Skip to content

SurVeil DCB SRV03513504015Recalls & reported problems

Surmodics, Inc. · SurVeil DCB 4.0mm x 150mm x 135cm · Class 3

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812339030388), received between Nov 19, 2025 and Nov 19, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2025Fracture; Fracture

    A 4.0MM X 150MM X 135CM SURVEIL DRUG-COATED BALLOON (LOT W240003) WAS USED TO TREAT A LESION IN THE COMMON FEMORAL ARTERY OF A 58-YEAR-OLD FEMALE PATIENT WITH A RELEVANT HISTORY OF DIALYSIS, HYPERTENSION, PERIPHERAL ARTERY DISEASE, CORONARY ARTERY DISEASE, HYPERCHOLESTEROLEMIA, AND DIABETES. DURING THE PROCEDURE, A NON-ABBOTT CLOSURE DEVICE WAS PRESENT PROXIMALLY IN THE ANATOMY. THE SURVEIL BALLOON WAS ADVANCED TO THE TARGET LESION AND INFLATED AS INTENDED. UPON INFLATION, A LOSS OF PRESSURE WAS NOTED AND BALLOON RUPTURE WAS SUSPECTED. THERE WAS RESISTANCE ENCOUNTERED WITH THE CLOSURE DEVICE DURING REMOVAL; THE BALLOON WAS PULLED AWAY FROM THE DEVICE BUT DID NOT DETACH, AND THE ENTIRE DEVICE WAS REMOVED FROM THE ANATOMY AS A SINGLE UNIT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT PROCEDURAL DELAY. THE PROCEDURE WAS COMPLETED AT THAT POINT. NO ADDITIONAL CLINICAL CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23595544 ↗

Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices

All Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page