AMPLATZER™ 9-ITV08F180/80Recalls & reported problems

Aga Medical Corporation · Delivery System · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00811806010823), received between May 23, 2023 and Aug 6, 2024 — classified as 1 death, 2 malfunctions.

The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device1
  • Material Twisted/Bent1
  • Positioning Problem1
  • Premature Separation1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Cardiac Tamponade1
  • Pericardial Effusion1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 6, 2024Premature Separation; Premature Separation

    IT WAS REPORTED THAT A 14-12MM AMPLATZER DUCT OCCLUDER WAS SELECTED FOR IMPLANT ON (B)(6) 2024, USING AN 8F AMPLATZER TORQVUE DELIVERY SYSTEM. THE DEFECT WAS MEASURED AS 7MM IN DIAMETER AND 9MM IN LENGTH VIA TRANSESOPHAGEAL ECHOCARDIOGRAPHY (TEE). THE DEVICES WERE PREPARED PER IFU. ECHOCARDIOGRAM AND FLUOROSCOPY WERE USED DURING THE PROCEDURE. DURING IMPLANT, THE CABLE CONNECTION WAS CHECKED PRIOR TO DEPLOYMENT. WHILE DEPLOYING THE DEVICE, IT WAS OBSERVED THROUGH FLUOROSCOPY THAT THE DEVICE PREMATURELY DETACHED FROM THE DELIVERY CABLE. THE DEVICE EMBOLIZED INTO THE DESCENDING AORTA. THE DEVICE WAS SNARED VIA TRANSCATHETER SNARE AND REMOVED FROM THE PATIENT. A NEW 16-14MM AMPLATZER DUCT OCCLUDER AND 8F AMPLATZER TORQVUE DELIVERY SYSTEM WERE USED TO COMPLETE THE PROCEDURE. THE PATIENT DID NOT PRESENT WITH ANY CLINICAL SIGNS OR SYMPTOMS. THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19924501 ↗
  • MalfunctionMay 24, 2023Retraction Problem; Device Damaged by Another Device; Material Twisted/Bent; Retraction Problem; Device Damaged by Another Device; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6)2023, AN 8MM AMPLATZER MUSCULAR VSD OCCLUDER WAS CHOSEN FOR IMPLANT VIA FEMORAL VEIN APPOACH WITH A 6F AMPLATZER TORQVUE DELIVERY SYSTEM. DURING DEPLOYMENT, THE 8MM DEVICE APPEARED TOO SMALL AND THE DEPLOYMENT ANGLE WAS NOT IDEAL SO THE 8MM DEVICE WAS REPLACED PRIOR TO RELEASE WHILE INSIDE THE PATIENT. A 10MM AMPLATZER MUSCULAR VSD OCCLUDER WAS THEN CHOSEN FOR IMPLANT WITH THE SAME 6F AMPLATZER TORQVUE DELIVERY SYSTEM. DURING DEPLOYMENT, THE 10MM OCCLUDER ALSO APPEARED TOO SMALL AND THE DEPLOYMENT ANGLE WAS ALSO NOT IDEAL. A DECISION WAS MADE TO REPLACE BOTH THE 10MM OCCLUDER AND 6F DELIVERY SYSTEM. IT WAS THEN REPORTED A 12MM AMPLATZER MUSCULAR VSD OCCLUDER WAS CHOSEN FOR IMPLANT VIA INTERNAL JUGULAR APPROACH WITH A 7F AMPLATZER TORQVUE DELIVERY SYSTEM. DURING DEPLOYMENT, THE 12MM OCCLUDER WAS NOT ABLE TO BE RELEASED FROM THE DELIVERY CABLE. BOTH 12MM OCCLUDER AND 7F DELIVERY SYSTEM WERE REMOVED FROM THE PATIENT AND THE 12MM OCCLUDER WAS REMOVED FROM THE DELIVERY CABLE WITHOUT DIFFICULTY. A DECISION WAS MADE TO IMPLANT THE SAME 12MM OCCLUDER WITH AN 8F AMPLATZER TORQVUE DELIVERY SYSTEM. DURING DEPLOYMENT, DUE TO THE DEPLOYMENT ANGLE NOT BEING IDEAL, T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16998318 ↗
  • DeathMay 23, 2023Positioning Problem; Positioning Problem

    IT WAS REPORTED THAT ON (B)(6) 2023, A 16-14MM AMPLATZER DUCT OCCLUDER WAS CHOSEN FOR PROCEDURE USING A 7F AMPLATZER TORQVUE DELIVERY SYSTEM FOR A PATIENT WITH A POST-INFARCT VENTRICULAR SEPTAL DEFECT (VSD). THE PATIENT HAD A MYOCARDIAL INFARCTION ON (B)(6) 2023. THE PHYSICIAN MADE THE DECISION TO CHOSE AN AMPLATZER DUCT OCCLUDER DUE TO THE VSD LOCALIZATION AT THE APEX OF THE HEART. HEPARIN WAS ADMINISTERED DURING PROCEDURE. THE OCCLUDER WAS UNABLE TO BE INTRODUCED INSIDE THE DELIVERY SYSTEM. THE DEVICE WAS BLOCKED AT THE ENTRANCE OF THE DELIVERY SYSTEM. AFTER A FEW ATTEMPTS, THE DECISION WAS MADE TO REPLACE THE DELIVERY SYSTEM WITH AN 8F AMPLATZER TORQVUE DELIVERY SYSTEM. THERE WERE MULTIPLE MANIPULATIONS TO FIND THE OPTIMAL POSITION OF THE OCCLUDER. THE OCCLUDER RECAPTURED ONCE BEFORE IT WAS IMPLANTED. A PERICARDIAL EFFUSION WAS NOTED AT THE APEX OF THE HEART. THE ACTIVATED CLOTTING TIME (ACT) LEVEL WAS 360 SECONDS. A PERICARDIOCENTESIS WAS PERFORMED WITH A REDUCTION TO THE EFFUSION. HOWEVER, THE PERICARDIAL EFFUSION THEN DEVELOPED INTO CARDIAC TAMPONADE AND WAS UNABLE TO BE REVERSED. THE PATIENT DIED AT THE END OF THE PROCEDURE. THE PERICARDIAL EFFUSION WAS BELIEVED TO BE DUE TO

    Patient problems: Cardiac Tamponade; Pericardial Effusion

    MAUDE report 16989583 ↗