ELI 380 ELI380-DCS11Recalls & reported problems

Mortara Instrument, Inc. · Electrocardiograph · Class 2

All models of ELI 380 (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Mortara Instrument ELI 380 is a resting electrocardiograph. Most of the 70 reports describe wireless connection trouble — the device dropping off Wi-Fi or failing to reconnect so patient records don't get transmitted — along with a few cases of data misreading and burn marks on the case. Reports mention 1 death and 2 injuries. One recall, started November 2020 and now closed, covered the radio losing its link to the wireless access point.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
1
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 00812345026917), received between Aug 4, 2020 and Aug 16, 2026 — classified as 1 death, 2 injuries, 67 malfunctions.

The most frequently reported specific device problem is “Wireless Communication Problem” (46 reports).

1 FDA recall record name this device's identifier.

3 other recalls cover devices cleared under the same FDA submission (K142105); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 1 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 11, 2020Class IITerminated

    The radio within the device can become disassociated with the wireless access point.

    Z-0553-2021 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Wireless Communication Problem46
  • Intermittent Communication Failure15
  • Patient Data Problem5
  • Overheating of Device2
  • Disconnection1
  • Display Difficult to Read1
  • Failure to Transmit Record1
  • Inaccurate Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions40
  • No Patient Involvement26
  • Cardiac Arrest2
  • Arrhythmia1
  • Death1
  • Myocardial Infarction1
  • Myocarditis1

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 16, 2026Overheating of Device

    Baxter received a report stating that an ELI380 had a burn mark above the touchpad on the top right. This occurred before use. There was no patient injury or death reported with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50267639 ↗
  • MalfunctionAug 16, 2026Overheating of Device

    Baxter received a report stating that an ELI380 had a burn mark above the touchpad on the top right. This occurred before use. There was no patient injury or death reported with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50263556 ↗
  • InjuryNov 25, 2025Inaccurate Information; Inaccurate Information

    IT WAS INITIALLY REPORTED THAT THE ELI 380 RESTING ELECTROCARDIOGRAPH READINGS WERE MISCALCULATING THE QTC AND THE PATIENT WENT INTO CARDIAC ARREST. THE CUSTOMER REPORTED THAT THE DEVICE INTERPRETED THE PATIENT¿S QTC VALUE AS 440 AND THE PHYSICIAN CALCULATED THE QTC VALUE AS 501.

    Patient problems: Cardiac Arrest

    MAUDE report 23642927 ↗
  • MalfunctionAug 25, 2025Wireless Communication Problem; Wireless Communication Problem

    CUSTOMER REPORTED DEVICE WOULD DROP WIFI CONNECTION AND WOULD NOT RECONNECT. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882263 ↗
  • MalfunctionAug 18, 2025Wireless Communication Problem; Wireless Communication Problem

    CUSTOMER REPORTED INTERMITTENT WIFI CONNECTIONS. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22825230 ↗
  • MalfunctionApr 24, 2025Wireless Communication Problem; Wireless Communication Problem

    CUSTOMER REPORTED WIFI DROPPING. THIS COMPLAINT WAS CAPTURED UNDER BAXTER REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898010 ↗
  • MalfunctionMar 17, 2025Wireless Communication Problem; Wireless Communication Problem

    CUSTOMER REPORTED DEVICE IS HAVING ISSUES CONNECTING TO THE WIRELESS LAN. THIS COMPLAINT WAS CAPTURED UNDER HILLROM REF #: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21625026 ↗
  • MalfunctionJan 2, 2025Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THE ED UNIT IS REPEATEDLY UNABLE TO TRANSMIT DUE TO BUTTONS BEING GREYED OUT ON THE ELI380. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21053155 ↗
  • MalfunctionNov 25, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THAT THE WIRELESS CARD WAS NOT WORKING. THE UNIT CONSTANTLY LOSES ITS IP ADDRESS. THERE WAS NO INJURY REPORTED. THIS INCIDENT WAS CAPTURED UNDER COMPLAINT REF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20770500 ↗
  • InjuryNov 14, 2024Failure to Transmit Record; Failure to Transmit Record

