Biomet SpinalPak Non-invasive Spine Fusion Stimulator System 1067716Recalls & reported problems

Ebi, LLC · SPINALPAK ASSEMBLY · Class 2

All models of Biomet SpinalPak Non-invasive Spine Fusion Stimulator System (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Biomet SpinalPak non-invasive spine fusion stimulator has 262 adverse event reports, including 261 injuries and 1 death. Reports mostly describe pain during treatment, including back and leg pain, along with headaches, skin irritation, and muscle cramps. There is 1 open recall naming this device, started in March 2024, over damage to the lead wire sheath that exposes the wires and could stop the treatment signal from reaching the electrodes.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
262
Deaths reported
1
Injuries reported
261
Recalls naming it
1

Summary

The FDA's MAUDE database contains 262 adverse event reports that list this exact device (DI 00812301020218), received between May 1, 2019 and Sep 29, 2026 — classified as 1 death, 261 injuries.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (110 reports).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (P850022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 4 reportsMar 2024: 4 reportsApr 2024: 6 reportsMay 2024: 7 reportsJun 2024: 6 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 5 reportsOct 2024: 6 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 5 reportsMar 2025: 5 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 4 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 6 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 22, 2024Class IIOpen, Classified

    Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

    Z-1616-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem162
  • Patient-Device Incompatibility110
  • Insufficient Device Problem Information99
  • Off-Label Use19
  • Appropriate Device Problem Term/Code Not Available2
  • Device-Device Incompatibility1
  • Inadequate or Insufficient Training1
  • Inappropriate/Inadequate Shock/Stimulation1

Patient problems

  • Pain204
  • Cramp(s) /Muscle Spasm(s)16
  • Headache15
  • Skin Inflammation/ Irritation14
  • Discomfort7
  • Shock7
  • Nausea6
  • Swelling/ Edema6

Recent adverse event reports

The 25 most recent of 262 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Patient-Device Incompatibility

    Patient problems: Pain

    MAUDE report 50934222 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Pain

    MAUDE report 50790521 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient was in pain. Attempts have been made to obtain additional information; however, no further information has been provided.

    Patient problems: Pain

    MAUDE report 50754219 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient was experiencing pain while wearing the bone growth stimulator. The patient conducted a time test but this did not help.It was later reported that the first time the patient used the unit the patient had to remove the device due to throbbing pain. The pain level was rated a 10 out of 10. The patient did not increase daily activities. The patient was taking Dilaudid and Cyclobenzaprine for the pain. The patient attempted the time test at 1 hour per day and still had pain. The patient was advised by the doctor to stop using the device due to pain. No further consequences were reported.

    Patient problems: Pain

    MAUDE report 50754161 ↗
  • InjurySep 10, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Pain

    MAUDE report 50724874 ↗
  • InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient had pain on his leg at the time of the fitting for the stimulator. The patient also had low blood sugar through the night and was not feeling well. The patient also stated that he is having back pain in the surgical area and his right leg will not work. The sales representative advised the patient to stop using the stimulator. The sales representative later spoke with the doctor¿s office who advised that the patient had been having pain prior to the surgery and had issues with numbness in his leg. This is not a new issue. The nurse stated that the patient had not called the doctor¿s office for pain medicine. It was also noted that the patient is listed as a diabetic. No further information was provided.

    Patient problems: Numbness; Pain

    MAUDE report 50275165 ↗
  • InjuryJun 30, 2026Adverse Event Without Identified Device or Use Problem

    THE SALES REP REPORTED THAT THE PATIENT IS EXPERIENCING ISSUES WHILE USING THE UNIT. THE PATIENT SPOKE WITH CUSTOMER SERVICE AND STATED THAT SHE SUFFERS FROM A PRE-EXISTING CONDITION, SEVERE RESTLESS LEG SYNDROME. THE PATIENT EXPERIENCED AN EXACERBATION OF THESE SYMPTOMS WHEN SHE STARTED TREATMENT WITH THE UNIT. THE PATIENT STATED IT WAS VERY UNCOMFORTABLE; SHE GOT NO SLEEP AND WAS IN PAIN. THE PAIN LEVEL WAS 4 OUT OF 10. THE PATIENT SPOKE WITH HER SURGEON'S OFFICE, AND THEY SUGGESTED SHE CONSULT WITH HER NEUROLOGIST. THE NEUROLOGIST ADVISED THE PATIENT TO DISCONTINUE THE SPAK TREATMENT AS THE BENEFITS DO NOT OUTWEIGH THE SYMPTOMS SHE IS EXPERIENCING. NO FURTHER CONSEQUENCES WERE REPORTED.

    Patient problems: Pain

    MAUDE report 25665573 ↗
  • InjuryJun 10, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT REPORTED THAT HE EXPERIENCED PAIN WHILE TREATING WITH THE DEVICE. NO FURTHER CONSEQUENCES WERE REPORTED.

    Patient problems: Pain

    MAUDE report 25462190 ↗
  • InjuryJun 5, 2026Adverse Event Without Identified Device or Use Problem

    THE SALES REP REPORTED THAT THE PATIENT EXPERIENCED A BURNING PAIN IN HER BUTTOCKS AFTER SLEEPING WITH THE DEVICE LAST WEEKEND. THE CUSTOMER SERVICE REPRESENTATIVE CALLED THE PATIENT AND FOUND THAT THE PADS HAD BURNED HER ON HER BACK. THE PATIENT STATED SHE HAD PAIN IN HER BUTTOCKS. THE 72R ELECTRODES WERE CLOSE TO EACH OTHER, 3 INCHES APART. THE PATIENT STATED SHE HAS 2 RED MARKS ON HER BACK. SHE HAD THE ELECTRODES ON FOR 2 OR 3 DAYS WHEN SHE GOT THE BURN ON (B)(6). THE PATIENT STATED SHE HAS PAIN ON HER BUTTOCKS. THE PATIENT WENT TO HER PCP TODAY AND GOT AN INJECTION FOR THE PAIN. THE PATIENT STATED THE PAIN HAS GONE AWAY. THE PATIENT DID NOT CALL THE SURGEON. THE PATIENT HAS INCREASED HER DAILY ACTIVITIES. HER FOOT STARTED SWELLING. THE PAIN LEVEL FROM THE BURN WAS A 10. THE PATIENT ASKED HER HUSBAND TO TAKE THE ELECTRODES OFF. THE PAIN IN THE BUTTOCKS WAS A 10 AS WELL. THE PATIENT STATED THAT SHE DID NOT HAVE THIS PAIN BEFORE. THE PATIENT CANNOT SLEEP BECAUSE OF THE PAIN. THE PATIENT DOES NOT WANT TO CONTINUE TREATMENT. NO FURTHER CONSEQUENCES WERE REPORTED.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 25414848 ↗
  • InjuryJun 4, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT REPORTED THAT SHE EXPERIENCED PAIN DURING TREATMENT. THE PATIENT STATED THAT SHE USED THE UNIT FOR 1 TO 2 HOURS AND HAD TRIED TO USE THE UNIT 3 TIMES TO TREAT, AND THEN SPOKE WITH HER DOCTOR. THE DOCTOR STATED THAT SHE DOES NOT NEED TO USE IT, AS SHE IS HEALING WITHOUT IT. NO FURTHER CONSEQUENCES WERE REPORTED. THE SPINALPAK ASSEMBLY WAS RETURNED FOR EVALUATION.

