Zephyr Delivery Catheter EDC-TS-4.0-JRecalls & reported problems
Pulmonx Corporation · The Zephyr® Endobronchial Valve (Zephyr Valve) is an endobronchial implant designed to occlude airflow in the target lobe and thereby reduce lung volume. The device consists of a one-way silicone duckbill valve attached to a nickel-titanium (Nitinol) self-expanding retainer that is covered with a silicone membrane. It is implanted in the target bronchus during a bronchoscopy procedure using the Zephyr Endobronchial Delivery Catheter (EDC) that is guided to the targeted bronchus by inserting it through the working channel of an adult bronchoscope. Successful lobar occlusion may require placement of multiple Zephyr Valves in bronchi of the target lobe. The Zephyr Valves permit air to vent out of the treated lobe during exhalation but do not permit refilling of this region during inhalation to reduce hyperinflation due to emphysema. Trapped air in the diseased lobe escapes through the Zephyr Valves until the lung volume of the treated lobe is reduced. The Zephyr Valve is designed to be a permanent implant, but can be removed, if necessary. · Class 3
All models of Zephyr Delivery Catheter (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Pulmonx Zephyr delivery catheter is used to place one-way valves inside the lung through a bronchoscope to shrink overinflated lung tissue. The 31 reports all describe the same problem: a small sizing wing on the catheter breaking off or detaching, sometimes inside the patient and then retrieved with forceps, with the procedure continuing using a new catheter. No deaths and no injuries are mentioned, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00811907030621), received between Dec 20, 2018 and Jul 27, 2026 — classified as 1 injury, 31 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (32 reports).
1 other recall cover devices cleared under the same FDA submission (P180002); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn32
Patient problems
- No Clinical Signs, Symptoms or Conditions21
- No Known Impact Or Consequence To Patient11
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
InjuryJul 27, 2026Material Split, Cut or Torn
An incident occurred during a valve placement procedure where one of the delivery catheter's wings broke off in the patient. The wing was located and removed from the patient. The following device deficiency was not associated with an adverse event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50019396 ↗MalfunctionFeb 5, 2026Material Split, Cut or Torn
AN INCIDENT OCCURRED DURING A VALVE PLACEMENT PROCEDURE WHERE PART OF THE DELIVERY CATHETER'S WING BROKE OFF IN THE PATIENT AND WAS REMOVED WITH A GRASPER. THE FOLLOWING DEVICE DEFICIENCY WAS NOT ASSOCIATED WITH AN ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24274078 ↗MalfunctionJun 21, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ONE OF THE SIZING WINGS BROKE OFF DURING SIZING OF THE FIRST AIRWAY. THE WING WAS FOUND AND RETRIEVED. THE PROCEDURE CONTINUED WITH A NEW CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19590369 ↗MalfunctionMay 24, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