    IT WAS REPORTED THAT DURING THE USE OF THE ELI 380, THE DEVICE WOULD NOT PRINT AND DISPLAYED A PAPER QUEUE FAULT ERROR. THE CUSTOMER REPORTED THE PATIENT HAD BEEN ADMITTED FOR ABNORMAL HEART RATE AND MYOCARDITIS AND AT THE TIME OF THE EVENT THE PATIENT WAS NOTED TO BE BRADYCARDIC. A CODE WAS INITIATED, AND AN ECG WAS REQUESTED. AT THAT TIME THE ELI 380 DID NOT PRINT. THEREFORE, A STAFF MEMBER OBTAINED AN ALTERNATE ECG DEVICE FROM A NEIGHBORING AREA, THUS RESULTING IN A 3-MINUTE DELAY IN OBTAINING THE ALTERNATE DEVICE AND IDENTIFYING THAT THE PATIENT WAS IN 2ND DEGREE HEART BLOCK, WHICH COULD NOT BE IDENTIFIED ON THE FACILITY¿S 3RD PARTY MONITORING DEVICE. SIMILARLY, IT WAS REPORTED THAT PRIOR TO OBTAINING THE ALTERNATE DEVICE, BASED ON THE PATIENT¿S CLINICAL STATUS, THE DECISION WAS MADE TO QUICKLY TRANSFER THE PATIENT TO A HIGHER LEVEL OF CARE, INCLUDING THE CARDIAC CATHETERIZATION LAB AND ICU. THE CUSTOMER CONFIRMED THAT NO INJURY OCCURRED, HOWEVER, INDICATED THAT A DELAY IN CARE AND CLINICAL DECISION MAKING OCCURRED AS A RESULT OF THIS EVENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: Arrhythmia; Myocarditis

    MAUDE report 20686202 ↗
  • MalfunctionOct 29, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED NETWORK ISSUES AND BEING UNABLE TO TRANSMIT DATA. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER BAXTER COMPLAINT REF # (B)(6).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20556405 ↗
  • MalfunctionSep 23, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THE ELI380 WHEN MOVED FROM 1 PART OF THE HOSPITAL TO ANOTHER, LOSES ITS CONNECTION. IT WAS ADDITIONALLY REPORTED THAT THE UNIT GAVE A BAD READING. THERE WAS NO PATIENT HARM REPORTED. THIS INCIDENT HAS BEEN CAPTURED UNDER COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20279451 ↗
  • MalfunctionJul 30, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED WIFI CONNECTIVITY ISSUES WITH THE ELI380 UNIT. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19860110 ↗
  • MalfunctionJul 18, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THE ELI380 WAS HAVING INCONSISTENT CONNECTION ISSUES. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19770433 ↗
  • MalfunctionJul 9, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED HAVING ONGOING CONNECTIVITY ISSUES WITH THE ELI380 UNIT. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19700906 ↗
  • MalfunctionJun 26, 2024Wireless Communication Problem; Wireless Communication Problem

    THE BIOMED AT THE FACILITY REPORTED THAT THE ELI380 UNIT INTERMITTENTLY LOSES CONNECTIVITY TO THE WLAN. IT WAS REPORTED THE CONNECTION DROPPED FOR 45 MINUTES, DEVICE SHOWED NO SIGNAL AND WAS UNABLE TO DOWNLOAD ANY ORDERS. IT WAS ADDITIONALLY REPORTED THAT THE SCREEN INTERMITTENTLY SHOWS FUZZY STATIC. THERE WAS NO ADVERSE EVENT REPORTED AS A RESULT OF THIS EVENT. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19615009 ↗
  • MalfunctionMay 1, 2024Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT ECG RECORDS WERE JUMPING TO DIFFERENT INCORRECT PATIENTS ON THE ELI380 CARTS. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219365 ↗
  • MalfunctionMay 1, 2024Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT ECG RECORDS WERE JUMPING TO DIFFERENT INCORRECT PATIENTS ON THE ELI380 CARTS. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219343 ↗
  • MalfunctionMay 1, 2024Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT ECG RECORDS WERE JUMPING TO DIFFERENT INCORRECT PATIENTS ON THE ELI380 CARTS. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER BAXTER COMPLAINT REFERENCE # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219335 ↗
  • MalfunctionApr 4, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THAT ELI380 WAS HAVING INTERMITENT NETWORK ISSUES. THERE WAS NO PATIENT USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19046114 ↗
  • MalfunctionFeb 29, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THAT THE ELI380 WAS INTERMITTENTLY DROPPING WIFI CONNECTION AND NOT TRANSMITTING DATA. THIS COMPLAINT WAS CAPTURED UNDER HILLROM REF (B)(4)..