    Patient problems: Pain

    MAUDE report 25396003 ↗
  • InjuryMay 21, 2026Adverse Event Without Identified Device or Use Problem

    THE SALES REP REPORTED THAT THE PATIENT EXPERIENCED PAIN DURING TREATMENT. THE PATIENT WORE THE DEVICE FROM HIS APPOINTMENT FITTING AND TOOK IT OFF AT 11PM; THERE WAS NO PAIN. THE PATIENT STATED THAT HE WOKE UP AT 1:30 AM IN SEVERE PAIN AND ALMOST WANTED TO GO TO THE EMERGENCY ROOM. THE PATIENT SAID THE PAIN WAS WHERE THE FUSION WAS AND POSSIBLY RADIATING DOWN HIS LEGS. THIS MORNING, THE PAIN HAS SUBSIDED AND IS NEAR NORMAL FOR SURGICAL HEALING PAIN. THE PATIENT STATED HE DID NOT TURN ON HIS MEDTRONIC PAIN STIMULATOR LAST NIGHT OR WHILE WEARING THE SPINAL PAK DEVICE. NO REPLACEMENT PRODUCT WAS SHIPPED. IT WAS LATER REPORTED THAT CUSTOMER SERVICE RECEIVED A CALL BACK FROM THE PATIENT, WHO STATED HE HAD PUT THE STIMULATOR ON YESTERDAY. HE STATED THE STIMULATOR CAUSED SIGNIFICANT, CONSTANT PAIN AND THAT THE PAIN KEPT HIM UP ALL NIGHT. THE PATIENT RATED THE PAIN LEVEL AT A 9. THE PATIENT DID NOT CALL THE DOCTOR. THE PATIENT ADVISED THAT THE SALES REP WOULD CONTACT HIS DOCTOR. THE PATIENT SAID THAT HE TOOK TRAMADOL AND TYLENOL FOR THE PAIN. THE PATIENT CALLED BACK AGAIN. CUSTOMER SERVICE EXPLAINED THAT ALL INFORMATION REGARDING THE DISCOMFORT HAD BEEN COLLECTED. THE PATIENT EXPLAINED HE

    Patient problems: Pain

    MAUDE report 25240909 ↗
  • InjuryMay 8, 2026Adverse Event Without Identified Device or Use Problem

    PER CS REP (B)(6), RECEIVED EMAIL FROM SALES REP, (B)(6). SUBJECT: PATIENT COMPLAINT (B)(4). HELLO, I RECEIVED THIS TEXT MESSAGE FROM THIS PATIENT OVER THE WEEKEND AND I CALLED HER BACK ON SATURDAY BUT DID NOT REACH HER AND JUST LEFT A VOICEMAIL. I JUST TRIED HER AGAIN TONIGHT AND SPOKE TO HER AND I TOLD HER THAT SOMEBODY FROM CUSTOMER SERVICE WOULD GIVE HER A CALL, BUT THAT THEY'RE PROBABLY GOING TO TRY AND RECOMMEND SHE START USING IT FOR JUST AN HOUR OR TWO A DAY AND THEN SLOWLY BUILD UP. SHE SAID THAT SHE EXPERIENCED SOME SPASM PAIN IN HER RIGHT HIP ALONG WITH SOME EDEMA. SHE TOOK IT OFF ON SATURDAY AND HASN'T PUT IT ON SINCE AND SHE ALSO HAS LOST SOME STRENGTH IN HER RIGHT LOWER EXTREMITY. IF SOMEBODY FROM CUSTOMER SERVICE, COULD PLEASE CALL HER ON WEDNESDAY TO DISCUSS THAT WOULD BE GREAT. (B)(6) THIS IS (B)(6). THE BONE GROWTH STIMULATOR CAUSED PAIN AND LOSS OF STRENGTH IN MY RIGHT LEG. FINALLY TOOK IT OFF LAST EVENING UNTIL I CAN TALK TO YOU. (B)(6). I CALLED THE PATIENT REGARDING PAIN AND LOSS OF STRENGTH IN THE RIGHT LEG. THE PATIENT STATED THAT THE STIMULATOR CAUSED PAIN ON THE RIGHT HIP AND SWELLING. THE PATIENT TOOK OFF THE UNIT AND SHE STARTED FEELING BETTER. THE PATIE

    Patient problems: Pain

    MAUDE report 25123373 ↗
  • InjuryMay 6, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN WHILE TREATING WITH THE SPINALPAK. THE PATIENT WORE THE DEVICE FOR 2 TO 3 HOURS. THE PATIENT RATED THE DISCOMFORT LEVEL AT A 2 OUT OF 10, WITH 10 BEING THE HIGHEST. THEN, AT NIGHT IN BED, THE PATIENT EXPERIENCED WORSE PAIN AT A 7 OR 8 OUT OF 10. THE PATIENT CONTINUED TREATMENT WITH THIS DISCOMFORT FOR ABOUT ONE WEEK. THEN THE PATIENT TOOK A BREAK FROM TREATMENT AND THE PAIN SUBSIDED. AT THIS POINT THE PATIENT CHOSE TO DISCONTINUE TREATMENT. THE PATIENT DID NOT DISCUSS THE PAIN WITH THEIR DOCTOR. THE PATIENT HAD NOT INCREASED THEIR ACTIVITY LEVEL. THE PATIENT REPORTED THE PAIN WAS BENEATH THE SURFACE OF THE SKIN AND FELT LIKE IT WAS IN THEIR BONES. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Pain

    MAUDE report 25093523 ↗
  • InjuryApr 15, 2026Adverse Event Without Identified Device or Use Problem

    THE PATIENT REPORTED THAT SHE HAS BEEN HAVING LOWER BACK PAIN. THE PATIENT STATED THAT SHE HAD THE SURGERY 6 AND A HALF MONTHS AGO. WHILE USING THE SPINALPAK UNIT CONSISTENTLY, THE PAIN IS A LEVEL 7. IF THE PATIENT DOES NOT USE THE SPINALPAK UNIT, THE PAIN GOES TO 4. ONCE THE SPINALPAK UNIT IS TAKEN OFF, IT TAKES ABOUT 2 DAYS FOR THE PAIN TO DISSIPATE A LITTLE BIT. THE PATIENT'S DOCTOR IS AWARE OF THIS. THE DOCTOR DOES NOT KNOW WHY SHE IS HAVING THIS PAIN AND WAS TOLD TO SPEAK TO HER PHYSICAL THERAPIST. THE PATIENT ALSO USES A TENS (TRANSCUTANEOUS ELECTRICAL NERVE STIMULATION) UNIT BUT NOT AT THE SAME TIME. THE PATIENT FEELS LIKE IT IS MUSCLE PAIN IN HER LOWER BACK. THE DOCTOR PRESCRIBED MELOXICAM AND CELEBREX FOR THE PAIN, BUT IT DID NOT WORK. THE PATIENT HAS SLOWLY INCREASED HER DAILY ACTIVITY. THE PATIENT STATED THAT THE PAIN INTENSIFIES THE MORE DAYS SHE USES THE SPINALPAK UNIT. THEN THE PATIENT STOPS USING THE SPINALPAK STIMULATOR TO USE THE TENS UNIT, AND AFTER A FEW DAYS, THE PAIN GOES DOWN, AND THEY START THE PROCESS AGAIN OF USING THE SPINALPAK STIMULATOR. THE PATIENT WAS ADVISED TO STOP USING THE SPINALPAK DEVICE FOR A FEW DAYS, THEN START THE TIME TEST AND CALL US BACK W

    Patient problems: Pain

    MAUDE report 24886390 ↗
  • InjuryMar 20, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the SpinalPak device burned the patient's leg when it was in her pocket. The unit was not hot to the touch and there was no mark on her leg. The patient advised that it was an internal burning pain. The patient noted her right leg was still sore the day after the incident. The pain level was rated an 8 out of 10. The patient called her doctor who advised the patient to stop using the unit for a week. No further consequences were reported.

    Patient problems: Pain; Skin Burning Sensation

    MAUDE report 24651901 ↗
  • InjuryMar 2, 2026Adverse Event Without Identified Device or Use Problem

    THE SALES REP ADVISED THAT THE PATIENT HAS DISCONTINUED USE OF THE UNIT PER HER DOCTOR DUE TO EAR RINGING AND VERTIGO ISSUES. THE PATIENT WAS CALLED BUT COULD NOT LEAVE A MESSAGE AS THE MAILBOX WAS FULL. NO NEW PRODUCT WAS SHIPPED TO THE PATIENT.

    Patient problems: Vertigo

    MAUDE report 24484569 ↗
  • InjuryFeb 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE PATIENT THAT THE DEVICE IS CAUSING HER PAIN. THE PATIENT STATES THAT IT STARTED ABOUT TWO WEEKS AFTER USING THE UNIT. THE PATIENT REPORTED THAT SHE HAD DISCONTINUED WEARING THE UNIT. THE PAIN LEVEL IS A 9. THE PATIENT IS TAKING PERCOCET FOR THE PAIN. THE PATIENT CALLED HER DOCTOR AND IS AWAITING A RETURN CALL. THE PATIENT WILL CALL BACK WITH AN UPDATE. NO FURTHER CONSEQUENCES WERE REPORTED. THE SPINALPAK WAS NOT RETURNED.

    Patient problems: Pain

    MAUDE report 24293973 ↗
  • InjuryJan 30, 2026Adverse Event Without Identified Device or Use Problem

    THE SALES REP ADVISED THAT THE PATIENT IS EXPERIENCING PAIN. THE PATIENT WAS CALLED AND ADVISED THAT HER NERVES AFTER 8 HOURS WERE ON EDGE AND NUMB. SHE HAS A PLATE IN HER NECK AND 2 STINTS, ASKING ABOUT LEVEL. IS TAKING NERVE MEDS AND DOES HAVE SCREWS. SHE HAS OSTEPINA. THE PATIENT WILL CONDUCT A TIMED TEST AND CALL BACK WITH ANY ISSUES. SAW THE DOCTOR AND WAS TOLD TO CALL CS. NO NEW PRODUCT WAS SHIPPED TO THE PATIENT. NO RETURN NOTED.

    Patient problems: Pain

    MAUDE report 24225508 ↗
  • InjuryJan 30, 2026Adverse Event Without Identified Device or Use Problem

    PER CS REP (B)(6) -RECEIVED EMAIL FROM SALES REP, (B)(6) " HI I JUST SPOKE TO THIS PATIENT AND HER DAUGHTER. I AM GOING TO GIVE YOU THE TIMELINE THEY GAVE ME, HER SX WAS 1/5. SHE RECEIVED THE BONE STIM AND STARTED USE ON (B)(6) - THE FIRST DAY USING IT SHE HAD INCREASE FEELING OF BOWEL MOVEMENT EVERY 20-30 MINS WEARING IT. DAY TWO ON (B)(6), SHE ENDED UP IN THE ER. DID NOT USE THE STIM THAT DAY. DAY THREE ON (B)(6), SHE WAS EXPERIENCING A LOT OF PAIN WITHIN 5 MINUTES OF WEARING THE STIM. TOOK STIM OFF AND PAIN STOPPED. SHE DID SAY SHE HAS SEVERE OSTEOARTHRITIS. NOT SURE IF THAT PLAYS A PART. I DO PERSONALLY THINK SOME OF THIS/MOST OF THIS MIGHT BE NORMAL HEALING ESPECIALLY WITH HER SURGERY BEING 7 DAYS AGO. I LET HER KNOW YOU WOULD BE IN TOUCH AND HOPEFULLY WE CAN FIGURE SOMETHING OUT SO SHE CAN STILL WEAR IT. PLEASE KEEP ME UPDATED THANK YOU. HER PHONE NUMBER IS (B)(6) WILL CALL LATER AS IT TOO EARLY TO CALL (B)(6). I CALLED THE PATIENT REGARDING THE COMPLAINT. I SPOKE WITH THE PATIENT'S DAUGHTER, (B)(6). THE FIRST DAY THE PATIENT HAD TO GO TO THE BATHROOM EVERY 30 MINUTES AND PAIN. THE PAIN LEVEL WAS LIKE AN 8 OR 9. THE PAIN SHOOTING DIRECTLY INTO HER LOWER SPINE. TWO DAYS LATER,

    Patient problems: Pain

    MAUDE report 24225347 ↗
  • InjuryJan 29, 2026Adverse Event Without Identified Device or Use Problem

