Amplatzer™ 9-VSD-MUSC-004Recalls & reported problems

St. Jude Medical, Inc. · Muscular VSD Occluder · Class 3 · Implant

All models of Amplatzer (37)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00811806011783), received between Feb 13, 2019 and May 16, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Difficult or Delayed Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Separation1
  • Difficult to Fold, Unfold or Collapse1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 16, 2025Difficult to Fold, Unfold or Collapse; Difficult to Fold, Unfold or Collapse

    IT WAS REPORTED THAT ON 23 APRIL 2025, A 4MM AMPLATZER MUSCULAR VSD OCCLUDER WAS SELECTED FOR IMPLANT USING A 6F NON-ABBOTT DELIVERY SHEATH. DURING PROCEDURE, ONE OF THE DISCS DID NOT TAKE PROPER SHAPE AND DEFORMED WITH A COBRA SHAPE. THE SHAPE WAS ATTEMPTED TO BE CORRECTED MULTIPLE TIMES, BUT THE DEFORMATION PERSISTED. THERE WERE NO INTERACTIONS WITH CARDIAC STRUCTURES DURING DEPLOYMENT AND NO ANGULATION/KINK WAS OBSERVED WITH THE DELIVERY SYSTEM. THE OCCLUDER WAS NEVER RELEASED FROM THE DELIVERY CABLE WHILE INSIDE THE PATIENT. THE DEVICE WAS REMOVED AND NO DEVICE WAS ULTIMATELY IMPLANTED. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22038214 ↗
  • MalfunctionFeb 13, 2019Difficult or Delayed Separation; Difficult or Delayed Separation

    A 4 MM AMPLATZER MUSCULAR VSD OCCLUDER (MUSCVSD) WAS DEPLOYED THROUGH A 5F AMPLATZER TORQVUE 180 DELIVERY SYSTEM (DTV180) BUT WOULD NOT RELEASE FROM THE DELIVERY CABLE. THE MUSCVSD WAS RETRACTED BACK IN THE SHEATH AND REMOVED FROM THE PATIENT. A SECOND 4 MM MUSCVSD (LOT: 5414451) WAS SUCCESSFULLY IMPLANTED USING A SECOND 5F DTV180 (LOT: 5984524). PATIENT WEIGHT WAS NOT PROVIDED BY THE HEALTHCARE PROVIDER AND WILL NOT BE MADE AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8336868 ↗

Other Transcatheter septal occluder devices

All Transcatheter septal occluder devices →

Sources and updates