Skip to content

LVIS 212517-LVIS-D-HDERecalls & reported problems

Microvention, Inc. · Stent · Class f · Implant

All models of LVIS (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00812636022185), received between Dec 15, 2023 and Feb 23, 2026 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Ischemia Stroke1
  • Numbness1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 23, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THROUGH THE LVIS PAS CLINICAL STUDY THAT THE FOLLOWING AE FOR AN LVIS SUBJECT WAS ADJUDICATED BY CEC STUDY DEVICE-RELATED: LEFT MCA ISCHEMIC STROKE ON POD 293. THE PATIENT REPORTED THAT THEIR PENMANSHIP HAS BEEN AFFECTED BY RIGHT HAND NUMBNESS. THE LEFT MCA ISCHEMIC STROKE WAS TREATED WITH PLAVIX AND COUMADIN. BASED ON THE REVIEW OF DISCHARGE NOTE, SUBJECT HAD ONGOING SYMPTOMS OF RIGHT-HAND NUMBNESS AS OF THEIR DISCHARGE DATE (B)(6) 2021.

    Patient problems: Ischemia Stroke; Numbness

    MAUDE report 24424478 ↗
  • InjuryDec 15, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED THROUGH A LVIS PAS CLINICAL TRIAL, DURING THE INDEX PROCEDURE IMMEDIATELY AFTER THE 3RD COIL WAS PLACED, THE SURGICAL TEAM NOTICED PROGRESSIVE NARROWING IN THE MIDPORTION OF THE STENT WITH IRREGULAR MARGINS SUGGESTIVE OF THROMBUS FORMATION. A REPEAT ANGIOGRAPHY WAS PERFORMED, AND IT CONFIRMED THE THROMBUS FORMATION. THE PATIENT WAS TREATED WITH INTEGRILIN INTRAARTERIALLY OVER 5 MINUTES. POST-INTEGRILIN ANGIOGRAPHY DEMONSTRATED IMPROVED FLOW ACROSS THE STENT, WITH PERSISTENT NONOCCLUSIVE THROMBUS WITHIN THE MIDPORTION. ADDITIONAL INTEGRILIN WAS ADMINISTERED AND FLOW WAS IMPROVED FURTHER. THE THROMBUS IS STATED TO BE RELATED TO THE STUDY DEVICE AND INDEX PROCEDURE. THE THROMBUS RESOLVED WITHOUT SEQUELAE.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 18337946 ↗

Other intracranial neurovascular stent devices

All intracranial neurovascular stent devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page