    PATIENT CALLED BACK STATED THAT THE PAIN WAS SO SEVERE THAT IF SHE COULD DRIVE TO THE ER, SHE WOULD HAVE. PATIENT STATED THAT SHE STARTED WEARING THE UNIT FOR 24 HOURS AND HAS CUT BACK TO 4 HOURS. THE PATIENT WILL SEE THE DOCTOR ON (B)(6) 2026 AND WILL DISCUSS WITH HIM. THE PATIENT WILL CONDUCT A TIMED TEST. PER PATIENT RETURN CALL TO PRODUCT SURVEILLANCE DEPARTMENT - THE PATIENT STATED THAT SHE SPOKE WITH HER DOCTOR AND HE STILL BELIEVES IN THE STIMULATOR. HE AGREED THAT SHE SHOULD TRY TO WEAR THE DEVICE LESS BUT WANTS HER TO CONTINUE TREATING. THE PATIENT WORE THE DEVICE FOR 8 HOURS YESTERDAY, (B)(6) 2026, AND STATED THAT HER LEGS FEEL VERY HEAVY AND TIGHT TODAY. THE PATIENT IS USING A CREAM SIMILAR TO BENGAY TO ALLEVIATE THE TIGHTNESS IN HER LEGS. PRIOR TO YESTERDAY, THE PATIENT HAD BEEN TREATING FOR 4-5 HOURS PER DAY AND NOTED THAT THE DEVICE WAS NOT IMPACTING HER AS HEAVILY. THE PATIENT NO LONGER TREATS WILL SLEEPING. THE PATIENT WILL CONTINUE TO TREAT AS MUCH AS SHE CAN TOLERATE AS SHE WANTS TO HEAL. PATIENT WILL CONTACT REP OR CUSTOMER SERVICE WITH ANY FURTHER CONCERNS. THE PATIENT CALLED WHEN SHE RECEIVED THE LETTER. THE PATIENT ADVISED THAT THE PAIN LEVEL WAS A 7 OUT OF 10

    Patient problems: Pain

    MAUDE report 24215316 ↗
  • InjuryJan 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PAIN WHEN TREATING WITH THE SPINALPAK DEVICE. THE PATIENT WAS EXPERIENCING PAIN SINCE THE BEGINNING OF TREATMENT. THE PATIENT COULD ONLY TOLERATE 30 MINUTES OF TREATMENT BEFORE REMOVING THE DEVICE. THE PAIN LEVEL WAS RATED 6 OUT OF 10 WITH 10 BEING THE HIGHEST. IT WAS LATER REPORTED THAT THE PATIENT SPOKE WITH THE DOCTOR WHO ADVISED HIM TO STOP USING THE DEVICE DUE TO THE PAIN. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Pain

    MAUDE report 24208969 ↗
  • InjuryJan 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN AND VOMITING WHILE USING THE STIMULATOR. THE PATIENT STARTED USING THE UNIT AROUND 5:00PM AND REPORTED FEELING A KNOT IN THEIR CHEST BY THE TIME THEY WENT TO BED. THE PATIENT REMOVED THE STIMULATOR AROUND 11:30PM. THE PATIENT LATER VOMITED THEN BEGAN TO FEEL BETTER AROUND 2:30AM. THE PATIENT DOES NOT KNOW IF THE EVENT WAS RELATED TO THE SPINALPAK. THE PATIENT ADVISED THAT IT WAS AN UNCOMFORTABLE PAIN AND RATED THAT PAIN LEVEL AN 8 OUT OF 10. THE PATIENT SAID THE FEELING WAS VERY SIMILAR TO HEARTBURN BUT STATED HE DOES NOT SUFFER FROM HEARTBURN. THE PATIENT REPORTED NO FURTHER ISSUES AFTER DISCONTINUING USE OF THE DEVICE. THE PATIENT DID NOT CONTACT THEIR DOCTOR. THE PATIENT ALSO NOTED THAT THEY USED THE DEVICE PREVIOUSLY IN 2011 AND NEVER HAD ANY ISSUES.IT WAS LATER REPORTED THAT THE PATIENT TRIED TO USE THE DEVICE AGAIN AND THE CHEST TIGHTNESS RETURNED AFTER 2 HOURS OF TREATMENT. THEY DID NOT EXPERIENCE ANY NAUSEA AT THIS TIME.THE SALES REPRESENTATIVE ADVISED THE PATIENT TO DISCONTINUE USE OF THE DEVICE UNTIL SPEAKING WITH THEIR PRIMARY CARE DOCTOR. NO FURTHER INFORMATION WAS PROVIDED.ADDITIONAL INFORMATION WAS RECEIVED THAT

