Amplatzer™ Septal Occluder (REF 9-ASD-004)Recalls & reported problems

Aga Medical Corporation · Septal Occluder · Class 3 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00811806010014), received between Jun 15, 2020 and Mar 28, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Fold, Unfold or Collapse” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Fold, Unfold or Collapse2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    IT WAS REPORTED THAT ON (B)(6) 2025, A 4MM AMPLATZER SEPTAL OCCLUDER WAS SELECTED FOR IMPLANT. DURING THE PROCEDURE, WHEN THE DEVICE WAS DEPLOYED, IT PRESENTED IN A COBRA DEFORMATION. THE DEVICE WAS REMOVED AND NEW 4MM AMPLATZER SEPTAL OCCLUDER WAS ATTEMPTED. IT WAS FOUND TO BE TOO SMALL TO PUSH THROUGH THE DEFECT, SO IT WAS REMOVED. A NEW 5MM AMPLATZER SEPTAL OCCLUDER WAS THEN SELECTED FOR USE. WHEN ATTEMPTED, THE 5MM DEVICE WAS ALSO FOUND TO BE TOO SMALL. THERE WAS ALSO LIMITED ROOM IN LEFT ATRIUM FOR A BIGGER DEVICE. THE DECISION WAS MADE TO SURGICALLY CLOSE THE ASD. NO ADVERSE EVENTS OCCURRED WITH THIS PATIENT AND THEY REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE, THE PATIENT WAS ON ECMO PRIOR TO PROCEDURE START. THERE WAS NO CLINICALLY SIGNIFICANT DELAY DURING THE PROCEDURE. THE PATIENT IS RECOVERING IN NORMAL FASHION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21722738 ↗
  • MalfunctionJul 8, 2022Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    IT WAS REPORTED THAT ON (B)(6) 2022 A 4MM AMPLATZER SEPTAL OCCLUDER WAS IMPLANTED IN THE PATIENT. IT WAS NOTED THAT IT HAD DEFORMED WHILE IN THE PATIENT, SO IT WAS REMOVED AND REPLACED WITH ANOTHER 4MM AMPLATZER SEPTAL OCCLUDER. THE PATIENT REMAINED HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14961936 ↗
  • MalfunctionJun 15, 2020Material Deformation; Material Deformation

    ON (B)(6) 2020 AN AMPLATZER SEPTAL OCCLUDER WAS SELECTED FOR AN IMPLANT. THE DEVICE FORMED A COBRA FORMATION WHILE DEPLOYING. THE PHYSICIAN REPLACED THE DEVICE WITH A COMPETITORS DEVICE TO COMPLETE THE PROCEDURE. THE PATIENT WAS STABLE THROUGH AND POST THE PROCEDURE. PATIENT WAS HEMODYNAMICALLY STABLE THROUGHOUT THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10155397 ↗