Skip to content

Pounce Venous Thrombectomy System VTS-0018-09010Recalls & reported problems

Surmodics, Inc. · Pounce Venous Thrombectomy System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812339030319), received between May 16, 2024 and May 16, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fracture1
  • Material Deformation1
  • Off-Label Use1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 16, 2024Fracture; Off-Label Use; Detachment of Device or Device Component; Material Deformation; Fracture; Off-Label Use; Detachment of Device or Device Component; Material Deformation

    THE POUNCE VENOUS DEVICE WAS USED IN THE PATIENT'S LEFT AND RIGHT LEGS FOR THROMBECTOMY TREATMENT. A 12FR STANDARD INTRODUCER WAS USED AT BOTH POPLITEAL ACCESS POINTS. APPROXIMATELY FOUR TO FIVE DEVICE PASSES WERE PERFORMED IN THE LEFT LEG BEFORE MOVING TO THE RIGHT LEG. PRIOR TO INSERTION INTO THE RIGHT POPLITEAL ARTERY, THE DEVICE WAS DEEMED ACCEPTABLE FOR FURTHER PASSES. AFTER THE FIRST PASS, THE PHYSICIAN ENCOUNTERED RESISTANCE PULLING THE BASKET BACK INTO THE INTRODUCER SHEATH AND WAS UNABLE TO BRING THE BASKET DOWN SUFFICIENTLY WITH THE HANDLE'S CAM OR REAR SLIDE. UPON ATTEMPT TO RE-SHEATH THE BASKET WITH THE OUTER SHEATH THE BASKET FURTHER DEFORMED, AND IT WAS CONFIRMED THAT THE HANDLE WAS NO LONGER ABLE TO COMMUNICATE WITH THE DISTAL END OF THE DEVICE (INABILITY TO EXPAND/COLLAPSE BASKET). FURTHER ATTEMPTS CAUSED THE MID-SHAFT TO FULLY FRACTURE, SEPARATING THE HANDLE FROM THE DEVICE LEAVING JUST THE SHAFT OF THE DEVICE. SEVERAL ATTEMPTS TO RECAPTURE OR REMOVE THE BASKET WERE UNSUCCESSFUL, ULTIMATELY REQUIRING A VENOUS CUT-DOWN AT THE ACCESS SITE TO REMOVE THE BASKET. THE BASKET/CATHETER WAS SUCCESSFULLY REMOVED, AND IT WAS NOTED THAT THE BASKET WAS PACKED FULL WITH CLOT. PA

    Patient problems: Foreign Body In Patient

    MAUDE report 19332250 ↗

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page