Tiger 212-40-003Recalls & reported problems
Trilliant Surgical, LLC · 3.0/4.0mm Cann. Screw Driver Bit: Headless · Class 1
All models of Tiger (20)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00812926021911), received between Jun 18, 2018 and Jun 18, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Entrapment of Device1
- Material Integrity Problem1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2018Break; Entrapment of Device; Material Integrity Problem; Material Integrity Problem
WHILE PERFORMING A LAPIDUS PROCEDURE ON A MALE PATIENT IN HIS (B)(6) WITH AVERAGE BONE QUALITY, THE SURGEON EXPERIENCED A 212-40-003 DRIVER (LOT # 19592-01-01) BREAK WHILE INSERTING A 202-30-034 TIGER HEADLESS SCREW. THE SALES REPRESENTATIVE AT THE CASE, NOTED THAT THE SURGEON PRE-DRILLED AND PROX-DRILLED PRIOR TO INSERTION. THE SURGEON UTILIZED A K-WIRE AND INSERTED THE 202-30-034 TIGER SCREW WITHOUT ANY ISSUES. ONCE THE SURGICAL SITE WAS SUTURED, IT WAS REALIZED THAT THE 212-40-003 DRIVER TIP HAD BROKEN OFF DURING SURGERY. FLUORO WAS UTILIZED TO DETECT THE BROKEN PIECE THAT REMAINED IN THE PATIENT (X-RAY ATTACHED). THE SALES REPRESENTATIVE IS FOLLOWING UP WITH THE SURGEON TO SEE HOW HE PLANS TO MOVE FORWARD. INFORMATION HAS NOT BEEN RECEIVED AT THIS TIME. THE BROKEN 212-40-003 DRIVER WILL BE RETURNED TO FOR REVIEW WHILE THE 202-30-034 TIGER HEADLESS SCREW AND DRIVER TIP REMAINS IN THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 7612352 ↗
Other Screwdriver devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page