Skip to content

Headway MC212156SRecalls & reported problems

Microvention, Inc. · Microcatheter · Class 2

All models of HEADWAY (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00812636020945), received between Feb 13, 2025 and Nov 24, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Protrusion/Extrusion” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Protrusion/Extrusion2
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Material Protrusion/Extrusion; Material Protrusion/Extrusion

    AS REPORTED: THE PHYSICIAN WAS TREATING A FUSIFORM ANEURYSM ON THE LEFT VERTEBRAL ARTERY. THE IMPLANTATION OF A SILK FLOWDIVERTER WAS PLANNED. A HEADWAY 21 WAS CHOSEN FOR THE PROCEDURE. IT WAS NOT POSSIBLE TO ADVANCE THE SILK (3,5-15) INTO THE HEADWAY 21. THE HEADWAY 21 WAS EXCHANGED AND THE SAME SILK WAS USED. THE EXAMINATION CONTINUED WITH GOOD RESULTS. THE PATIENT IS DOING WELL. NO INJURY CAUSED BY THIS ISSUE. NO OTHER INCIDENT INFORMATION WAS RECEIVED. WHEN THE DEVICE WAS RECEIVED FOR EVALUATION, THE INVESTIGATION FINDINGS FOUND AN EXPOSED LUMEN COIL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23636885 ↗
  • InjuryFeb 13, 2025Material Protrusion/Extrusion; Unintended Movement; Material Protrusion/Extrusion; Unintended Movement

    IT WAS REPORTED THAT THE PATIENT WAS BEING TREATED FOR AN ANTERIOR COMMUNICATING (ACOM) ARTERY ANEURYSM. THE PLANNED TECHNIQUE WAS AN H-STENT CONSTRUCT WITH 2 FLOW DIVERTING STENTS. 1 FRED X GOING FROM RIGHT A2 ¿ A1 AND 1 X SILK VISTA BABY ON LEFT A2- A1. THE SILK VISTA BABY WAS ABLE TO BE IMPLANTED SUCCESSFULLY THROUGH A HEADWAY 17. DESPITE THE PATIENT ANATOMY BEING QUITE CHALLENGING AROUND THE ARCH, AND EVEN MORE CHALLENGING AROUND THE TARGETED SITE, THE PHYSICIAN CONTINUALLY EXPERIENCED EXCESSIVE FORCES AND COMPLAINED OF THE CONCERNED HEADWAY 21 MICROCATHETER CONTINUALLY SKIPPING. ACCESS TO THE TREATMENT SITE WAS QUITE TORTUOUS AND CHALLENGING AND THE PHYSICIAN WAS UNABLE TO RE-ACCESS THE LOCATION. THE PHYSICIAN ORIGINALLY DECIDED TO STOP THE PROCEDURE ON SAFETY GROUNDS AFTER ABOUT 4 HOURS, WITH THE INTENTION OF CHANGING TO SURGICAL APPROACH. SUBSEQUENTLY DECIDED TO TRY AN ENDOVASCULAR APPROACH AGAIN IN APPROXIMATELY 6 WEEKS TIME. THE PATIENT IS FINE AND THERE HAS BEEN NO CLINICAL SEQUELAE FROM THIS CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21379767 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates