Skip to content

HEARTMATE®, POWER MODULE PATIENT CABLE 12V 2225Recalls & reported problems

Thoratec Corporation · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813024010777), received between Jan 28, 2026 and Jan 28, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Electrical Power Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical Power Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2026Electrical Power Problem

    IT WAS REPORTED THAT THE LOG FILES WERE SUBMITTED FOR EVALUATION. THE VENTRICULAR ASSIST DEVICE (VAD) COORDINATOR HAD FEEDBACK THAT ALL ALARMS WERE NOT BEING CAPTURED AND SHOWN ON THE SYSTEM MONITOR. A LOG FILES REVIEW SHOWED DATA FROM (B)(6) 2025 AT 4:21:23 TO (B)(6) 2025 AT 11:01:13. MOST OF THE EVENTS WERE ASSOCIATED WITH A (F56) V_UNABLE_CHARGE_EBB POWER FAULT FLAGS, WHICH WAS DUE TO WORN OR DAMAGED PATIENT CABLE. IT WAS RECOMMENDED THAT THE PATIENT CABLE BE REMOVED AND DISCARDED. IT WAS PROVIDING ABOUT 12.5 VOLTS TO THE SYSTEM CONTROLLER. THIS WAS NOT ENOUGH VOLTAGE FOR THE SYSTEM CONTROLLER TO OPERATE THE CHARGING CIRCUITRY FOR THE EMERGENCY BACKUP BATTERY (EBB). MOTOR FUNCTION WAS NOT AFFECTED BY THESE EVENTS. THE POWER FAULT FLAGS WERE RECORDABLE EVENTS BUT NOT ALARMS. THE ALARM HISTORY ON THE SYSTEM MONITOR WOULD NOT SHOW THESE EVENTS. THESE EVENTS WERE OCCURRING FREQUENTLY AND WAS QUICKLY FILLING UP THE LOG FILE HISTORY. THE LATEST LOG FILES WERE PROVIDED. THE PATIENT CABLE WAS EXCHANGED. THE EVENT AND PERIODIC LOGS WERE ABLE TO BE SEEN AND TALLIED ACCORDINGLY. THE LOG FILES SHOWED THE PATIENT CABLE STILL DELIVERING LESS THAN 13.5 VOLTS. THIS NUMBER SHOULD BE 14.5 V FOR A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24201412 ↗

Other Ventricular (Assisst) Bypass devices

All Ventricular (Assisst) Bypass devices →

Sources and updates