Skip to content

Amplatzer™ TorqVue™ 9-TVLP4F90/080Recalls & reported problems

St. Jude Medical, Inc. · LP Delivery System · Class 2

All models of Amplatzer TorqVue (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00811806011578), received between Jul 14, 2025 and Oct 29, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 17, 2025Class IIOpen, Classified

    Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

    Z-1643-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1
  • Fracture1
  • Material Split, Cut or Torn1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Cardiac Arrest1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2025Fracture; Material Split, Cut or Torn; Fracture; Material Split, Cut or Torn

    IT WAS REPORTED THAT ON (B)(6) 2025, A 4F AMPLATZER TORQVUE LP DELIVERY SYSTEM WAS SELECTED FOR USE. DURING USE, AFTER INSERTION INTO THE PATIENT, A HOLE IN THE TUBE OF THE Y-CONNECTOR WAS OBSERVED. LEAK OCCURRED IN A CONTINUOUS DRIP. NO OTHER PARTS OF THE DELIVERY SYSTEM WERE OBSERVED TO BE DAMAGED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23411268 ↗
  • InjuryJul 24, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON 01 JULY 2025, A 4F AMPLATZER TORQVUE LP DELIVERY SYSTEM WAS CHOSEN FOR A PATENT DUCTUS ARTERIOSUS (PDA) CLOSURE PROCEDURE. DURING PROCEDURE, WHILE CROSSING THE PDA, IT WAS NOTED THAT THE DELIVERY SHEATH WAS KINKED. THE PATIENT HAD A CARDIAC ARREST AND IMMEDIATELY RESUSCITATED. THE PHYSICIAN BELIEVED THE KINKED SHEATH CAUSED THE CARDIAC ARREST. A NEW 4F AMPLATZER TORQVUE LP DELIVERY SYSTEM WAS CHOSEN AND COMPLETED THE PROCEDURE. THE PATIENT WAS REPORTED STABLE.

    Patient problems: Cardiac Arrest

    MAUDE report 22592088 ↗
  • MalfunctionJul 14, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT ON (B)(6) 2025, A 4F AMPLATZER TORQVUE LP DELIVERY SYSTEM WAS SELECTED FOR USE IN A PDA OCCLUSION PROCEDURE. THE DELIVERY SYSTEM WAS INSPECTED PRIOR TO USE. DURING PROCEDURE, HEMOSTATIC VALVE LEAKAGE OCCURRED AND SALINE WAS OBSERVED TO BE LEAKING. NO PARTS OF THE DELIVERY SYSTEM WERE OBSERVED TO BE DAMAGED. THE DELIVERY SYSTEM WAS EXCHANGED FOR A NEW 4F AMPLATZER TORQVUE LP DELIVERY SYSTEM, WHICH WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THE PATIENT WAS REPORTED TO BE IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22503577 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page