THE CATHETER WAS USED FOR ONE SUCCESSFUL DEPLOYMENT OF A VALVE. IT WAS THEN NOTED THAT ONE OF THE SIZING WINGS WAS MISSING. THE WING WAS FOUND IN THE FLUID CONTAINER. THE PROCEDURE CONTINUED WITH A NEW CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19397535 ↗MalfunctionFeb 7, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A PROCEDURE TO PLACE ZEPHYR VALVES, THE DOCTOR NOTICED THAT ONE OF THE LARGER SIZING WINGS BECAME DETACHED FROM THE CATHETER. HE TRIED TO REMOVE THE WING WITH GRASPERS BUT LOST SIGHT OF THE WING. THE PROCEDURE CONTINUED WITH A NEW CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16315231 ↗MalfunctionOct 18, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A BRONCHOSCOPIC LUNG VOLUME REDUCTION PROCEDURE ON (B)(6) 2022 WITH ZEPHYR VALVES, ONE SIZING WING ON THE ZEPHYR 4.0 J ENDOBRONCHIAL DELIVERY CATHETER (EDC) DETACHED WHILE MEASURING AND SIZING THE AIRWAYS. THE SIZING WING WAS RECOVERED AND REMOVED FROM THE PATIENT. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY COMPLETE THE PROCEDURE WITH A NEW CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15626204 ↗MalfunctionSep 28, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A BRONCHOSCOPIC LUNG VOLUME REDUCTION PROCEDURE ON (B)(6) 2022 WITH ZEPHYR VALVES, ONE SIZING WING ON THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) FELL OFF WHILE SIZING THE APICAL SEGMENT OF THE RIGHT UPPER LOBE. THE SIZING WING WAS RECOVERED AND REMOVED FROM THE BODY. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY COMPLETE THE PROCEDURE WITH A NEW CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15499159 ↗MalfunctionSep 20, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A BRONCHOSCOPIC LUNG VOLUME REDUCTION PROCEDURE ON (B)(6) 2022 WITH ZEPHYR VALVES, ONE SIZING WING ON THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) FELL OFF DURING THE PROCEDURE. THE SIZING WING WAS RETRIEVED SUCCESSFULLY WITH NO PATIENT CONSEQUENCES. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY COMPLETE THE PROCEDURE WITH A NEW CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15457350 ↗MalfunctionSep 14, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2022, THE PATIENT HAD ZEPHYR VALVES IMPLANTED. DURING THE INITIAL SIZING PART OF THE PROCEDURE, ONE OF THE LARGER SIZING WINGS FROM THE CATHETER FELL OFF. THE SIZING WING WAS EXTRACTED AND THE PROCEDURE CONTINUED WITH A DIFFERENT CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15419027 ↗MalfunctionJun 30, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A ZEPHYR VALVE PROCEDURE, A ZEPHYR 4.0 J EDC WAS USED TO SIZE. IT WAS NOTICED THAT ONE OF THE SIZING WINGS WAS BROKEN OFF AND IT COULD BE SEEN INSIDE THE PATIENT'S AIRWAY. THE BROKEN WING WAS SAFELY RETRIEVED USING BIOPSY FORCEPS. THE SIZING PROCEDURE WAS SUCCESSFULLY COMPLETED USING A NEW 4.0 J EDC AND THE VALVE WAS PLACED WITHOUT ANY OTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14881175 ↗MalfunctionJun 29, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
DURING A ZEPHYR VALVE PROCEDURE, A ZEPHYR 4.0 J EDC WAS USED TO DEPLOY A VALVE. FOLLOWING THE DEPLOYMENT, IT WAS NOTICED THAT ONE OF THE SIZING WINGS WAS MISSING. THE MISSING WING WAS NOT NOTICED INSIDE THE AIRWAYS. THE VALVE DEPLOYMENT PROCEDURE WAS SUCCESSFULLY COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14861895 ↗MalfunctionJan 6, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, A LARGE SIZING WING FROM THE CATHETER SHEARED OFF UPON RETRACTIOIN FROM THE MAIN BRONCHUS. THE SIZING WING WAS EXTRACTED FROM THE PATIENT VIA FORCEPS AND THE PROCEDURE CONTINUED WITH A DIFFERENT CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13179264 ↗MalfunctionNov 10, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, UPON INITIAL MEASURING OF THE AIRWAY, THE PHYSICIAN NOTICED A PIECE OF THE SIZING WING SHEARED OFF. THE SIZING WING WAS EXTRACTED FROM THE PATIENT AND THE PROCEDURE CONTINUED