEEN REVERSED. THE PRELIMINARY INVESTIGATION INDICATED THAT THE ISSUE WAS RESTRICTED TO 2 LOTS, EACH CONTAINING A QUANTITY OF 10 UNITS. AS PART OF THE ONGOING COMPLAINT INVESTIGATION AND CONTAINMENT ACTIONS, THE SUSPECT LOT AND THE ASSOCIATED LOT (PACKAGING LABELED AS SIZE 32 L HEADS) WERE REQUESTED TO BE RETURNED FOR EVALUATION. UPON INSPECTION OF THE RETURNED UNITS, NOVOSOURCE IS AWARE THAT THE 28 L CERAMIC HEADS WERE INDEED ALSO PACKAGED WITH LABELING THAT INDICATED SIZE 32 L. THUS, THE DECISION WAS BEEN MADE THAT A FIRM-INITIATED RECALL WAS APPROPRIATE AND THE APPROPRIATE FDA DIVISION RECALL COORDINATOR WAS NOTIFIED WITHOUT DELAY. FINALLY, INFORMATION OBTAINED THROUGH A HEALTH HAZARD EVALUATION SUPPORTED THAT IF THE MALFUNCTION WERE TO

    Patient problems: Joint Dislocation; Joint Laxity; Metal Related Pathology; Osteolysis

    MAUDE report 20412105 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K132158. One submission often covers many models, so check the product and lot details in each record.

  • Z-0483-2025 · Sep 19, 2024 · Open, Classified
    NovoHip Biolox Ball Head 32mmL Ref: 1345-2803 NovoHip Total Hip System - NovoHip Total Hip System is a non-cemented hip prosthesis that consists of a 4-part total hip replacement

Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented devices →