    Patient problems: Chest Pain; Myocardial Infarction; Vomiting

    MAUDE report 24172992 ↗
  • InjuryJan 22, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT CONTINUED TO EXPERIENCE PAIN WITH THE REPLACEMENT SPINALPAK DEVICE. THE PATIENT TRIED LIMITING THEIR USE BUT ULTIMATELY COULD NOT USE THE STIMULATOR. THE PATIENT¿S DOCTOR ALSO ADVISED TO NOT USE THE STIMULATOR. IT WAS LATER REPORTED THAT THE PATIENT RATED THE PAIN LEVEL AS A 7 OUT OF 10, WITH 10 BEING THE HIGHEST. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Pain

    MAUDE report 24147907 ↗
  • InjuryDec 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED BACK PAIN WHEN TREATING WITH THE SPINALPAK DEVICE. THE PAIN WAS BAD ENOUGH THAT THE PATIENT COULD BARELY WALK. THE PATIENT STATED THAT THE PAIN GOES AWAY WITHIN THREE HOURS OF STOPPING TREATMENT. THE PATIENT SPOKE WITH THEIR DOCTOR WHO SAID THE PAIN COULD BE RELATED TO THE WEATHER BECAUSE OF THE PATIENT¿S ARTHRITIS. THE PATIENT REPORTED THAT THE PAIN LEVEL WAS A 10 WHEN TREATING WITH THE SPINALPAK DEVICE ON THE NIGHT OF (B)(6) 2025. BY THE TIME THE PATIENT SPOKE WITH EBI CUSTOMER SERVICE ON (B)(6) 2025, THE PAIN LEVEL HAD SUBSIDED TO A 2. THE PATIENT REPORTED THAT THEY HAD BEEN USING THE DEVICE 24 HOURS PER DAY FOR 7 DAYS. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Pain

    MAUDE report 23883019 ↗
  • InjuryNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT STARTED GETTING MUSCLE SPASMS THE SECOND TIME USING THE SPINALPAK UNIT BACK IN MAY. THE PAIN LEVEL WAS RATED A 10 OUT OF 10. THE PATIENT STATED THE PAIN WAS OFF THE CHART. THE PATIENT TRIED TREATING FOR 5 MINUTES AND WAS OKAY, BUT AFTER TREATING FOR ANOTHER 10 MINUTES THE PAIN WAS EXCRUCIATING. THE PATIENT USED PAIN MEDICATION SHE HAD AT HOME. THE PATIENT STOPPED USING THE UNIT IN JUNE. IT WAS LATER REPORTED THAT THE PATIENT WAS FITTED WITH THE DEVICE AND ONLY USED IT FOR ABOUT A WEEK. THE SALES REPRESENTATIVE ADVISED THAT THE PATIENT WOULD BE RETURNING THE UNIT. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Cramp(s) /Muscle Spasm(s); Pain

    MAUDE report 23570170 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P850022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1617-2024 · Mar 22, 2024 · Open, Classified
    Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bon
  • Z-1619-2024 · Mar 22, 2024 · Open, Classified
    48" Lead Wires, Replacement Part Number: 1067724-4
  • Z-1618-2024 · Mar 22, 2024 · Open, Classified
    20" Lead Wires, Replacement Part Number: 1067724-2

Other Stimulator, Bone Growth, Non-Invasive devices

All Stimulator, Bone Growth, Non-Invasive devices →