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18806430 ↗
  • MalfunctionJan 30, 2024Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THAT THE ELI380 HAS HAD PERSISTENT NETWORK ISSUES. IT WILL GO FROM HAVING A STRONG SIGNAL TO A WEAK SIGNAL OR NO SIGNAL. THERE WAS NO ASSOCIATED INJURY REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18606807 ↗
  • MalfunctionNov 30, 2023Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED A NUMBER OF ISSUES WITH THE ELI380 INCLUDING THE WIRED AND WIRELESS NETWORK CONNECTIONS WOULD SIMPLY STOP WORKING BUT THEN COME BACK WITH NO ACTION TAKEN. THE CUSTOMER ALSO STATED THAT WHEN ANY OF THOSE CONDITIONS OCCUR, THE SCREEN GOES BLANK. THERE WAS NO ASSOCIATED ADVERSE EVENT REPORTED WITH THIS COMPLAINT. THIS INCIDENT WAS CAPTURED UNDER HILL-ROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18236782 ↗
  • MalfunctionAug 23, 2023Patient Data Problem; Patient Data Problem

    THE CUSTOMER REPORTED THAT THEIR ELI 380 RESTING ELECTROCARDIOGRAPH READINGS WERE INCORRECT, PARTICULARLY THE CORRECTED QT (QTC) VALUE OF 430, AND ¿THE PATIENT SUBSEQUENTLY WENT INTO CARDIAC ARREST RELATED TO INAPPROPRIATE TIKOSYN ADMINISTRATION BASED ON QTC MEASUREMENT.¿ THE CUSTOMER REPORTED THAT THE DEVICE INTERPRETED THE PATIENT¿S QTC VALUE AS 430 AND THE PHYSICIAN CALCULATED THE QTC VALUE AS OVER 500. FOLLOW-UP WITH THE CUSTOMER CONFIRMED THAT THERE WERE NO ALTERATIONS IN THE PATIENT¿S PLAN OF CARE ASSOCIATED WITH THE RESULTS FROM THE ECG PERFORMED AT 1038 ALONG WITH THE HOSPITALIST¿S ¿WET READ¿ OF SAID ECG AT THAT SAME TIME. ADDITIONALLY, THE CUSTOMER REPORTED THAT NO SUBSEQUENT PRESCRIBED DOSES OF TIKOSYN WERE ADMINISTERED AND THAT THIS ECG WAS PERFORMED, PER FACILITY PRACTICE, TWO HOURS FOLLOWING THE LAST ADMINISTRATION OF TIKOSYN (PO AT 0913, Q12HOURS) TO DETERMINE IF THE NEXT PRESCRIBED DOSE IS INDICATED. BASED ON ADDITIONAL INFORMATION FROM THE CUSTOMER THAT NO MEDICAL DECISIONS WERE MADE TO CHANGE THE COURSE OF TREATMENT BASED ON THE ECG ¿WET READ¿ BY THE HOSPITALIST AND THE SEQUENCE OF EVENTS REPORTED TO HAVE OCCURRED ON 21JUL2023 (TIKOSYN GIVEN 0913, ECG PERFORMED 103

    Patient problems: Cardiac Arrest

    MAUDE report 17604669 ↗
  • MalfunctionMay 5, 2023Wireless Communication Problem; Wireless Communication Problem

    THE CUSTOMER REPORTED THAT THE DEVICE IS INTERMITTENTLY NOT CONNECTING TO THE WLAN NETWORK. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16875014 ↗

FDA clearance and classification

Cleared through Special 510(k) K142105 on Sep 5, 2014 (decision: Substantially Equivalent). Submitted by Mortara Instrument, Inc. as “Eli 380 Electrocardiograph”. The FDA review took 35 days. FDA record ↗

The FDA regulates product code DPS (electrocardiograph) under 21 CFR 870.2340 as a Class II device. 21 CFR 870.2340 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K142105. One submission often covers many models, so check the product and lot details in each record.

  • Z-1013-2023 · Oct 31, 2022 · Terminated
    Baxter Electrocardiograph, ELI 380 -DCS21
  • Z-1483-2021 · Mar 31, 2021 · Terminated
    ELI 380 Electrocardiograph, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediat
  • Z-1796-2015 · May 29, 2015 · Terminated
    ELI 380 Electrocardiograph. The ELI 380 is intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously a

Common questions about the ELI 380 ELI380-DCS11

Has the ELI 380 ELI380-DCS11 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0553-2021 (Class II), began on Nov 11, 2020: The radio within the device can become disassociated with the wireless access point. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ELI 380 ELI380-DCS11?

The FDA's MAUDE database lists 70 adverse event reports for this device, received between Aug 4, 2020 and Aug 16, 2026. In those reports, the most common specific device problems are “Wireless Communication Problem” (46), “Intermittent Communication Failure” (15) and “Patient Data Problem” (5); the most common patient problems are “Cardiac Arrest” (2), “Arrhythmia” (1) and “Death” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ELI 380 ELI380-DCS11?

1 report is classified as a death, 2 as injuries and 67 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ELI 380 ELI380-DCS11 FDA approved?

It was cleared through Special 510(k) K142105 on Sep 5, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the ELI 380 ELI380-DCS11, and is it still sold?

The FDA's device database lists Mortara Instrument, Inc. as the company responsible for it (brand ELI 380, model ELI380-DCS11). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrocardiograph devices

All Electrocardiograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page