WITH A DIFFERENT CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12788307 ↗MalfunctionOct 25, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, A PIECE OF THE SMALL SIZING WING WAS VISIBLE INSIDE THE PATIENT THROUGH THE BRONCHOSCOPE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12694024 ↗MalfunctionAug 23, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, WHILE SIZING THE APICAL SEGMENT OF THE LEFT UPPER LOBE, A SIZING WING OF THE DELIVERY CATHETER BROKE IN THE PATIENT. THE SIZING WING WAS SUCCESSFULLY REMOVED USING FORCEPS, AND A NEW EDC WAS USED TO COMPLETE THE PROCEDURE WITHOUT ANY FURTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12357054 ↗MalfunctionJul 27, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. THE PHYSICIAN HAD USED THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER TO SIZE THE AIRWAY AND DELIVER A VALVE. ONE OF THE LARGER SIZING WINGS DETACHED FROM THE CATHETER, AND THE PHYSICIAN RETRIEVED IT FROM THE PATIENT WITH A SET OF GRASPERS. A NEW DELIVERY CATHETER WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12230020 ↗MalfunctionJul 18, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. THE PHYSICIAN HAD USED THE ZEPHYR 4.0-J CATHETER TO SIZE THE AIRWAY, AND THEN A 4.0-LP VALVE WAS LOADED INTO THE CATHETER. AFTER LOADING THE VALVE INTO THE CATHETER, THEY NOTICED ONE OF THE SIZING WINGS ON THE CATHETER WAS MISSING. THE VALVE WAS DEPLOYED, AND THEN A NEW DELIVERY CATHETER WAS USED TO COMPLETE THE PROCEDURE. ADDITIONAL DETAILS REGARDING THE LOCATION OF THE SIZING WING AND/OR ITS RETRIEVAL HAVE BEEN REQUESTED FROM THE PHYSICIAN, BUT HAVE NOT YET BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12186708 ↗MalfunctionMay 27, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES FOR THE RIGHT UPPER LOBE. A ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) WAS INSERTED, AND ONCE VISUALIZED, IT WAS NOTICED THAT ONE LARGE SIZING WING SHEARED OFF. THE DAMAGED DEVICE WAS SUCCESSFULLY REMOVED AND THE DETACHED WING WAS REMOVED WITH FORCEPS. A NEW ZEPHYR 4.0-J EDC WAS USED TO COMPLETE THE PROCEDURE UNEVENTFULLY, WITHOUT ANY SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11898744 ↗MalfunctionMay 17, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. WHEN ATTEMPTING TO PLACE A VALVE, THE PHYSICIAN NOTICED THAT ONE OF THE SIZING WINGS OF THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) HAD FALLEN OFF. THE SIZING WING WAS SUCCESSFULLY REMOVED WITH BIOPSY FORCEPS AND THE PROCEDURE WAS COMPLETED WITHOUT ANY FURTHER ISSUES. THE PATIENT TOLERATED THE PROCEDURE WITHOUT INCIDENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11838030 ↗MalfunctionMay 4, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, WHILE SIZING THE LB6 AIRWAY WITH EDC 4.0-J CATHETER, A SIZING TAB FELL OFF THE EDC INTO THE AIRWAY WHEN THE CATHETER WAS PULLED INTO THE BRONCHOSCOPE. THE SIZING TAB WAS REMOVED FROM AIRWAY VIA BRONCHOSCOPE SUCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11773762 ↗MalfunctionMar 11, 2021Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2021, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. ONE OF THE SMALL SIZING WINGS OF THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) BROKE OFF WHILE MEASURING AN AIRWAY. IT WAS SUCCESSFULLY REMOVED WITH FORCEPS AND DISCARDED. THERE WAS NO ADVERSE EFFECT ON THE CASE OR PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11467124 ↗MalfunctionAug 26, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2020, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. WHILE ATTEMPTING TO INSERT THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) INTO AN AIRWAY THROUGH THE BRONCHOSCOPE WHILE THE BRONCHOSCOPE WAS SEVERELY ARTICULATED, THE EDC MET RESISTANCE AND WAS EVENTUALLY FORCED INTO THE AIRWAY. ONCE THE EDC ENTERED THE AIRWAY, IT WAS OBSERVED THAT A SMALL PIECE OF THE HOUSING ON THE DISTAL END HAD BROKEN AWAY FROM THE CATHETER. THE EDC WAS REMOVED AND THEY OPTED TO USE A NEW ZEPHYR 4.0-J EDC. THE TERRITORY MANAGER REMINDED THE PHYSICIAN AND SUPPORT TEAM TO ADVANCE THE EDC TO THE END OF THE BRONCHOSCOPE PRIOR TO ARTICULATING THE BRONCHOSCOPE. THERE WERE NO FURTHER ISSUES DURING THE PROCEDURE. THE BROKEN FRAGMENT WAS NOT LOCATED OR RETRIEVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10455217 ↗MalfunctionAug 4, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2020, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. DURING AIRWAY SIZING USING THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC), SMALL SIZING WING FRAGMENTS HAD BECOME DETACHED AND WERE NOTICED IN THE AIRWAY. THESE WERE PROMPTLY SUCTIONED OUT, THE EDC WAS REMOVED, AND A NEW ZEPHYR 4.0-J EDC WAS USED FOR THE REMAINDER OF THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10365306 ↗MalfunctionJul 14, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2020, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. ONE OF THE SIZING WINGS OF THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER (EDC) DETACHED DURING THE PROCEDURE AND WAS DEPOSITED INTO THE PATIENT'S AIRWAY. THE PIECE WAS PROMPTLY REMOVED FROM THE PATIENT, AND A DIFFERENT ZEPHYR 4.0-J EDC WAS OPENED AND USED FOR THE REMAINDER OF THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10273992 ↗MalfunctionFeb 13, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ON (B)(6) 2020, THE PATIENT UNDERWENT A BRONCHOSCOPIC LUNG VOLUME REDUCTION (BLVR) PROCEDURE WITH ZEPHYR VALVES. ONE OF THE SIZING WINGS OF THE ZEPHYR 4.0-J ENDOBRONCHIAL DELIVERY CATHETER CAME OFF WHILE DEPLOYING A ZEPHYR 4.0 VALVE INTO THE APICAL SEGMENT OF THE RIGHT UPPER LOBE (RUL). THE DETACHED SIZING WING BECAME STUCK NEXT TO THE VALVE IN THE AIRWAY, AND THE PHYSICIAN WAS UNABLE TO RETRIEVE THE SIZING WING WITH RAT TOOTH GRASPERS. THE VALVE WAS EXPLANTED IN ORDER TO RETRIEVE THE DETACHED SIZING WING WHICH WAS SUCCESSFULLY REMOVED. AFTERWARDS, ANOTHER ZEPHYR 4.0 VALVE WAS SUCCESSFULLY IMPLANTED AS A REPLACEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9707222 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P180002. One submission often covers many models, so check the product and lot details in each record.
- Z-0235-2024 · Aug 31, 2023 · Open, ClassifiedZephyr Endobronchial Valve (EBV) 5.5-LP, REF: EBV-TS-5.5-LP
Other Valve, Pulmonary devices
- Zephyr Endobronchial Valve EBV-TS-4.0Pulmonx Corporation · Valve, Pulmonary36 deaths387 reports
- Zephyr Endobronchial Valve EBV-TS-4.0-LPPulmonx Corporation · Valve, Pulmonary22 deaths295 reports
- Zephyr Endobronchial Valve EBV-TS-5.5Pulmonx Corporation · Valve, Pulmonary31 deaths282 reports
- Zephyr Endobronchial Valve EBV-TS-5.5-LPPulmonx Corporation · Valve, Pulmonary1 recall24 deaths205 reports
- Spiration Valve System SVS-V9-00Gyrus Acmi, Llc · Valve, Pulmonary1 recall5 deaths100 reports
- Spiration Valve System SVS-V7-00Gyrus Acmi, Llc · Valve, Pulmonary3 deaths84 reports
- Spiration Valve System SVS-V6-00Gyrus Acmi, Llc · Valve, Pulmonary1 death31 reports
- Zephyr Delivery Catheter EDC-TS-5.5-DMPulmonx Corporation · Valve, Pulmonary23 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page